Skip to content

Biograph mMR 10433372Recalls & reported problems

Siemens Healthcare GmbH · MRI-PET Scanner · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K133226, K163234, K200213); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Aug 28, 2025Class IOpen, Classified

    There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    Z-2654-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K133226, K163234, K200213. One submission often covers many models, so check the product and lot details in each record.

  • Z-0467-2016 · Dec 7, 2015 · Terminated
    MAGNETOM systems, magnetic resonance diagnostic devices (MRDD).

Other Tomographic imager combining emission computed tomography with nuclear magnetic resonance devices

All Tomographic imager combining emission computed tomography with nuclear magnetic resonance devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page