therascreen® KRAS RGQ PCR Kit 1Recalls & reported problems
QIAGEN GmbH · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Aug 22, 2023Class IIOpen, Classified
Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid
Z-2665-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 25, 2022Class IIOpen, Classified
False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.
Z-0951-2022 · Root cause: Device Design · Recall details · FDA record ↗
Other Somatic Gene Mutation Detection System devices
- cobas® EGFR Mutation Test v2 2.0Roche Molecular Systems, Inc. · Somatic gene mutation detection system129 reports
- RealTime 08N90-090Abbott Molecular Inc. · Somatic gene mutation detection system2 reports
- therascreen® EGFR RGQ PCR Kit 1QIAGEN GmbH · Somatic Gene Mutation Detection System2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page