PANOVIEW 8650415Recalls & reported problems
Richard Wolf GmbH · TELESCOPE 70° Ø 4MM WL 302MM rigid, TL 364mm, rod lens system · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04055207018354), received between Sep 13, 2019 and Sep 14, 2019 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Failure to Clean Adequately3
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2019Failure to Clean Adequately; Failure to Clean Adequately
ON JULY 11, 2019, A CUSTOMER REQUESTED FIVE (5) SCOPES BE RETURNED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) FOR REPAIR. THE CUSTOMER REPORTED THAT A SUBSTANCE WAS BEING COOKED AT THE END AND REQUESTED TO KNOW WHAT THE SUBSTANCE WAS. THE DEVICES WERE RETURNED TO RWMIC ON JULY 16, 2019. UPON RETURN, PHOTOGRAPHS WERE TAKEN OF THE SCOPES. PHOTOGRAPHS SHOWED WAS APPEARED TO BE BLOOD AND TISSUE COOKED ON 4 OUT OF 5 OF THE SCOPE'S DISTAL TIPS. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED FOR TELESCOPE PART # 8650.415, SERIAL # (B)(4). AS THE DEVICE HAD ALREADY BEEN CLEANED PRIOR TO LEAVING THE RECEIVING DEPARTMENT, NO RESIDUE COULD BE DETECTED. THE HOUSING WAS FOUND TO BE BENT AND DENTED, THE "PROXIMAL" WAS LEAKING, AND THE OBJECTIVE WAS DAMAGED FROM MOISTURE. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USER ERROR/MISUSE AND/OR USER HANDLING; THE DEVICE MAY HAVE BEEN EXPOSED TO SOME KIND OF PHYSICAL FORCE. ON JULY 14, 2019, THE CUSTOMER PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: WHAT KIND OF PROCEDURES WERE BEING PERFORMED WITH THESE DEVICES? - CYSTOSCOPE CASES DO YOU HAVE ANY EVIDENCE OF A FUNCTIONAL TEST PRIOR TO USING THESE INSTRUMENTS? (SEE IFU GA-S001 SECTION 8.2)
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9056752 ↗MalfunctionSep 13, 2019Failure to Clean Adequately; Failure to Clean Adequately
ON JULY 11, 2019, A CUSTOMER REQUESTED FIVE (5) SCOPES BE RETURNED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) FOR REPAIR. THE CUSTOMER REPORTED THAT A SUBSTANCE WAS BEING COOKED AT THE END, AND REQUESTED TO KNOW WHAT THE SUBSTANCE WAS. THE DEVICES WERE RETURNED TO RWMIC ON JULY 16, 2019. UPON RETURN, PHOTOGRAPHS WERE TAKEN OF THE SCOPES. PHOTOGRAPHS SHOWED WAS APPEARED TO BE BLOOD AND TISSUE COOKED ON 4 OUT OF 5 OF THE SCOPE'S DISTAL TIPS. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED FOR TELESCOPE PART # 8650.415, SERIAL # (B)(4). AS THE DEVICE HAD ALREADY BEEN CLEANED PRIOR TO LEAVING THE RECEIVING DEPARTMENT, NO RESIDUE COULD BE DETECTED. THE SCOPE PRESENTED DAMAGE THAT MAY HAVE BEEN DUE TO A THIRD PARTY REPAIR, CUSTOMER CONFIRMED THE DEVICE HAD BEEN REPAIRED BY A THIRD PARTY. ON JULY 14, 2019, THE CUSTOMER PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: WHAT KIND OF PROCEDURES WERE BEING PERFORMED WITH THESE DEVICES? CYSTOSCOPE CASES. DO YOU HAVE ANY EVIDENCE OF A FUNCTIONAL TEST PRIOR TO USING THESE INSTRUMENTS? (SEE IFU GA-S001 SECTION 8.2) ONLY FUNCTION BEFORE PLACING IN PAN TO BE STERILIZED, SPD PERSONNEL LOOK THROUGH THE SCOPE, SURGICAL TECHS IN THE OR DO NO
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9050571 ↗MalfunctionSep 13, 2019Failure to Clean Adequately; Failure to Clean Adequately
ON JULY 11, 2019, A CUSTOMER REQUESTED FIVE (5) SCOPES BE RETURNED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) FOR REPAIR. THE CUSTOMER REPORTED THAT A SUBSTANCE WAS BEING COOKED AT THE END AND REQUESTED TO KNOW WHAT THE SUBSTANCE WAS. THE DEVICES WERE RETURNED TO RWMIC ON JULY 16, 2019. UPON RETURN, PHOTOGRAPHS WERE TAKEN OF THE SCOPES. PHOTOGRAPHS SHOWED WAS APPEARED TO BE BLOOD AND TISSUE COOKED ON 4 OUT OF 5 OF THE SCOPE'S DISTAL TIPS. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED FOR TELESCOPE PART # 8650.415, SERIAL # (B)(4). AS THE DEVICE HAD ALREADY BEEN CLEANED PRIOR TO LEAVING THE RECEIVING DEPARTMENT, NO RESIDUE COULD BE DETECTED. THE HOUSING WAS FOUND TO BE BENT AND DENTED AND ONE (1) OF THE LENS WAS FOUND TO BE BROKEN. ADDITIONALLY, THE EYEPIECE AND COLOR RING WERE THIRD PARTY PARTS, CUSTOMER CONFIRMED THE DEVICE HAD BEEN REPAIRED BY A THIRD PARTY. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USER ERROR/MISUSE AND/OR USER HANDLING; THE DEVICE MAY HAVE BEEN EXPOSED TO SOME KIND OF PHYSICAL FORCE AND/OR THE DAMAGE MAY HAVE DUE TO THE THIRD PARTY REPAIR. ON JULY 14, 2019, THE CUSTOMER PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: WHAT KIND OF PROCEDURES WERE B
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9050517 ↗
Other ENDOSCOPE, RIGID devices
- Olympus WA50082AOLYMPUS Winter & Ibe GmbH · Endoscope, Rigid1,056 reports
- Olympus WA50080AOLYMPUS Winter & Ibe GmbH · Endoscope, Rigid245 reports
- PANOVIEW 8654422Richard Wolf GmbH · ENDOSCOPE, RIGID5 reports
- PANOVIEW 8654431Richard Wolf GmbH · ENDOSCOPE, RIGID1 report
- NA 8834075Richard Wolf GmbH · ENDOSCOPE, RIGID1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.