PANOVIEW 8654422Recalls & reported problems
Richard Wolf GmbH · TELESCOPE 30° Ø 4MM WL 300MM rigid, TL 364mm, rod lens system · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04055207018798), received between Sep 13, 2019 and Aug 19, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Failure to Clean Adequately” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately2
- Device Handling Problem1
- Insufficient Device Problem Information1
- Material Opacification1
- Positioning Failure1
- Use of Device Problem1
Patient problems
- No Known Impact Or Consequence To Patient2
- Air Embolism1
- Brain Injury1
- Cardiac Arrest1
- Electrolyte Imbalance1
- Hyperventilation1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 19, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING A PROSTATE RESECTION PROCEDURE USING A BIPOLAR RESECTOSCOPE AND AN ERBE VIO 300D GENERATOR, THE SURGEON OBSERVED A STRONG ELECTROLYTIC REACTION AT THE BEGINNING OF THE PROCEDURE. THE PATIENT SUFFERED SUDDEN HYPOCAPNIA, SEVERE GAS EMBOLISM, AND THREE CARDIAC ARRESTS. THE PATIENT IS CURRENTLY IN INTENSIVE CARE WITH SEVERE BRAIN DAMAGE.
Patient problems: Air Embolism; Brain Injury; Cardiac Arrest; Electrolyte Imbalance; Hyperventilation
MAUDE report 22839667 ↗MalfunctionOct 27, 2021Use of Device Problem; Device Handling Problem; Positioning Failure; Use of Device Problem; Device Handling Problem
PLEASE SEE MANUFACTURERS NARRATIVE FOR DEVICE EVALUATION/INVESTIGATION RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12705292 ↗MalfunctionOct 1, 2021Material Opacification; Material Opacification
THE PURPOSE OF THIS SUBMISSION IS TO REPORT THE RESULTS OF THE DEVICE EVALUATION. PLEASE SEE MANUFACTURERS NARRATIVE.
Patient problems: Insufficient Information
MAUDE report 12562567 ↗MalfunctionSep 14, 2019Failure to Clean Adequately; Failure to Clean Adequately
ON JULY 11, 2019, A CUSTOMER REQUESTED FIVE (5) SCOPES BE RETURNED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) FOR REPAIR. THE CUSTOMER REPORTED THAT A SUBSTANCE WAS BEING COOKED AT THE END AND REQUESTED TO KNOW WHAT THE SUBSTANCE WAS. THE DEVICES WERE RETURNED TO RWMIC ON JULY 16, 2019. UPON RETURN, PHOTOGRAPHS WERE TAKEN OF THE SCOPES. PHOTOGRAPHS SHOWED WAS APPEARED TO BE BLOOD AND TISSUE COOKED ON 4 OUT OF 5 OF THE SCOPE'S DISTAL TIPS. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED FOR TELESCOPE PART # 8650.415, SERIAL # (B)(4). AS THE DEVICE HAD ALREADY BEEN CLEANED PRIOR TO LEAVING THE RECEIVING DEPARTMENT, NO RESIDUE COULD BE DETECTED. THE HOUSING WAS FOUND TO BE BENT AND DENTED AND THE NEGATIVE WAS CHIPPED. ADDITIONALLY, THE EYEPIECE WAS NOTED TO BE A THIRD PARTY PART, CUSTOMER CONFIRMED THE DEVICE HAD BEEN REPAIRED BY A THIRD PARTY. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USER ERROR/MISUSE AND/OR USER HANDLING; THE DEVICE MAY HAVE BEEN EXPOSED TO SOME KIND OF PHYSICAL FORCE AND/OR THE DAMAGE MAY HAVE DUE TO THE THIRD PARTY REPAIR. ON JULY 14, 2019, THE CUSTOMER PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: WHAT KIND OF PROCEDURES WERE BEING PERFORMED WITH T
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9056746 ↗MalfunctionSep 13, 2019Failure to Clean Adequately; Failure to Clean Adequately
ON JULY 11, 2019, A CUSTOMER REQUESTED FIVE (5) SCOPES BE RETURNED TO RICHARD WOLF MEDICAL INSTRUMENTS CORPORATION (RWMIC) FOR REPAIR. THE CUSTOMER REPORTED THAT A SUBSTANCE WAS BEING COOKED AT THE END AND REQUESTED TO KNOW WHAT THE SUBSTANCE WAS. THE DEVICES WERE RETURNED TO RWMIC ON JULY 16, 2019. UPON RETURN, PHOTOGRAPHS WERE TAKEN OF THE SCOPES. PHOTOGRAPHS SHOWED WAS APPEARED TO BE BLOOD AND TISSUE COOKED ON 4 OUT OF 5 OF THE SCOPE'S DISTAL TIPS. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED FOR TELESCOPE PART # 8654.422, SERIAL #: (B)(4). AS THE DEVICE HAD ALREADY BEEN CLEANED PRIOR TO LEAVING THE RECEIVING DEPARTMENT, NO RESIDUE COULD BE DETECTED. THE HOUSING WAS FOUND TO BE BENT AND DENTED, THE "PROXIMAL" WAS LEAKING, AND THE OBJECTIVE WAS CHIPPED. PROBABLE ROOT CAUSE WAS DETERMINED TO BE USER ERROR/MISUSE AND/OR USER HANDLING; THE DEVICE MAY HAVE BEEN EXPOSED TO SOME KIND OF PHYSICAL FORCE. ON JULY 14, 2019, THE CUSTOMER PROVIDED THE FOLLOWING ADDITIONAL INFORMATION: WHAT KIND OF PROCEDURES WERE BEING PERFORMED WITH THESE DEVICES? CYSTOSCOPE CASES. DO YOU HAVE ANY EVIDENCE OF A FUNCTIONAL TEST PRIOR TO USING THESE INSTRUMENTS? ONLY FUNCTION BEFORE PLACING IN PAN TO BE S
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9050423 ↗
Other ENDOSCOPE, RIGID devices
- Olympus WA50082AOLYMPUS Winter & Ibe GmbH · Endoscope, Rigid1,056 reports
- Olympus WA50080AOLYMPUS Winter & Ibe GmbH · Endoscope, Rigid245 reports
- PANOVIEW 8650415Richard Wolf GmbH · ENDOSCOPE, RIGID3 reports
- PANOVIEW 8654431Richard Wolf GmbH · ENDOSCOPE, RIGID1 report
- NA 8834075Richard Wolf GmbH · ENDOSCOPE, RIGID1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page