EVA, combined machine including Laser (DORC connector) 8000.COM02Recalls & reported problems

D.O.R.C. Dutch Ophthalmic Research Center (International) B.V. · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This D.O.R.C. EVA combined laser system is used in eye surgery. Most of the 44 reports describe the machine not working as expected — irrigation or air flow stopping on its own without any alert, foot pedals that go dead mid-procedure, and material separating or breaking. Reports also mention eye injuries and pressure inside the eye rising sharply, in one case leading to retinal ischemia and loss of the eye. There are 14 injuries and no deaths reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
44
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 44 adverse event reports that list this exact device (DI 08717872019659), received between Oct 12, 2017 and Jul 5, 2018 — classified as 14 injuries, 30 malfunctions.

The most frequently reported specific device problem is “Device Operational Issue” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Operational Issue33
  • Device Operates Differently Than Expected3
  • Device Inoperable2
  • Improper Flow or Infusion2
  • Intermittent Infusion2
  • Material Separation2
  • Device Contamination with Chemical or Other Material1
  • Device Displays Incorrect Message1

Patient problems

  • No Consequences Or Impact To Patient27
  • Eye Injury4
  • Intraocular Infection4
  • Intraocular Pressure Increased4
  • Intraocular Pressure, Delayed, Uncontrolled4
  • No Code Available3
  • Loss of Vision1
  • Vitreous Hemorrhage1

Recent adverse event reports

The 25 most recent of 44 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 5, 2018Device Operational Issue; Device Operational Issue

    LOW VOLTAGE ON FOOT PEDAL, CANNOT CLEAR MESSAGE AND UNABLE TO PROCEED WITH PROCEDURE. PEDAL DOES NOT WORK WIRELESSLY NOR HARDWIRED. NEW PEDAL AND CABLES REQUESTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7664982 ↗
  • MalfunctionJul 5, 2018Device Operational Issue; Device Operational Issue

    DURING VITRECTOMY PRESSURE ROSE TO 111 MMHG AND WOULD NOT REDUCE, HIGHLY DANGEROUS TO OPTIC NERVE. UNFORTUNATELY WE CAN'T SEE WHY THE SYSTEM GOT STUCK. THEY DID A LOT OF PROGRAMMING THIS DAY MAYBE THIS CAUSED THE PROBLEM, WE DON'T KNOW. RECOMMEND TO JOIN THE SURGEONS A COMPLETE DAY AND MONITOR THE UNIT, BETTER SHOULD BE TO CHECK UP THE UNIT BEFORE USE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7664894 ↗
  • InjuryJun 29, 2018Device Operational Issue; Device Operational Issue

    DUE TO A LASER DEFECT THEY HAD TO REPRIME THE MACHINE AT A STAGE OF THE AIR INJECTION SURGICAL STEP. THIS HAS CAUSED AN INCREASE IN THE EYE PRESSURE AS THE TUBING WAS STILL CONNECTED. ADDITIONALLY INFORMATION REPORTED THAT THE INCREASE IN PRESSURE CAUSED RETINAL ISCHEMIA AND LOSS OF EYE.

    Patient problems: Eye Injury; Intraocular Pressure Increased

    MAUDE report 7651875 ↗
  • MalfunctionJun 29, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THERE WAS SOMETHING ON THE CUTTER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7651440 ↗
  • MalfunctionJun 28, 2018Device Operational Issue; Device Operational Issue

    IRRIGATION SPONTANEOUSLY SWITCHES OFF DURING OIL EXTRACTION, NO ALERT FROM THE EVA THAT THE IRRIGATION WAS SWITCHED OFF. THIS CAUSED A SOFT EYE.

    Patient problems: Intraocular Pressure, Delayed, Uncontrolled

    MAUDE report 7648215 ↗
  • MalfunctionJun 28, 2018Device Operational Issue; Device Operational Issue

    IRRIGATION SPONTANEOUSLY SWITCHED OFF. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7648088 ↗
  • InjuryJun 28, 2018Device Operational Issue; Device Operational Issue

    PLEASE NOTE: THIS INCIDENT IS RELATED TO MEDWATCH (B)(4) SUBMITTED TO FDA BY THE USER FACILITY. THIS WAS RECEIVED ON 06/19/2018. WE HAVE REACHED OUT TO THE FACILITY FOR ADDITIONAL INFORMATION.

    Patient problems: Eye Injury

    MAUDE report 7646534 ↗
  • MalfunctionJun 26, 2018Device Operational Issue; Device Operational Issue

    COMPLAINT: IRRIGATION SPONTANEOUSLY SWITCHES OFF, NO VERBAL ANNOUNCEMENT FROM EVA. IRRIGATION ICON TURNS STATIS (NO FADING EFFECT), TEXT: "IRRIGATION OFF" IS SHOWN.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7638733 ↗
  • MalfunctionJun 26, 2018Device Operational Issue; Device Operational Issue

    AFTER SALES ASKED US TO ENTER THE FOLLOWING COMPLAINT: THE SURGEON EXPERIENCED A SOFT EYE DURING A VITRECTOMY PROCEDURE (PEELING). THE EYE BECAME SOFT AND HE WAS LOOKING AT THE SCREEN AND NOTICED THAT THE IRRIGATION WAS TURNED OFF BY ITSELF. HE WAS NOT INFORMED BY THE SYSTEM ( VOICE) THAT IT WAS TURNED OFF.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7638671 ↗
  • MalfunctionJun 25, 2018Device Displays Incorrect Message; Device Operates Differently Than Expected

    DURING SURGERY PHACO WENT LOWER AND LOWER; CHANGED HP NO US POWER. ERROR CODE PHA 103/104.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7635389 ↗
  • MalfunctionJun 22, 2018Device Operational Issue; Device Operational Issue

    PUMP IN ERROR "PUMP152" AND SYSTEM NOT WORKING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7631365 ↗
  • InjuryJun 19, 2018Device Operational Issue; Device Operational Issue

    THEY STOP THE SURGERY AND TRANSFER THE PATIENT. PCB HOLDER CHARGED 2 TIMES, MAIN PEDAL CHANGED 5 TIMES . FOP ERROR BLOCKAGE DURING SURGERY. IN A REVIEW THE INFORMATION REASONABLY SUGGESTS THAT THIS PATIENT WAS UNDERGOING SURGERY WHEN THE INCIDENT OCCURRED AND THE PATIENT WAS TRANSFERRED TO ANOTHER UNIT FOR COMPLETION OF THE SURGERY. ALL INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Code Available

    MAUDE report 7616007 ↗
  • MalfunctionJun 13, 2018Material Separation; Material Separation

    VIT 3 SEVERAL PRIMING FAILS AND A BREAKDOWN DURING SURGERY. MODULE HAS BEEN REPLACED. MODULE (B)(4) HAS BEEN RETURNED. NO REPORT OF PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7597998 ↗
  • InjuryJun 11, 2018Intermittent Infusion; Improper Flow or Infusion; Intermittent Infusion; Improper Flow or Infusion

    DURING THE SURGERY, THE EVA GAVE IRREGULAR IRRIGATION PRESSURE, WITH NO BACKFLUSH. HYPOTONIC EYE OCCURRED. IT WAS REPORTED NO PATIENT HARM HAD OCCURRED.

    Patient problems: Intraocular Pressure Increased

    MAUDE report 7586343 ↗
  • InjuryJun 8, 2018Intermittent Infusion; Improper Flow or Infusion; Intermittent Infusion; Improper Flow or Infusion

    DURING THE SURGERY, THE EYE PRESSURE WAS TOO HIGH. WITH ONLY AT 4 MMHG, THE VESSELS BECAME RED. NO BACKFLUSH. IRREGULAR IRRIGATION PRESSURE.

    Patient problems: Intraocular Pressure Increased

    MAUDE report 7582754 ↗
  • MalfunctionJun 5, 2018Device Operational Issue; Device Operational Issue

    A REPORT FROM (B)(6) HAS BEEN RECEIVED AND REPORT A SOFT EYE ISSUE WHEN SWITCHING FROM PHACO TO VITRECTOMY IN 25 COMBINED PROG. AT OUR CUSTOMER THERE ARE INFUSION PROBLEMS ON EVA. THE SURGEON STILL HEARS THE SOUND OF THE DROPS/DROP NOISE, BUT THERE IS NO INFUSION, WHY THE EYES COLLAPSE. THERE IS NO INFORMATION ON INTERVENTION PROVIDED.

    Patient problems: Intraocular Pressure, Delayed, Uncontrolled

    MAUDE report 7568989 ↗
  • MalfunctionJun 5, 2018Device Operational Issue; Device Operational Issue

    AT OUR CUSTOMER THERE ARE INFUSION PROBLEMS ON EVA. THE SURGEON STILL HEARS THE SOUND OF THE DROPS/DROP NOISE, BUT THERE IS NO INFUSION, WHY THE EYES COLLAPSE.

    Patient problems: Intraocular Pressure, Delayed, Uncontrolled; No Consequences Or Impact To Patient

    MAUDE report 7568814 ↗
  • MalfunctionMay 24, 2018Device Operates Differently Than Expected

    MAINBOARD IS GOING TO BOOTLOADER MODUS BY ITSELF DURING SHAVING (NEAR RETINA ) PROCEDURE. ALL WENT RED AND UNIT MUST BE SWITCHED OFF FROM BACKSIDE DURING SURGERY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7541299 ↗
  • MalfunctionMay 23, 2018Device Operational Issue; Device Operational Issue

    DURING SURGERY, THE SURGEON WAS SWITCHING BETWEEN ANTERIOR (I/A) AND POSTERIOR (VITRECTOME) MODES. WHEN HE SWITCHED OFF THE INFUSION ON HIS FOOT PEDAL DURING I/A MODE, THE MACHINE ASPIRATED AT THE POSTERIOR (CORE VITRECTOMY) MODE, BUT THERE WAS NO INFUSION (THE INFUSION WAS OFF) IN THE MEANWHILE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7536664 ↗
  • MalfunctionMay 22, 2018Device Operational Issue; Device Operational Issue

    PUMP ERROR 152 - HIGH TEMPERATURE = MOTOR FAILURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7535386 ↗
  • MalfunctionMay 22, 2018Material Separation; Material Separation

    THEIR PHACO FRAGMENTATION TIP BROKE DURING SURGERY. MOST LIKELY LOT NUMBERS (B)(6) 2015 NO SAMPLE AND NO REPORT OF INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7535346 ↗
  • MalfunctionMay 22, 2018Device Operational Issue; Device Operational Issue

    HARDWARE ERROR DETECTED IN THE MAIN PEDAL, ERROR MESSAGE FOP 5007. MAIN PEDAL MUST BE REPLACED. PLEASE NOTE IT WILL BE RETURNED TO ZUIDLAND TOGETHER WITH THE PUMP.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7535289 ↗
  • MalfunctionMay 18, 2018Device Operational Issue; Device Operational Issue

    THE HEAD SURGEON INDICATES, THAT AT SEVERAL OCCASIONS DURING SURGERY, HE SAW THE BLOOD VESSELS CONTRACTING AND FELT THAT THE PRESSURE IS MUCH HIGHER THAN THE 20MMHG DISPLAYED IN THE SETTINGS. HE WAS QUESTIONING IF THE BOTTLE PRESSURE DISPLAYED IS CORRECT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7526214 ↗
  • MalfunctionMay 15, 2018Device Operational Issue; Device Operational Issue

    PUMP ERROR CAR1600 ( SERVO ERROR PUMP ACTUATOR ) OCCURS DURING PHACO. IT FOLLOWS AFTER EACH OTHER. SAMPLE AVAILABLE. ALL INFORMATION AVAILABLE HAS BEEN DISCLOSED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7516818 ↗
  • MalfunctionMay 15, 2018Device Operational Issue; Device Operational Issue

    DORC BECAME AWARE OF A POTENTIAL PERFORMANCE ISSUE CONCERNING THE VFI ADAPTER OF THE VFI PACK FOR SILICONE OIL SYRINGES DURING USE IN COMBINATION WITH GLASS PREFILLED SILICONE OIL SYRINGE. INVESTIGATION SHOWED THAT THE DORC VFI PACK FOR SILICONE OIL SYRINGE (1363.DD) CAN LEAD TO BREAKING OF THE GLASS SYRINGE WHEN MOUNTING THE SYRINGE IN THE ADAPTER. NO INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7515521 ↗

FDA clearance and classification

Cleared through 510(k) K142877 on Mar 13, 2015 (decision: Substantially Equivalent). Submitted by Dutch Ophthalmic Research Center International BV as “EVA,anterior machine, EVA, combined machine, EVA combined machine with laser(DORC connector), EVA combined machine with laser SMA connector)”. The FDA review took 163 days. FDA record ↗

The FDA regulates product code HQF (laser, ophthalmic) under 21 CFR 886.4390 as a Class II device. 21 CFR 886.4390 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the EVA, combined machine including Laser (DORC connector) 8000.COM02

Has the EVA, combined machine including Laser (DORC connector) 8000.COM02 been recalled?

No FDA recall record in our data names this exact device identifier (DI 08717872019659), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the EVA, combined machine including Laser (DORC connector) 8000.COM02?

The FDA's MAUDE database lists 44 adverse event reports for this device, received between Oct 12, 2017 and Jul 5, 2018. In those reports, the most common specific device problems are “Device Operational Issue” (33), “Device Operates Differently Than Expected” (3) and “Device Inoperable” (2); the most common patient problems are “Eye Injury” (4), “Intraocular Infection” (4) and “Intraocular Pressure Increased” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EVA, combined machine including Laser (DORC connector) 8000.COM02?

None of the reports are classified as deaths, 14 as injuries and 30 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EVA, combined machine including Laser (DORC connector) 8000.COM02 FDA approved?

It was cleared through 510(k) K142877 on Mar 13, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the EVA, combined machine including Laser (DORC connector) 8000.COM02, and is it still sold?

The FDA's device database lists D.O.R.C. Dutch Ophthalmic Research Center (International) B.V. as the company responsible for it (brand EVA, combined machine including Laser (DORC connector), model 8000.COM02). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Laser, Ophthalmic devices

All Laser, Ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page