HistoCore PEGASUS 000000014048858005Recalls & reported problems
Leica Biosystems Nussloch GmbH · The HistoCore PEGASUS is an automated tissue processor specifically designed for fixation, dehydration, infiltration with intermedium and paraffin infiltration of human tissue specimens used for histological medical diagnosis by a pathologist, e.g. for cancer diagnosis. The HistoCore PEGASUS is designed for in vitro diagnostic applications. · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 14
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04049188216397), received between Apr 12, 2023 and Nov 19, 2025 — classified as 14 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 10, 2024Open, Classified
Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
Z-1947-2024 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information7
- Device Ingredient or Reagent Problem5
- Use of Device Problem4
Patient problems
- Insufficient Information8
- Unspecified Tissue Injury6
- Chemical Exposure1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryNov 19, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
ON 05 NOVEMBER 2025, LEICA BIOSYSTEMS RECEIVED INFORMATION FROM THE CUSTOMER THAT SOME PATIENT CASES ARE NOT DIAGNOSABLE. NO FURTHER INFORMATION REGARDING THE NUMBER OF UNDIAGNOSABLE CASES OR PATIENT IMPACT HAS BEEN PROVIDED TO THE MANUFACTURER TO DATE. ON 23 OCTOBER 2025, LEICA BIOSYSTEMS RECEIVED A COMPLAINT DETAILING "UNDER PROCESSED BIOPSY TISSUE ON BOTH RETORTS. THE RUNS COMPLETED WITH NO ERRORS." ON 31. OCTOBER, LEICA BIOSYSTEMS WAS INFORMED THAT THE AFFECTED SAMPLES HAD BEEN REPROCESSED. AT THIS TIME ADDITIONAL INFORMATION WAS PROVIDED STATING THAT SOME IMPACTED TISSUES STILL HAD TO BE DIAGNOSED. ON 05 NOVEMBER 2025, LEICA BIOSYSTEMS RECEIVED THE INFORMATION FROM THE CUSTOMER THAT THEY UNFORTUNATELY WILL HAVE SOME CASES THAT ARE NOT DIAGNOSABLE. THE CUSTOMER IS IN THE PROCESS OF DETERMINING WHAT CASES ARE NOT DIAGNOSABLE, AND THEY WILL THEY WILL INFORM THE LEICA BIOSYSTEMS TECHNICAL SPECIALIST ONCE THEY GET CONFIRMATION FROM THE PATHOLOGISTS.
Patient problems: Unspecified Tissue Injury
MAUDE report 23591892 ↗InjuryOct 17, 2025Use of Device Problem; Device Ingredient or Reagent Problem; Use of Device Problem; Device Ingredient or Reagent Problem
LEICA BIOSYSTEMS (LBS) RECEIVED A COMPLAINT OF UNDERPROCESSED TISSUE. ON 18 SEPTEMBER 2025, THE COMPLAINANT PROVIDED THE FOLLOWING INFORMATION TO LBS, "THERE WAS 1 CASE THAT WE WERE NOT ABLE TO GET A DEFINITIVE DIAGNOSIS. I HAVE NOT HEARD IF THEY ARE GOING TO RE-BIOPSY OR NOT THOUGH. I CAN MAKE SURE TO LET YOU KNOW ONCE I KNOW MORE." ON 23 SEPTEMBER 2025, THE LBS FIELD APPLICATIONS SPECIALIST (FAS) INVESTIGATING THIS EVENT WITH THE COMPLAINANT RECEIVED THE FOLLOWING UPDATE, "THE SUPERVISOR JUST SENT ME AN UPDATE. THERE WERE ACTUALLY 2 CASES THAT WERE UNABLE TO BE DIAGNOSED, AND THEY ARE GOING TO RE-BIOPSY." THE FOLLOWING IS THE FULL PATIENT IDENTIFIER INFORMATION FOR IMPORTER MDR 3022446399-2025-00010: FEMALE/59 SU25: MEN:3209. REFER TO IMPORTER MDR 3022446399-2025-00009 FOR INFORMATION ON THE FIRST PATIENT CASE IMPACTED.
Patient problems: Unspecified Tissue Injury
MAUDE report 23324738 ↗InjuryOct 17, 2025Use of Device Problem; Device Ingredient or Reagent Problem; Use of Device Problem; Device Ingredient or Reagent Problem
LEICA BIOSYSTEMS (LBS) RECEIVED A COMPLAINT OF UNDERPROCESSED TISSUE. ON (B)(6) 2025, THE COMPLAINANT PROVIDED THE FOLLOWING INFORMATION TO LBS, "THERE WAS 1 CASE THAT WE WERE NOT ABLE TO GET A DEFINITIVE DIAGNOSIS. I HAVE NOT HEARD IF THEY ARE GOING TO RE-BIOPSY OR NOT THOUGH. I CAN MAKE SURE TO LET YOU KNOW ONCE I KNOW MORE." ON (B)(6) 2025, THE LBS FIELD APPLICATIONS SPECIALIST (FAS) INVESTIGATING THIS EVENT WITH THE COMPLAINANT RECEIVED THE FOLLOWING UPDATE, "THE SUPERVISOR JUST SENT ME AN UPDATE. THERE WERE ACTUALLY 2 CASES THAT WERE UNABLE TO BE DIAGNOSED, AND THEY ARE GOING TO RE-BIOPSY." THE FOLLOWING IS THE FULL PATIENT IDENTIFIER INFORMATION FOR IMPORTER MDR (B)(4): FEMALE/67 SU25:CM:3233. REFER TO IMPORTER MDR (B)(4) FOR INFORMATION ON THE SECOND PATIENT CASE IMPACTED.
Patient problems: Unspecified Tissue Injury
MAUDE report 23324720 ↗InjuryOct 16, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
ON (B)(6) 2025, LEICA BIOSYSTEMS RECEIVED A COMPLAINT THAT A CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS TISSUE PROCESSOR. ACCORDING TO THE CUSTOMER, THERE WERE TWO RUNS PERFORMED ON (B)(6) 2025, AND ONE RUN PERFORMED ON (B)(6) 2025, WHICH WERE AFFECTED. THE TISSUE WAS DESCRIBED BY THE CUSTOMER AS UNDER PROCESSED. AFTER THE RUN PERFORMED ON (B)(6) 2025, THE CUSTOMER SUSPENDED THE USE OF THE DEVICE. ON (B)(6) 2025, LEICA BIOSYSTEMS NUSSLOCH GMBH WAS INFORMED THAT ONE PATIENT SAMPLE WAS UNABLE TO BE DIAGNOSED. ON (B)(6) 2025, LEICA BIOSYSTEMS NUSSLOCH GMBH RECEIVED FURTHER INFORMATION THAT THE NUMBER OF NON-DIAGNOSABLE CASES HAD INCREASED TO TWO AND THAT FOR BOTH CASES A RE-BIOPSY WOULD BE PERFORMED.
Patient problems: Unspecified Tissue Injury
MAUDE report 23308217 ↗InjuryOct 16, 2025Device Ingredient or Reagent Problem; Device Ingredient or Reagent Problem
ON (B)(6) 2025, LEICA BIOSYSTEMS RECEIVED A COMPLAINT THAT A CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS TISSUE PROCESSOR. ACCORDING TO THE CUSTOMER, THERE WERE TWO RUNS PERFORMED ON (B)(6) 2025, AND ONE RUN PERFORMED ON (B)(6) 2025, WHICH WERE AFFECTED. THE TISSUE WAS DESCRIBED BY THE CUSTOMER AS UNDER PROCESSED. AFTER THE RUN PERFORMED ON (B)(6) 2025, THE CUSTOMER SUSPENDED THE USE OF THE DEVICE. ON (B)(6) 2025, LEICA BIOSYSTEMS NUSSLOCH GMBH WAS INFORMED THAT ONE PATIENT SAMPLE WAS UNABLE TO BE DIAGNOSED. ON (B)(6) 2025, LEICA BIOSYSTEMS NUSSLOCH GMBH RECEIVED FURTHER INFORMATION THAT THE NUMBER OF NON-DIAGNOSABLE CASES HAD INCREASED TO TWO AND THAT FOR BOTH CASES A RE-BIOPSY WOULD BE PERFORMED.
Patient problems: Unspecified Tissue Injury
MAUDE report 23308216 ↗InjuryJan 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2024 LEICA BIOSYSTEMS RECEIVED THE CONFIRMATION, THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGAUS, TISSUE PROCESSOR. AS A RESULT, 5 TISSUE SAMPLES WERE UNDIAGNOSABLE, AND FIVE PATIENTS NEED RE-BIOPSY.
Patient problems: Insufficient Information; Unspecified Tissue Injury
MAUDE report 21206736 ↗InjuryDec 27, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON 28 NOVEMBER 2024 LEICA BIOSYSTEMS RECEIVED THE CONFIRMATION, THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS, TISSUE PROCESSOR. AS A RESULT THREE (3) TISSUES WERE UNDIAGNOSABLE.
Patient problems: Insufficient Information
MAUDE report 21025046 ↗InjurySep 18, 2024Use of Device Problem; Use of Device Problem
ON (B)(6) 2024 LEICA BIOSYSTEMS RECEIVED THE CONFIRMATION, THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS, TISSUE PROCESSOR. AS A RESULT, 6 SIX) TISSUE SAMPLES WERE UNDIAGNOSABLE.
Patient problems: Insufficient Information
MAUDE report 20244640 ↗InjuryFeb 7, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
ON 09 JANUARY 2024 LEICA BIOSYSTEMS RECEIVED A COMPLAINT THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS. AS A RESULT, 2 TISSUE SAMPLES NEED TO BE RECOLLECTED.
Patient problems: Insufficient Information
MAUDE report 18665108 ↗InjuryDec 22, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2023, THE COMPLAINANT CONTACTED LEICA AND DETAILED: "RETORT A TISSUE OVER PROCESSED, SLIDES NUCLEAR DETAILS ARE GONE. USER SUSPECTING SOME TYPE OF REAGENT CONTAMINATION, CARRY OVER ISSUE. HYDROMETER CHECK PERFORMED, ALL REAGENTS ARE WITHIN LIMITS. NO REAGENTS CHANGED. RETORT B TISSUE WAS FINE, AND IT WAS RUNNING AT AN HOUR APART, SAME PROTOCOL, SAME TYPE OF TISSUE." THE LAB HAS CHANGED ALL THE REAGENT INCLUDING WAX. THEY RAN TEST RUNS ON RETORT A ( 4 CASSETTES) AND RETORT B (1 CASSETTE), ALL TISSUE PROCESSED AS IT SHOULD. THEY WILL BE RUNNING MORE TESTS. THEY ARE SEEING NUCLEAR MELTDOWN WITH 150 SPECIMENS AFFECTED. THEY HAVE REQUESTED AN FSE ON SITE BEFORE THEY BEGIN TO USE THE PROCESSOR AGAIN. ON 20 DECEMBER 2023, LEICA BIOSYSTEMS NUSSLOCH RECEIVED THE FOLLOWING INFORMATION PROVIDED BY THE HISTOLOGY TECHNICAL SPECIALIST - CANADA TO THE LEICA BIOSYSTEMS SHANGHAI QUALITY ENGINEER: "DIAGNOSIS WAS MADE FOR ALL CASES HOWEVER 1 PROSTATE CASE ( 18 CASSETTES) WAS IMPACTED TO THE DEGREE IN WHICH ONE PATIENT WILL REQUIRE A REPEAT BIOPSY."
Patient problems: Insufficient Information
MAUDE report 18387677 ↗InjuryOct 13, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2023 LEICA BIOSYSTEMS RECEIVED A COMPLAINT THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR ASP6025, TISSUE PROCESSOR. THE CUSTOMER INFORMED LEICA BIOSYSTEMS, THAT AS A RESULT, 20 TISSUE SAMPLES HAVE NOT BEEN DIAGNOSED.
Patient problems: Insufficient Information
MAUDE report 17929712 ↗InjuryOct 7, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2023 LEICA BIOSYSTEMS (LBS) RECEIVED A COMPLAINT THAT THE CUSTOMER EXPERIENCED SUBOPTIMAL TISSUE PROCESSING ON THEIR HISTOCORE PEGASUS. ON (B)(6) 2023, THE COMPLAINANT INITIALLY REPORTED TO LBS THAT ALL TISSUE WAS DIAGNOSABLE. HOWEVER, ON (B)(6) 2023 THE CUSTOMER INFORMED LEICA BIOSYSTEMS, THAT IHC (IMMUNOHISTOCHEMISTRY) WAS UNABLE TO BE PROPERLY PERFORMED AND FURTHER RECOMMENDED RE-BIOPSY. ONE (1) BREAST SPECIMEN WAS NOT ENTIRELY DIAGNOSABLE DUE TO INABILITY TO RUN TESTING MARKER. WHEN ASKED IF THE STATE OF HEALTH OF THE PATIENT HAD BEEN ADVERSELY AFFECTED BY A MEDICAL DECISION OR ACTION EITHER TAKEN OR NOT TAKEN DUE TO THIS EVENT, THE COMPLAINANT REPLIED, "N/A ALTHOUGH DELAYS IN CARE, FULL DIAGNOSIS AND HAVING TO UNDERGO ANOTHER BIOPSY HAVE ALL OCCURRED."
Patient problems: Insufficient Information
MAUDE report 17892560 ↗InjuryOct 6, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE LEICA BIOSYSTEMS (LBS) MANUFACTURER CONDUCTED AN INVESTIGATION OF THIS EVENT AND DETERMINED THE INCIDENT WAS USER RELATED AND WAS MOST LIKELY CAUSED BY AN AIR TUBE BLOCKAGE, WHICH IN TURN CAUSED THE INSTRUMENT TO REPORT AN ERROR. THIS LED TO THE AFFECTED TISSUE SITTING IDLE FOR 2 HOURS IN XYLENE IN RETORT A AND CAUSED THE OBSERVED TISSUE DAMAGE TO OCCUR. BASED ON THE INVESTIGATION RESULTS AND THE ACTIONS CONDUCTED WITH THE INSTRUMENT AT THE CUSTOMER SITE, THE MANUFACTURER CONFIRMS THE INSTRUMENT, AT THIS TIME, IS WORKING ACCORDING TO FACTORY SPECIFICATIONS. TO AVOID A RECURRENCE OF THE SAME ISSUE IN THE FUTURE, A CUSTOMER FACING LETTER WAS SENT OUT TO THE CUSTOMER WITH RECOMMENDATIONS TO CONSIDER THE COUNTERMEASURES DEFINED IN THE INSTRUMENT INSTRUCTION FOR USE (IFU) REGARDING THE AIR TUBE BLOCKAGE. THE LBS MANUFACTURER'S REPORT, 8010478-2023-00011, CONTAINS ADDITIONAL INFORMATION REGARDING THE INVESTIGATION AND ROOT CAUSE.
Patient problems: Insufficient Information
MAUDE report 17886741 ↗InjuryApr 12, 2023Use of Device Problem; Use of Device Problem
ON (B)(6) 2023, THE COMPLAINANT CONTACTED LEICA BIOSYSTEMS AND REPORTED INCREASING FORMALIN FUMES IN THE TISSUE PROCESSOR, HISTOCORE PEGASUS AND THE CUT-UP ROOM IN THEIR HISTOPATHOLOGY ROOM. THE COMPLAINANT REPORTED THAT THEIR STAFF WERE FEELING LIGHT HEADED, NAUSEOUS, AND HAD HEADACHES.
Patient problems: Chemical Exposure
MAUDE report 16722353 ↗
Other PROCESSOR, TISSUE, AUTOMATED devices
- HistoCore PELORIS HistoCore PELORIS 3 (100-120V)Leica Biosystems Melbourne Pty Ltd · Processor, Tissue, Automated1 recall28 reports
- ASP6025 230V/50-60 Hz 000000014049543267Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED10 reports
- HistoCore PEGASUS Plus 000000014048858007Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED9 reports
- Leica ASP300 S - 230-240V 000000014047643514Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED5 reports
- HistoCore PEARL 000000014049350667Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED4 reports
- Leica TP1020 VF/accessories 1491020US02Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED4 reports
- Tissue-Tek VIP® 6 6040Sakura Finetek U.S.A., Inc. · PROCESSOR, TISSUE, AUTOMATED2 reports
- ASP6025 120V/50-60 Hz 000000014049543268Leica Biosystems Nussloch GmbH · PROCESSOR, TISSUE, AUTOMATED1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page