Carl Zeiss Meditec AG · The VisuMax laser keratome is a femtosecond laser. It generates a beam of ultra-short laser pulses which are guided through the aperture in the treatment objective . The optical system focuses the laser beam into the corneal tissue. The VisuMax femtosecond laser is indicated for use in ophthalmic surgery for the creation of corneal incisions in patients undergoing the following treatments: - In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring initial lamellar resection of the cornea - In patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea - In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty - In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting - Surgery or other treatment requiring initial lamellar resection of the cornea to create tunnels for placement of corneal ring segments. Additionally under class III: The VisuMax Femtosecond Laser is indicated for use in small incision lenticule extraction (SMILE) for the reduction or elimination of myopia with or without astigmatism • for spherical refractive error (in minus cylinder format) from -1.00 diopters through -10.00 diopters, • for cylinder from -0.75 diopters through -3.00 diopters, • when refraction spherical equivalent is no greater in magnitude than 10.00 diopters, in patients 22 years of age or older with documentation of stable manifest refraction over the past year as demonstrated by a change in sphere and cylinder of lower or equal 0.50 D in magnitude. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04049471095036), received between Aug 27, 2021 and Apr 7, 2026 — classified as 9 injuries, 1 malfunction.
The most frequently reported specific device problem is “Use of Device Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Use of Device Problem3
- Device Fell1
- Improper or Incorrect Procedure or Method1
- Unintended Movement1
Patient problems
- Visual Impairment7
- Eye Injury2
- Corneal Scar1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryApr 7, 2026Use of Device Problem
IT WAS REPORTED THAT DURING TREATMENT, FLAP WAS CREATED BUT IT APPEARED THAT THERE WAS NO HINGE WHEN LIFTED. CAP COMPLETELY CAME OFF.
Patient problems: Visual Impairment
MAUDE report 24811483 ↗InjuryJun 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER RECEIVED AN ANONYMOUS NOTIFICATION THROUGH BRAZILIAN AUTHORITY ANVISA, IN WHICH THE CLIENT REPORTED THAT THE DEVICE DISPLAYED A ERROR MESSAGE WHEN A FLAP TREATMENT SHOULD BE STARTED AND THE SYSTEM WAS SHUT OFF. THE CLIENT HAD TO RESTART THE EQUIPMENT TO ATTEMPT THE PROCEDURE AGAIN. AFTER REBOOTING, NO ERROR MESSAGE APPEARED, AND THEY WERE ABLE TO PERFORM THE PROCEDURE ON THE FIRST EYE (OD) SUCCESSFULLY. FOR THE SECOND EYE, THERE WAS NO ERROR MESSAGE EITHER, BUT THE FLAP TURNED OUT TOO THIN AND ENDED UP TEARING. AS A RESULT, CLIENT DECIDED TO ABORT THE PROCEDURE ON THE SECOND EYE.
Patient problems: Visual Impairment
MAUDE report 22312084 ↗InjuryMay 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT A PATIENT MAY HAVE RETAINED LENTICULE FRAGMENT, AND HIS VISION IS DECREASING.
Patient problems: Visual Impairment
MAUDE report 21998874 ↗InjuryApr 30, 2025Use of Device Problem; Use of Device Problem
A DIFFICULT CASE WAS REPORTED WITH A SUBSTANDARD OUTCOME FOR ONE PATIENT OD. ORIGINAL TREATMENT WAS A SMILE ATTEMPT WITH SUCTION LOSS IN THE POSTERIOR CUT. SURGEON PROCEEDED WITH A FLAP REPAIR CUT, LIFTED THE FLAP AND TREATED WITH AN EXCIMER LASER. THE PATIENT'S VISION WAS REPORTED TO BE 20/150, HOWEVER, IT WAS NOT SHARED IF THIS WAS UNCORRECTED VISION OR BEST CORRECTED VISION. THUS, IT IS UNCLEAR WHETHER THERE WAS A NEGATIVE IMPACT ON THIS PATIENT. IT CANNOT BE DETERMINED WHETHER THE PATIENT HAS LOST MORE THAN 2 LINES OF THE BEST CORRECTED VISUAL ACUITY (BCVA).
Patient problems: Insufficient Information
MAUDE report 21936889 ↗MalfunctionJul 16, 2024Unintended Movement; Device Fell; Unintended Movement; Device Fell
DURING SURGERY, WHILE THE PATIENT WAS LYING ON THE PSS, THE Z-MOTOR BROKE CAUSING THE PSS TO FALL DOWN. WHEN THE INCIDENT HAPPENED THE SURGERY PROCESS HAD NOT START YET. THERE IS NO REPORT OF ANY NEGATIVE IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19757494 ↗InjuryOct 24, 2023Use of Device Problem; Use of Device Problem
IT WAS REPORTED BY THE SURGEON THAT THERE WAS A COMPLICATION WITH A LENTICULE REMOVAL, AND STICKY DISSECTION. LENTICULAR INTERFACE WAS EXTREMELY STICKY ; ONE QUARTER OF SIDE CUT APPEARED TO BE MISSING. THE DOCTOR MENTIONED AN INJURY: IRREGULAR ASTIGMATISM ON PATIENT CORNEA.
Patient problems: Visual Impairment
MAUDE report 17997105 ↗InjurySep 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2023, WE RECEIVED A REPORT THAT A FOLLOW UP OD SURGERY WAS PERFORMED ON A PATIENT. THE PATIENT HAD UNDERWENT A PTK SURGERY AFTER HAVING SMILE SURGERY ON (B)(6) 2023. THE PTK PROCEDURE IS AN AFTER-TREATMENT AND AN ADDITIONAL MEDICAL INTERVENTION WHICH WAS TAKEN IN ORDER TO PREVENT PERMANENT AND/OR TEMPORARY IMPAIRMENT. AT PRESENT, THE PATIENT IS STILL IN THE RECOVERY PERIOD AFTER SURGERY, AND NO MORE TREATMENTS HAVE BEEN PERFORMED.
Patient problems: Visual Impairment
MAUDE report 17673869 ↗InjuryOct 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT THE PATIENT HAD AN UNEVENTFUL SMILE PROCEDURE OF THE LEFT EYE ON (B)(6), 2022. RIGHT EYE WAS SLIGHTLY OVERCORRECTED AND AN ENHANCEMENT WAS PERFORMED ON (B)(6), 2022, WITH FLAP AND MEL80. FLAP CREATION WAS PERFECT AND IT LIFTED FINE. HYPEROPIC MEL80 ABLATION WAS PERFORMED IMMEDIATELY AFTER THE LIFT. ONCE COMPLETED, A THIN CORNEAL "WAFER" OR "LAYER" WAS NOTICED IN THE STROMAL BED WHEN THE FLAP WAS REPOSITIONED. TWO DAYS LATER, AFTER CONSULTING WITH ANOTHER SURGEON, THE FLAP WAS REPOSITIONED AGAIN WITH CAREFUL ATTENTION TO THE WAFER.
Patient problems: Eye Injury
MAUDE report 15642280 ↗InjuryOct 13, 2022Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
THE SURGEON REPORTED THAT THE PATIENT DEVELOPED A BUTTONHOLE PARACENTRAL TO THE VISUAL AXIS WHICH HAS RESULTED IN A SMALL STROMAL SCAR. EVALUATION OF THE TREATMENT VIDEO SHOWED THAT THE REPORTED SCAR WAS A RESULT OF REPEATED SUCTION LOSS AND THE IMMEDIATE RESTART OF THE PROCEDURE WITH A DIRTY SURFACE OF THE USED TREATMENT PACK. THE CONSEQUENCE OF THE DIRTY SURFACE WAS AN INCOMPLETE CUT FOLLOWED BY MANIPULATIONS DURING LENTICULE EXTRACTIONS THAT EVENTUALLY CAUSED THE STROMAL SCAR. AN ASSESSMENT OF VISUAL ACUITY OF THE PATIENT CAN ONLY BE MADE CAUTIOUSLY BASED ON THE AVAILABLE DATA. CURRENTLY, THERE IS A LOSS OF VISUAL ACUITY AND THE PATIENT NOTES AN OPACITY THAT HAS CONTINUED TO BE MONITORED.
Patient problems: Corneal Scar; Eye Injury; Visual Impairment
MAUDE report 15597276 ↗InjuryAug 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A HEALTH CARE PROFESSIONAL (HCP) REPORTED THAT A PATIENT WHO WENT THROUGH A SMILE PROCEDURE LOST 3 LINES OF THE BEST CORRECTED VISUAL ACUITY (BCVA). THE PATIENT'S ONE EYE IS 20/20 BUT THE OTHER EYE'S BCVA IS 20/50.
Patient problems: Visual Impairment
MAUDE report 12380094 ↗
FDA clearance and classification
The FDA regulates product code HQF (laser, ophthalmic) under 21 CFR 886.4390 as a Class II device. 21 CFR 886.4390 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VisuMax
Has the VisuMax been recalled?
No FDA recall record in our data names this exact device identifier (DI 04049471095036), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VisuMax?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Aug 27, 2021 and Apr 7, 2026. In those reports, the most common specific device problems are “Use of Device Problem” (3), “Device Fell” (1) and “Improper or Incorrect Procedure or Method” (1); the most common patient problems are “Visual Impairment” (7), “Eye Injury” (2) and “Corneal Scar” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VisuMax?
None of the reports are classified as deaths, 9 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VisuMax, and is it still sold?
The FDA's device database lists Carl Zeiss Meditec AG as the company responsible for it (brand VisuMax, model VisuMax). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Laser, Ophthalmic devices
- Illuminated Stepped Laser Probe, Curved With tapered tip for use in combinatio 7510.ALCD.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic1 recall3 reports
- Directional Laser Probe with Alcon / Lumenis connector.(25 gauge / 0.5 mm) 7225.ALCD.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic1 recall0 reports
- Directional Laser Probe with Alcon / Lumenis connector.(27 gauge / 0.4 mm) 7227.ALCD.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic1 recall0 reports
- Directional Laser Probe with Alcon / Lumenis connector.(23 gauge / 0.6 mm) 7223.ALCD.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic1 recall0 reports
- EVA, combined machine including Laser (DORC connector) 8000.COM02D.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic44 reports
- EVA, combined machine including Laser (SMA connector) 8000.COM03D.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Laser, Ophthalmic20 reports
- Alcon, Engauge 8065751709Alcon Laboratories, Inc. · Laser, ophthalmic14 reports
- Alcon, Engauge 8065750991Alcon Laboratories, Inc. · Laser, ophthalmic10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page