Skip to content

Fluency® Endovascular Stent Graft FEM12100Recalls & reported problems

Bard Peripheral Vascular, Inc. · FLUENCY® PLUS Endovascular Stent Graft 12 mm x 100 mm (80 cm delivery system) · Class 3 · Implant

All models of Fluency Endovascular Stent Graft (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04049519008660), received between May 11, 2017 and Jun 12, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component3
  • Positioning Failure2
  • Adverse Event Without Identified Device or Use Problem1
  • Malposition of Device1

Patient problems

  • Foreign Body In Patient2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2024Break; Malposition of Device; Detachment of Device or Device Component; Break; Malposition of Device; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE FOR A DISEASED ACCESS SITE SEGMENT THROUGH THE ARTERIAL LIMB OF A THIGH LOOP GRAFT, THE STENT ALLEGEDLY JUMPED MISSING THE TARGET LESION. IT WAS FURTHER REPORTED THAT THE CATHETER TIP ALLEGEDLY BROKE OFF IN THE VESSED DURING DEPLOYMENT. REPORTEDLY, THE BROKEN VISIBLE MARKER WAS CAPTURED AGAINST THE WALL OF THE GRAFT BY USING ANOTHER STENT. THE CURRENT STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 19516851 ↗
  • InjuryJan 11, 2024Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A STENT GRAFT PLACEMENT PROCEDURE IN THE LEFT FOREARM, A FOREIGN BODY WAS ALLEGEDLY FOUND UNDER X-RAY EXAMINATION. REPORTEDLY, THE PATIENT HAD TO HAVE THE VEIN REMOVED. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 18490671 ↗
  • MalfunctionOct 14, 2019Break; Positioning Failure; Detachment of Device or Device Component; Break; Positioning Failure; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES ONE MALFUNCTION. A REVIEW OF THE REPORTED INFORMATION INDICATED THAT MODEL FEM12100, ENDOVASCULAR STENT GRAFT, ALLEGEDLY FAILED TO DEPLOY, EXPERIENCED A BREAK, DETACHMENT, AND A DEVICE-DEVICE INCOMPATIBILITY. THIS INFORMATION WAS RECEIVED FROM ONE SOURCE. THIS MALFUNCTION INVOLVED A PATIENT WITH NO CONSEQUENCES. THE PATIENT'S WEIGHT, AGE, AND GENDER WERE NOT PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9188965 ↗
  • MalfunctionMay 11, 2017Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A STENTING PROCEDURE FOR TREATMENT OF A LEFT ARM FISTULA, RESISTANCE WAS MET AND THE ENDOVASCULAR STENT GRAFT COULD NOT BE DEPLOYED. THERE WAS NO REPORTED DIFFICULTY IN RETRACTING THE DELIVERY SYSTEM FROM THE PATIENT. ANOTHER STENT WAS USED TO CONTINUE THE TREATMENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6559622 ↗

Other System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices

All System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page