Fluency Endovascular Stent GraftRecalls & reported problems across all models
Bard Peripheral Vascular, Inc. · 51 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 532
- Deaths reported
- 3
- Injuries reported
- 49
- Recalls
- 0
We found 51 models of Fluency Endovascular Stent Graft by Bard Peripheral Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 532 adverse event reports in the FDA's MAUDE database, received between Sep 23, 2015 and Sep 29, 2026 — classified as 3 deaths, 49 injuries, 480 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (301 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Fluency Endovascular Stent Graft problems
Device problems
- Positioning Failure301
- Misfire238
- Fracture163
- Material Perforation118
- Break84
- Detachment of Device or Device Component54
- Material Deformation28
- Malposition of Device24
Patient problems
- No Consequences Or Impact To Patient331
- No Clinical Signs, Symptoms or Conditions160
- No Known Impact Or Consequence To Patient33
- Foreign Body In Patient8
- Injury5
- Extravasation4
- No Patient Involvement4
- Vascular System (Circulation), Impaired4
Fluency Endovascular Stent Graft models
51 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Fluency® Endovascular Stent Graft FEM12060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 death46 reports
- Fluency® Endovascular Stent Graft FEM10120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment39 reports
- Fluency® Endovascular Stent Graft FEM12040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment35 reports
- Fluency® Endovascular Stent Graft FEM14060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment32 reports
- Fluency® Endovascular Stent Graft FEM14120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 deaths29 reports
- Fluency® Endovascular Stent Graft FEM08060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment26 reports
- Fluency® Endovascular Stent Graft FEM14040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment25 reports
- Fluency® Endovascular Stent Graft FEM14080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment24 reports
- Fluency® Endovascular Stent Graft FEM08040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment23 reports
- Fluency® Endovascular Stent Graft FEM10080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment23 reports
- Fluency® Endovascular Stent Graft FEM10100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment19 reports
- Fluency® Endovascular Stent Graft FEM10060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment18 reports
- Fluency® Endovascular Stent Graft FEM12080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment16 reports
- Fluency® Endovascular Stent Graft FEM07060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment15 reports
- Fluency® Endovascular Stent Graft FEM08080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment15 reports
- Fluency® Endovascular Stent Graft FEM08120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment13 reports
- Fluency® Endovascular Stent Graft FEM12120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment12 reports
- Fluency® Endovascular Stent Graft FEM08100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment9 reports
- Fluency® Endovascular Stent Graft FEM09080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment8 reports
- Fluency® Endovascular Stent Graft FEM07080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment7 reports
- Fluency® Endovascular Stent Graft FEL10080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment6 reports
- Fluency® Endovascular Stent Graft FEM06080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment6 reports
- Fluency® Endovascular Stent Graft FEM09060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment6 reports
- Fluency® Endovascular Stent Graft FEM09120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment6 reports
- Fluency® Endovascular Stent Graft FEM10040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment6 reports
- Fluency® Endovascular Stent Graft FEM09040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment5 reports
- Fluency® Endovascular Stent Graft FEM09100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment5 reports
- Fluency® Endovascular Stent Graft FEM14100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment5 reports
- Fluency® Endovascular Stent Graft FEL12060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEL12120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEM06060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEM06100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEM07120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEM12100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment4 reports
- Fluency® Endovascular Stent Graft FEL08060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment3 reports
- Fluency® Endovascular Stent Graft FEL10120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment3 reports
- Fluency® Endovascular Stent Graft FEL06080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEL07080System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEL09120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEL14040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEL14100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEM06120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEM07040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEM07100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment2 reports
- Fluency® Endovascular Stent Graft FEL06100System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL06120System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL07060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL08040System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL09060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL10060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
- Fluency® Endovascular Stent Graft FEL14060System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 report
Recent Fluency Endovascular Stent Graft adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Positioning Failure; Misfire
On (b)(6)2026, a patient underwent a stent placement procedure using a Flunecy Plus device for subclavian and recurrent stenosis. Reportedly, the stent did not deploy, resulting in partial deployment misfire of the stent delivery system, with the stent remaining in a partially deployed condition and the outer sheath becoming heavily elongated. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50928448 ↗MalfunctionJul 27, 2026Fracture; Misfire; Malposition of Device; Detachment of Device or Device Component
On (b)(6) 2026, a 74-year-old male patient underwent treatment for femoral artery stenosis/iliac artery aneurysm via angiography using a Fluency Plus Endovascular Stent Graft at the left upper arm venous graft anastomosis site. During the stent procedure, the Fluency stent failed to fully deploy. The delivery system detached and became loose, resulting in unintended stent movement during the attempted deployment. Additionally, the sheath fractured during the procedure. The patient is stable, and there was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50021161 ↗MalfunctionJul 15, 2026Difficult or Delayed Positioning; Detachment of Device or Device Component
On (b)(6) 2026, a patient underwent a stent placement procedure using a Fluency Plus Endovascular Stent Graft for the treatment of stenosis at the venous anastomosis via a left upper arm graft. During the procedure, while the stent was being deployed, it was observed that the clear plastic knob on the back of the Tuohy was loose and became disconnected, making stent deployment more difficult. The patient is doing well, with no reported complications during or after the procedure. There was no reported patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25822439 ↗MalfunctionApr 24, 2026Break; Difficult to Advance
ON (B)(6) 2026, A PATIENT UNDERWENT STENT PLACEMENT PROCEDURE USING FLUENCY PLUS. REPORTEDLY, FLUENCY DEPLOYED IN THE COMMON ILIAC, RESISTANCE ADVANCING THEY DELIVERY SYSTEM, STENT BROKE IN THE PATIENT, AND THEY HAD TO SNARE OUT PIECES OF THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24979530 ↗MalfunctionApr 13, 2026Positioning Failure; Misfire
ON (B)(6) 2026, A PATIENT UNDERWENT ANGIOPLASTY STENT PLACEMENT PROCEDURE USING FLUENCY PLUS ENDOVASC STENT GRAFT. DURING THE PROCEDURE, IT WAS NOTICED THAT THE FLUENCY STENT ALLEGEDLY DID NOT DEPLOY. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24856543 ↗MalfunctionApr 2, 2026Failure to Advance; Misfire; Material Deformation
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING A FLUENCY STENT. DURING ADVANCEMENT OF THE STENT OVER THE GUIDEWIRE, RESISTANCE WAS ENCOUNTERED, PROMPTING REMOVAL OF THE DEVICE. UPON REMOVAL, A KINK IN THE STENT WAS IDENTIFIED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24761946 ↗MalfunctionMar 25, 2026Premature Activation
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING THE FLUENCY PLUS ENDOVASCULAR STENT. IT WAS REPORTED THAT DURING THE PROCEDURE, THEY ATTEMPTED TO BAR BACK THE STENT, BUT UPON ENTERING THE PATIENT, THE STENT ALLEGEDLY PREMATURELY DEPLOYED. THE PHYSICIAN WAS ABLE TO SUCCESSFULLY REMOVE THE STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24685099 ↗MalfunctionMar 3, 2026Break; Migration or Expulsion of Device; Malposition of Device; Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE BY USING THE FLUENCY STENT SYSTEM. IT WAS REPORTED THAT DURING DEPLOYMENT, THE DEPLOYMENT SYSTEM EXPERIENCED A MECHANICAL FAILURE AND ALLEGEDLY BROKE OFF. DESPITE THE BREAKAGE, THE FLUENCY STENT WAS FULLY DEPLOYED, AND THE DEPLOYMENT SYSTEM WAS REMOVED FROM THE PATIENT WITHOUT COMPLICATION. IT WAS NOTED THAT THE STENT WAS INADVERTENTLY RETRACTED APPROXIMATELY 1TO 2 CM FROM THE INTENDED LANDING ZONE. THE IMPLANTING PHYSICIAN CONFIRMED THAT ALTHOUGH THE FINAL POSITION WAS NOT IDEAL, THE CLINICAL OUTCOME REMAINED ACCEPTABLE. SUBSEQUENT INVESTIGATION INDICATED THAT THE HEMOSTATIC VALVE OF THE TOUHY BORST Y ADAPTER MAY HAVE BEEN EXCESSIVELY LOOSENED, LEADING TO THE VALVE DETACHING DURING THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24493969 ↗MalfunctionFeb 17, 2026Fracture; Misfire
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING THE FLUENCY DEVICE. THE DEVICE WAS PREPARED IN THE STANDARD MANNER AND ADVANCED TO THE INTENDED POSITION. DEPLOYMENT WAS INITIATED; HOWEVER, THE DEVICE WAS UNABLE TO COMPLETE THE DEPLOYMENT PROCESS. SIGNIFICANT PULLING FORCE WAS APPLIED IN AN ATTEMPT TO ACHIEVE FULL DEPLOYMENT, DURING WHICH THE DEVICE FRACTURED. THE DEVICE WAS SUBSEQUENTLY REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24361886 ↗MalfunctionFeb 16, 2026Positioning Failure; Fracture
ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE FOR INTRAGRAFT VIA THE RIGHT UPPER ARM USING FLUENCY PLUS ENDOVASCULAR STENT GRAFT. DURING THE PROCEDURE, IT WAS REPORTED THAT THE STENT ALLEGEDLY DID NOT DEPLOY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24342490 ↗MalfunctionFeb 13, 2026Break; Misfire
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING FLUENCY FOR POPLITEAL ANEURYSM IN PATIENT WITH HEPARIN ALLERGY VIA CONTRALATERAL FEMORAL ARTERY. REPORTEDLY, THE STENT MALFUNCTIONED AND ALLEGEDLY BROKEN. THE DEVICE WAS PARTIALLY DEPLOYED AND EVENTUALLY REMOVED FROM THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24330033 ↗MalfunctionFeb 10, 2026Break; Detachment of Device or Device Component
ON (B)(6) 2026, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING A FLUENCY PLUS ENDOVASCULAR STENT GRAFT DEVICE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE STENT DEPLOYED SUCCESSFULLY, BUT WHILE REMOVING THE STENT SHAFT, THE TIP BROKE OFF AND HAD TO BE RETRIEVED USING A SNARE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24301885 ↗MalfunctionJan 9, 2026Entrapment of Device; Material Perforation; Misfire; Activation Failure
ON (B)(6) 2025, A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING FLUENCY STENT. DURING THE PROCEDURE THE STENT HAD STRUT STICKING OUT PRE DEPLOYMENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24018126 ↗MalfunctionDec 12, 2025Positioning Failure; Material Separation; Positioning Failure; Material Separation
A PATIENT UNDERWENT A STENT PLACEMENT PROCEDURE USING FLUENCY PLUS STENT. IT WAS REPORTED THAT THE FLUENCY DID NOT DEPLOY. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23787762 ↗MalfunctionDec 4, 2025Misfire; Misfire
A PATIENT UNDERWENT STENT GRAFT PLACEMENT PROCEDURE USING FLUENCY STENT GRAFT. DURING PROCEDURE, THE STENT PARTIAL DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23714961 ↗
Other Bard Peripheral Vascular, Inc. products
- MAX-CORE11 models · 8,845 reports · 1 recall
- DENALI Vena Cava Filter2 models · 1,021 reports
- MAGNUM19 models · 571 reports · 1 recall
- Conquest32 models · 560 reports
- Ultraverse RX56 models · 536 reports
- Atlas Gold33 models · 427 reports
- Ultraverse 035 PTA Dilatation Catheter92 models · 393 reports
- Dorado49 models · 387 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.