Skip to content

AcuNav G 10F Ultrasound Catheter 10043342Recalls & reported problems

Siemens Medical Solutions USA, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04056869001036), received between Feb 6, 2026 and Aug 14, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K071234); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • No Display/Image2

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261485 ↗
  • MalfunctionFeb 18, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE CALLER, CLINICAL ACCOUNT SPECIALIST THAT AFTER THE ACUSON ACUNAV¿ ULTRASOUND CATHETER WAS CONNECTED TO THE "GE" ULTRASOUND SYSTEM, THE CATHETER WAS NOT RECOGNIZED BY THE SYSTEM. THE SWIFLINK CONNECTER WAS RESEATED WITHOUT RESOLUTION. THE ULTRASOUND SYSTEM WAS RESTARTED WITHOUT RESOLUTION. THE CATHETER WAS REPLACED WITHOUT RESOLUTION. THE SWIFTLINK CONNECTOR WAS REPLACED, AND THE ISSUE PERSISTED. THE CALLER STATED THAT THE ULTRASOUND SYSTEM AND CATHETER WERE BOTH REPLACED WITH ANOTHER (VUE FLEX CATHETER) AND THE ISSUE RESOLVED. THE PROCEDURE CONTINUED WITH NO FURTHER ISSUES. THE CALLER IS REQUESTING TWO REPLACEMENT CATHETERS. THE CALLER STATED THAT THE ACUSON ACUNAV¿ ULTRASOUND CATHETERS ARE AVAILABLE FOR RETURN. RETURN KIT REQUESTED. THE CALLER NOTED THAT THE CARTO¿ 3 SYSTEM WAS NOT USED. THE CALLER NOTED THAT THE MEDTRONIC AFFERA WAS USED. DOCUMENTATION APPROVED BY CSS WS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24376638 ↗
  • MalfunctionFeb 18, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE S70 ULTRASOUND SYSTEM DISPLAYED A MESSAGE "PROBE NOT SUPPORTED" WHEN THE CATHETER WAS PLUGGED INTO THE S70 ULTRASOUND SYSTEM. CALLER RESEATED THE TRANSDUCER WITHOUT RESOLUTION. CALLER REBOOTED S70 ULTRASOUND SYSTEM WITHOUT RESOLUTION. THE CATHETER WAS REPLACED, AND THE ISSUE WAS NOT RESOLVED. CALLER SWITCHED TO THE MEDTRONIC ULTRASOUND MACHINE AND THE ISSUE WAS RESOLVED. THE PROCEDURE WAS CONTINUED. 2 REPLACEMENT CATHETER REQUESTED. SOFTWARE VERSION 8.1.1.944 MEDTRONIC GENERATOR WAS IN USE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 24376624 ↗
  • MalfunctionFeb 6, 2026No Display/Image

    DURING AN UNKNOWN PROCEDURE, NO IMAGE WAS DISPLAYED WHEN THE CATHETER WAS CONNECTED. SURGERY WAS DELAYED BY 10 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24284759 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071234. One submission often covers many models, so check the product and lot details in each record.

  • Z-1745-2011 · Mar 20, 2009 · Terminated
    Siemens Medical Solutions USA, Inc., ACUSON AcuNav ultrasound catheters when using SwiftLink catheter connectors; Model Number: 10348560.

Other Catheter, ultrasound, intravascular devices

All Catheter, ultrasound, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page