NA SIGNA PET MR (UDI 00840682105378)Recalls & reported problems

GE Medical Systems, L.L.C. · PET MR System Medical Device Assembly · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
11

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682105378), received between Apr 15, 2026 and Apr 15, 2026 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

11 FDA recall records name this device's identifier, 2 still open.

6 other recalls cover devices cleared under the same FDA submission (K142098, K163619); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 20, 2025Open, Classified

    GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.

    Z-2309-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗

  2. May 23, 2025Open, Classified

    GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

    Z-2133-2025 · Root cause: Device Design · Recall details · FDA record ↗

  3. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3029-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  4. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3021-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  5. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3023-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  6. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3025-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  7. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3028-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  8. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3027-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  9. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3024-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  10. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3022-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

  11. May 11, 2017Terminated

    Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

    Z-3026-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Use of Device Problem1

Patient problems

  • Limb Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 15, 2026Use of Device Problem; Insufficient Device Problem Information

    AFTER COMPLETION OF AN EXAM, A TECHNOLOGIST TRIPPED OVER TUBING FROM THE PATIENT ALERT BALL THAT HAD EXTENDED ONTO THE SCAN ROOM FLOOR WHILE THE PATIENT WAS REMOVED FROM THE BORE OF THE SCANNER. THE TECHNOLOGIST FELL, LANDING ON THE RIGHT KNEE, STRIKING THE NOSE, AND INJURING THE LEFT SHOULDER WHILE ATTEMPTING TO BRACE THE FALL. THE TECHNOLOGIST SOUGHT MEDICAL EVALUATION AT AN URGENT CARE FACILITY THE FOLLOWING DAY AND WAS DIAGNOSED WITH A LEFT CHEST CONTUSION AND LEFT SHOULDER STRAIN, WITH SUBSEQUENT INITIATION OF PHYSICAL THERAPY. NO PATIENT INJURY OCCURRED.

    Patient problems: Limb Pain

    MAUDE report 24884715 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142098, K163619. One submission often covers many models, so check the product and lot details in each record.

  • Z-0520-2025 · Oct 28, 2024 · Open, Classified
    Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
  • Z-2466-2024 · Jun 14, 2024 · Open, Classified
    SIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
  • Z-0595-2022 · Dec 23, 2021 · Open, Classified
    GE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA PET/MR System, tomographic imager combining emission computed tomography with nuclear magnetic resonanc
  • Z-2057-2021 · Jun 4, 2021 · Open, Classified
    SIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance
  • Z-0131-2021 · Sep 8, 2020 · Terminated
    Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance - Product Usage: The combined system maintains independent functionality of the MR and PET
  • Z-0260-2016 · Aug 31, 2015 · Terminated
    GE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners tha

Other Tomographic imager combining emission computed tomography with nuclear magnetic resonance devices

All Tomographic imager combining emission computed tomography with nuclear magnetic resonance devices →