Skip to content

neon³ CT spinal stabilization CS 3901-01Recalls & reported problems

Ulrich GmbH & Co. KG · Locking screw · Class 2 · Implant

All models of neon³ CT spinal stabilization (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04052536084159), received between Mar 29, 2019 and Jun 28, 2019 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K150650, K161032); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1
  • Fracture1
  • Material Fragmentation1

Patient problems

  • Failure of Implant2
  • Device Embedded In Tissue or Plaque1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2019Break; Break

    PIECE OF SCREW HEAD BROKE OFF INSIDE LOCKING SCREW. REPORT 2/2.

    Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient

    MAUDE report 8745542 ↗
  • InjuryMar 29, 2019Fracture; Material Fragmentation; Fracture; Material Fragmentation

    DURING THE CASE, THE SCREW'S TULIP WAS SHEERED OFF DURING FINAL TIGHTENING. THE REP BELIEVES THIS MAY HAVE BEEN CAUSED BY CROSS THREADING THE LOCKING SCREW. THE SCREW WAS REMOVED AND REPLACED AND THE CASE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Device Embedded In Tissue or Plaque; Failure of Implant

    MAUDE report 8465475 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150650, K161032. One submission often covers many models, so check the product and lot details in each record.

  • Z-1350-2020 · Dec 31, 2019 · Terminated
    Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
  • Z-1126-2019 · Mar 8, 2019 · Terminated
    ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page