Fluency® Endovascular Stent Graft FEM08100Recalls & reported problems
Bard Peripheral Vascular, Inc. · FLUENCY® PLUS Endovascular Stent Graft 8 mm x 100 mm (80 cm delivery system) · Class 3 · Implant
All models of Fluency Endovascular Stent Graft (51)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04049519008516), received between Sep 27, 2016 and Apr 6, 2023 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Fracture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture6
- Misfire4
- Positioning Failure4
- Detachment of Device or Device Component3
- Break2
- Difficult to Remove2
- Malposition of Device2
- Difficult or Delayed Positioning1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2023Break; Fracture; Difficult to Remove; Misfire; Break; Fracture; Difficult to Remove; Misfire
IT WAS REPORTED THAT DURING TREATMENT OF A LEFT ARM STENT GRAFT CANNULATION ZONE AREA, THE BLUE PORTION OF THE CATHETER CLOSEST TO DEPLOYMENT TUOHY BROKE. IT WAS FURTHER REPORTED THAT THE STENT COULD NOT BE DEPLOYED AND HAD TO BE REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16689903 ↗MalfunctionMar 3, 2023Misfire; Malposition of Device; Detachment of Device or Device Component; Misfire; Malposition of Device; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE FOR THE LEFT UPPER ARM LOOP GRAFT REPAIR OF ENLARGING ANEURYSM. THE STENT ALLEGEDLY PARTIALLY DEPLOYED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16478590 ↗MalfunctionOct 17, 2022Positioning Failure; Misfire; Difficult to Advance; Positioning Failure; Misfire; Difficult to Advance
IT WAS REPORTED THAT DURING A STENT PLACEMENT PROCEDURE, THE STENT ALLEGEDLY FAILED TO DEPLOY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15613714 ↗MalfunctionJan 9, 2020Break; Positioning Failure; Fracture; Detachment of Device or Device Component; Break; Positioning Failure; Fracture; Detachment of Device or Device Component
THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL FEM08100 ENDOVASCULAR STENT GRAFT ALLEGEDLY EXPERIENCED A BREAK, FAILURE TO DEPLOY, AND DETACHMENT OF DEVICE OR DEVICE COMPONENT. THIS INFORMATION WAS RECEIVED FROM A SINGLE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO REPORTED PATIENT CONSEQUENCE. THE PATIENT WAS REPORTED AS A (B)(6) MALE WEIGHING (B)(6) LBS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9570385 ↗MalfunctionJul 11, 2019Positioning Failure; Fracture; Material Perforation; Positioning Failure; Fracture; Material Perforation
THIS REPORT SUMMARIZES ONE (1) MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL FEM08100 ENDOVASCULAR STENT GRAFT ALLEGEDLY EXPERIENCED FAILURE TO DEPLOY, A FRACTURE, AND MATERIAL PERFORATION. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THE MALFUNCTION INVOLVED A PATIENT WITH NO KNOWN IMPACT TO THE PATIENT. THE MALE PATIENT'S AGE AND WEIGHT WERE NOT PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8783335 ↗MalfunctionDec 4, 2018Positioning Failure; Fracture; Positioning Failure; Fracture
IT WAS REPORTED DURING STENT GRAFT DEPLOYMENT IN CEPHALIC ARCH, THE STENT GRAFT ALLEGEDLY DID NOT DEPLOY. IT WAS FURTHER REPORTED THAT ANOTHER STENT GRAFT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8131180 ↗MalfunctionOct 31, 2018Fracture; Difficult to Remove; Misfire; Fracture; Difficult to Remove; Misfire
IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE TO DECLOT THE AVF USING A BAREBACK APPROACH, THE STENT GRAFT ALLEGEDLY PARTIALLY DEPLOYED AND THE SHAFT ALLEGEDLY BROKE. IT WAS FURTHER REPORTED THAT THE STENT GRAFT WAS REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8026116 ↗MalfunctionJun 1, 2017Difficult or Delayed Positioning; Fracture; Malposition of Device; Difficult or Delayed Positioning; Fracture; Malposition of Device
IT WAS REPORTED THAT DURING DEPLOYMENT OF THE STENT GRAFT FOR TREATMENT OF A PRE-DILATED THROMBOSIS OF AN EXTRAVASATION OF AN OUTFLOW VEIN OF 100 MM LENGTH ACROSS A TIGHT BEND VIA ACCESS THROUGH THE LEFT UPPER ARM, THE STENT GRAFT COULD NOT BE RELEASED ANY FURTHER AFTER BEING DEPLOYED 3/4 AND GOT STUCK INSIDE THE DELIVERY SYSTEM. DURING THE ATTEMPT TO DEPLOY THE STENT GRAFT, THE SHEATH OF THE DELIVERY SYSTEM BROKE. FINALLY, THE STENT GRAFT COULD BE DEPLOYED BUT WAS NOT IN THE CORRECT POSITION. TWO ADDITIONAL STENT GRAFTS WERE REQUIRED TO BE PLACED; ONE STENT WAS USED TO RE-STENT THE FRONT END TO COVER THE TARGET AREA AND ANOTHER STENT WAS IMPLANTED TO COVER THE BACK END OF THE STENT GRAFT WHICH WAS IN THE WRONG POSITION. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6606406 ↗MalfunctionSep 27, 2016Material Invagination; Failure to Advance; Detachment of Device or Device Component; Material Invagination; Failure to Advance; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE ATTEMPT TO LOAD THE STENT DELIVERY SYSTEM ONTO A 0.035" GUIDE WIRE PRIOR TO THE PROCEDURE, THE OUTER TIP OF THE DELIVERY SYSTEM BECAME DETACHED. REPORTEDLY, THE BLUE PORTION OF THE CATHETER WAS NOTICEABLE. ANOTHER STENT GRAFT WAS USED WITH THE SAME GUIDE WIRE TO PERFORM THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 5980142 ↗
Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page