therascreen KRAS RGQ PCR Kit (24) recallZ-0951-2022 · Qiagen Sciences LLC · initiated Feb 25, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Qiagen Sciences LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
False positive or false negative G12C (12CYS) mutation result for the KRAS Mutation Status resulting from Flags/Warnings may invalidate the result for one or more individual mutation target of the kit.
FDA root cause category: Device Design
What the company is doing
Qiagen issued Urgent Medical Device Correction therascreen KRAS RGQ PCR Kit (24), REF 870021 letter on 2/25/22. Letter states reason for recall, health risk and action to take: This notice is only relevant for the analysis of samples originating from non-small cell lung cancer (NSCLC) tissue. For continued use of the therascreen KRAS RGQ PCR Kit (24), check the Flags/Warnings in the Sample Result table for automated interpretation of results. Sample results with one or more of the Flags/Warnings listed in this notice should be regarded as invalid with respect to the KRAS status and KRAS Mutation status. Setup a new PCR run to repeat the affected sample(s). To minimize Flags/Warnings for the controls and samples, strict compliance to the IFU guidance should be followed, including but not limited to: Correct reagent mixing is required, and must be ensured at each mixing step during the assay setup. Placement of the Rotor-Gene Q MDx according to Installation Procedures site requirements Forward this information to all individuals and departments within your organization who are using the therascreen KRAS RGQ PCR Kit (24), REF. 870021. If you are not the end user, please forward this notice to the product end user. Review this notice with your laboratory/medical director. Complete the Acknowledgement of Receipt Form attached to this letter by March 10, 2022 and email it to [email protected]. QIAGEN is revising the Instructions for Use, as described in this notice, to reduce any risk resulting from the calling of a false positive or false negative G12C (12CYS) mutation result for NSCLC samples due to the Flags/Warnings listed in this notice and to emphasize strict compliance to the IFU guidance to minimize Flags/Warning. If you have any questions or concerns, pleas.se contact your local QIAGEN Technical Service Department through any of the following: Telephone: 800 362 7737 Email: [email protected]
Recall details
- Recall number
- Z-0951-2022
- Status
- Open, Classified
- Initiated by the company
- Feb 25, 2022
- Recalling firm
- Qiagen Sciences LLC
- Quantity
- 435 kits
- Product code
- OWD
- FDA submissions
- P110027
- Distribution
- AZ CA CT FL GA IL IN MA MN NC NY OK OR PA TN TX WV
Product description
therascreen KRAS RGQ PCR Kit (24) Model Number(s): REF 870021 (US IVD)
Code information (models, lots, serial numbers)
All lot numbers. GTIN number: 04053228002048
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- therascreen® KRAS RGQ PCR Kit 1QIAGEN GmbH · Somatic Gene Mutation Detection System2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page