NA SIGNA PET MR (UDI 00840682105378)Recalls & reported problems
GE Medical Systems, L.L.C. · PET MR System Medical Device Assembly · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 11
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00840682105378), received between Apr 15, 2026 and Apr 15, 2026 — classified as 1 injury.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
11 FDA recall records name this device's identifier, 2 still open.
6 other recalls cover devices cleared under the same FDA submission (K142098, K163619); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jun 20, 2025Open, Classified
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
Z-2309-2025 · Root cause: Equipment maintenance · Recall details · FDA record ↗
- May 23, 2025Open, Classified
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Z-2133-2025 · Root cause: Device Design · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3029-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3021-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3023-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3025-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3028-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3027-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3024-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3022-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
- May 11, 2017Terminated
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
Z-3026-2017 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information1
- Use of Device Problem1
Patient problems
- Limb Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 15, 2026Use of Device Problem; Insufficient Device Problem Information
AFTER COMPLETION OF AN EXAM, A TECHNOLOGIST TRIPPED OVER TUBING FROM THE PATIENT ALERT BALL THAT HAD EXTENDED ONTO THE SCAN ROOM FLOOR WHILE THE PATIENT WAS REMOVED FROM THE BORE OF THE SCANNER. THE TECHNOLOGIST FELL, LANDING ON THE RIGHT KNEE, STRIKING THE NOSE, AND INJURING THE LEFT SHOULDER WHILE ATTEMPTING TO BRACE THE FALL. THE TECHNOLOGIST SOUGHT MEDICAL EVALUATION AT AN URGENT CARE FACILITY THE FOLLOWING DAY AND WAS DIAGNOSED WITH A LEFT CHEST CONTUSION AND LEFT SHOULDER STRAIN, WITH SUBSEQUENT INITIATION OF PHYSICAL THERAPY. NO PATIENT INJURY OCCURRED.
Patient problems: Limb Pain
MAUDE report 24884715 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K142098, K163619. One submission often covers many models, so check the product and lot details in each record.
- Z-0520-2025 · Oct 28, 2024 · Open, ClassifiedTomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2) DISCOVERY PET/MR,
- Z-2466-2024 · Jun 14, 2024 · Open, ClassifiedSIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance
- Z-0595-2022 · Dec 23, 2021 · Open, ClassifiedGE Healthcare MR superconducting magnets, a component of GE Healthcare SIGNA PET/MR System, tomographic imager combining emission computed tomography with nuclear magnetic resonanc
- Z-2057-2021 · Jun 4, 2021 · Open, ClassifiedSIGNA PET/MR, Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance
- Z-0131-2021 · Sep 8, 2020 · TerminatedTomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance - Product Usage: The combined system maintains independent functionality of the MR and PET
- Z-0260-2016 · Aug 31, 2015 · TerminatedGE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners tha
Other Tomographic imager combining emission computed tomography with nuclear magnetic resonance devices
- Not Applicable 5494440GE Medical Systems, L.L.C. · Tomographic imager combining emission computed tomography with nuclear magnetic resonance9 recalls0 reports
- NA 5789577GE Medical Systems, L.L.C. · Tomographic imager combining emission computed tomography with nuclear magnetic resonance1 recall0 reports
- Biograph mMR 10433372Siemens Healthcare GmbH · Tomographic imager combining emission computed tomography with nuclear magnetic resonance1 recall0 reports
- NA SIGNA PET MRGE Medical Systems, L.L.C. · Tomographic imager combining emission computed tomography with nuclear magnetic resonance1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page