Skip to content

therascreen® KRAS RGQ PCR Kit 1Recalls & reported problems

QIAGEN GmbH · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 22, 2023Class IIOpen, Classified

    Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid

    Z-2665-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 25, 2022Class IIOpen, Classified

    False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.

    Z-0951-2022 · Root cause: Device Design · Recall details · FDA record ↗

Other Somatic Gene Mutation Detection System devices

All Somatic Gene Mutation Detection System devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page