AcuNav G 10F Ultrasound Catheter 10043342Recalls & reported problems
Siemens Medical Solutions USA, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04056869001036), received between Feb 6, 2026 and Aug 14, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K071234); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available2
- No Display/Image2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 14, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50261485 ↗MalfunctionFeb 18, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE CALLER, CLINICAL ACCOUNT SPECIALIST THAT AFTER THE ACUSON ACUNAV¿ ULTRASOUND CATHETER WAS CONNECTED TO THE "GE" ULTRASOUND SYSTEM, THE CATHETER WAS NOT RECOGNIZED BY THE SYSTEM. THE SWIFLINK CONNECTER WAS RESEATED WITHOUT RESOLUTION. THE ULTRASOUND SYSTEM WAS RESTARTED WITHOUT RESOLUTION. THE CATHETER WAS REPLACED WITHOUT RESOLUTION. THE SWIFTLINK CONNECTOR WAS REPLACED, AND THE ISSUE PERSISTED. THE CALLER STATED THAT THE ULTRASOUND SYSTEM AND CATHETER WERE BOTH REPLACED WITH ANOTHER (VUE FLEX CATHETER) AND THE ISSUE RESOLVED. THE PROCEDURE CONTINUED WITH NO FURTHER ISSUES. THE CALLER IS REQUESTING TWO REPLACEMENT CATHETERS. THE CALLER STATED THAT THE ACUSON ACUNAV¿ ULTRASOUND CATHETERS ARE AVAILABLE FOR RETURN. RETURN KIT REQUESTED. THE CALLER NOTED THAT THE CARTO¿ 3 SYSTEM WAS NOT USED. THE CALLER NOTED THAT THE MEDTRONIC AFFERA WAS USED. DOCUMENTATION APPROVED BY CSS WS.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24376638 ↗MalfunctionFeb 18, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE S70 ULTRASOUND SYSTEM DISPLAYED A MESSAGE "PROBE NOT SUPPORTED" WHEN THE CATHETER WAS PLUGGED INTO THE S70 ULTRASOUND SYSTEM. CALLER RESEATED THE TRANSDUCER WITHOUT RESOLUTION. CALLER REBOOTED S70 ULTRASOUND SYSTEM WITHOUT RESOLUTION. THE CATHETER WAS REPLACED, AND THE ISSUE WAS NOT RESOLVED. CALLER SWITCHED TO THE MEDTRONIC ULTRASOUND MACHINE AND THE ISSUE WAS RESOLVED. THE PROCEDURE WAS CONTINUED. 2 REPLACEMENT CATHETER REQUESTED. SOFTWARE VERSION 8.1.1.944 MEDTRONIC GENERATOR WAS IN USE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24376624 ↗MalfunctionFeb 6, 2026No Display/Image
DURING AN UNKNOWN PROCEDURE, NO IMAGE WAS DISPLAYED WHEN THE CATHETER WAS CONNECTED. SURGERY WAS DELAYED BY 10 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24284759 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071234. One submission often covers many models, so check the product and lot details in each record.
- Z-1745-2011 · Mar 20, 2009 · TerminatedSiemens Medical Solutions USA, Inc., ACUSON AcuNav ultrasound catheters when using SwiftLink catheter connectors; Model Number: 10348560.
Other Catheter, ultrasound, intravascular devices
- OptiCross H749518100Boston Scientific Corporation · Catheter, ultrasound, intravascular2 deaths457 reports
- OptiCross™ HD H74939352030Boston Scientific Corporation · Catheter, ultrasound, intravascular1 death436 reports
- ViewFlex™ D087031Irvine Biomedical, Inc. · Catheter, ultrasound, intravascular3 deaths341 reports
- OptiCross 18 H7493932800180Boston Scientific Corporation · Catheter, ultrasound, intravascular186 reports
- OptiCross™ 6 HD H74939354080Boston Scientific Corporation · Catheter, ultrasound, intravascular4 deaths177 reports
- Visions PV .014P RX Digital IVUS Catheter 014RPhilips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall1 death84 reports
- Visions PV .018 Digital IVUS Catheter 86700Philips Image Guided Therapy Corporation · Catheter, Ultrasound, Intravascular1 recall67 reports
- OptiCross™ HD H74939352040Boston Scientific Corporation · Catheter, ultrasound, intravascular3 deaths116 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page