MAGNETOM AlteaRecalls & reported problems across all models
Siemens Healthcare GmbH · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one MAGNETOM Altea model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 1
We found 2 models of MAGNETOM Altea by Siemens Healthcare GmbH in the FDA's device registry (GUDID). Together they are named in 10 adverse event reports in the FDA's MAUDE database, received between Apr 28, 2022 and Sep 8, 2026 — classified as 9 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
MAGNETOM Altea recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 1, 2023Open, Classified
In some cases, when lubricating grease is present within the headband of the over-ear headphones, it may cause visible, dot- or lineshaped, fat-isointense artifacts during head examinations only. These potential artifacts depend on the position of the headband, the amount of lubricating grease, as well as the sequence parameters used
Z-2648-2023 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
Most reported MAGNETOM Altea problems
Device problems
- Insufficient Device Problem Information4
- Improper or Incorrect Procedure or Method2
- Use of Device Problem2
- Accessory Incompatible1
- Adverse Event Without Identified Device or Use Problem1
- Defective Component1
- Patient Device Interaction Problem1
Patient problems
- Burn(s)4
- Full thickness (Third Degree) Burn2
- Crushing Injury1
- Hearing Impairment1
- No Clinical Signs, Symptoms or Conditions1
- Partial thickness (Second Degree) Burn1
- Tinnitus1
MAGNETOM Altea models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- MAGNETOM Altea 11344915System, nuclear magnetic resonance imaging1 recall8 reports
- MAGNETOM Altea 11410371Coil, magnetic resonance, specialty1 recall2 reports
Recent MAGNETOM Altea adverse event reports
The 10 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 8, 2026Improper or Incorrect Procedure or Method
Siemens Healthineers (Siemens) was notified of a potential patient injury associated with a MAGNETOM Altea MRI system. According to the report, the patient complained of hearing issues following an MRI examination on (b)(6)2026. The patient wore only headphones during the scan and did not use the combination of headphones and earplugs specified for hearing protection. It was confirmed that the hearing protection warning signage was posted on the examination room door. The patient was evaluated by an ENT specialist on (b)(6)2026. Based on additional information received on September 2, 2026, Siemens was informed by the customer that the patient was diagnosed with Stage 2 temporomandibular joint (TMJ) disorder. The customer indicated that no additional information is available from the patient's ENT specialist or directly from the patient. The customer also confirmed that the patient did not use the hearing protection prescribed in the system Instructions for Use and wore only the standard Siemens headphones during the examination. As sufficient information is not available to rule out the occurrence of a serious injury, this event is being reported.
Patient problems: Hearing Impairment
MAUDE report 50666638 ↗InjuryAug 18, 2026Insufficient Device Problem Information
Siemens Healthineers (Siemens) was notified of a patient injury associated with a MAGNETOM Altea system. The reported event indicated that the patient sustained a radiofrequency burn. The patient was evaluated at an urgent care facility, where antibiotics were prescribed and burn ointment was provided.Additional information about the severity of the burn has been requested, but has not been received to this date.Therefore, in-doubt reporting is performed.
Patient problems: Burn(s)
MAUDE report 50299373 ↗InjuryNov 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
SIEMENS HEALTHINEERS WAS NOTIFIED OF A PATIENT INJURY INVOLVING THE MAGNETOM ALTEA SYSTEM. IT WAS STATED IN THE INITIAL COMPLAINT THAT A PATIENT OBTAINED A BURN DURING AN MR EXAMINATION OF THE HIP. ACCORDING TO THE INFORMATION PROVIDED IN THE QUESTIONNAIRE, THERE WAS AN AREA (10X10 CM IN SIZE) OF SKIN REDNESS AND BURNING SENSATION ON BOTH INNER SIDES OF THE PATIENT¿S LEGS. A CREAM FOR BURNS WAS PRESCRIBED. SIEMENS HEALTHINEERS WAS INFORMED ON (B)(6) 2025 THAT A CRUST HAS FORMED OVER THE BURNED AREA AND THAT IT IS STILL NOT HEALED COMPLETELY. THE SIEMENS HEALTHINEERS MEDICAL OFFICER ASSESSED THE AVAILABLE INFORMATION. THE FORMATION OF A CRUST AND A HEALING TIME OF MORE THAN 12 DAYS SUGGEST THAT THE PATIENT OBTAINED A BURN OF GRADE 2B OR HIGHER.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 23451105 ↗InjuryNov 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
SIEMENS HEALTHINEERS WAS NOTIFIED OF A POTENTIAL PATIENT INJURY INVOLVING THE MAGNETOM ALTEA SYSTEM. PER THE INFORMATION RECEIVED FROM THE CUSTOMER IT WAS STATED THAT THE PATIENT WAS PLACED FEET FIRST INTO THE MRI. AFTER THE EXAMINATION, WHEN THE PATIENT WAS REMOVED FROM THE BORE, IT WAS NOTED THAT THE PATIENT HAD BLISTERS ON THE INSIDE OF BOTH THIGHS. MEDICAL TREATMENT WAS PROVIDED, HOWEVER, THE EXTENT OF THE BURNS AND TYPE OF TREATMENT PROVIDED WERE NOT REPORTED TO SIEMENS HEALTHINEERS.
Patient problems: Burn(s)
MAUDE report 23450720 ↗InjurySep 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
SIEMENS HEALTHINEERS WAS NOTIFIED OF A PATIENT INJURY INVOLVING THE MAGNETOM ALTEA SYSTEM. PER THE RECEIVED INFORMATION, THE PATIENT RECEIVED A 25MM X 25MM BURN ON THE RIGHT SHOULDER DURING AN MRI EXAMINATION USING A THIRD-PARTY ELECTRODE HEART MONITORING DEVICE. SIEMENS HEALTHINEERS HAS REQUESTED ADDITIONAL INFORMATION REGARDING THE EVENT AND THE EXTENT OF THE PATIENT¿S INJURY. THE ADDITIONAL INFORMATION IS PENDING RECEIPT AS OF THE DATE OF THIS REPORT.
Patient problems: Burn(s)
MAUDE report 22976145 ↗MalfunctionJan 15, 2025Defective Component; Defective Component
IT WAS REPORTED TO SIEMENS THAT AN ISSUE WAS DISCOVERED ON THE MAGNETOM ALTEA (CN). IT WAS STATED THAT THE CUSTOMER FOUND THE LOWER PART OF THE BM HEAD/NECK 20 COIL SURFACE HAD BEEN BURNED WITH A HOLE AND THE DIAMETER WAS ABOUT 10 MM. THERE WAS NO INJURY ASSOCIATED WITH THIS ISSUE. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO INVESTIGATE THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21156537 ↗InjurySep 4, 2024Accessory Incompatible; Use of Device Problem; Accessory Incompatible; Use of Device Problem
SIEMENS BECAME AWARE OF AN INCIDENT THAT INVOLVED THE MAGNETOM ALTEA SYSTEM. AN OXYGEN TANK WAS BROUGHT INSIDE THE MRI EXAMINATION ROOM BY HOSPITAL PERSONNEL. THE TANK WAS PULLED BY THE MRI MACHINE; A DOCTOR¿S LEG WAS PINNED TO THE MACHINE BY AN OXYGEN TANK. THE MRI UNIT QUENCHED. THE DOCTOR REQUIRED A MEDICAL TREATMENT FOLLOWING THE INCIDENT AND NEEDED A SURGERY. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION REGARDING THIS EVENT.
Patient problems: Crushing Injury
MAUDE report 20137272 ↗InjuryAug 23, 2024Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem; Improper or Incorrect Procedure or Method
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE MAGNETOM ALTEA MR SYSTEM. IT WAS STATED THAT A PATIENT REPORTED A SKIN BURN SEVEN DAYS AFTER AN MR SCAN. THE PATIENT NOTED PAIN AND BLISTERS ON THE INDEX KNUCKLE AND LEFT FOREARM WHEN HE GOT IN THE CAR AFTERWARDS. LATER IT WAS DIAGNOSED THAT THE PATIENT SUFFERED A FULL THICKNESS BURN ON THE LEFT FOREARM AND A PARTIAL THICKNESS BURN ON THE RIGHT INDEX KNUCKLE. BOTH BURNS HAVE A SIZE LIKE A KNUCKLE. IT WAS STATED THAT THE POINTS OF BURN WERE TOUCHING EACH OTHER DURING EXAMINATION.
Patient problems: Burn(s); Full thickness (Third Degree) Burn
MAUDE report 20046046 ↗InjuryAug 3, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
SIEMENS HEALTHINEERS WAS INITIALLY INFORMED ON JULY 27, 2023, THAT A PATIENT SUSTAINED BURNS ON BOTH INNER THIGHS DURING AN MR PROSTATE EXAMINATION ON (B)(6) 2023. INITIALLY, IT WAS STATED THAT THE PATIENT SUSTAINED OVAL BURNS APPROXIMATELY 1.5CM CM LONG ON BOTH INNER THIGHS. THE SEVERITY OF THE BURNS WAS NOT INITIALLY REPORTED TO SIEMENS HEALTHINEERS. THE INITIAL COMPLAINT INFORMATION AND PHOTOS OF THE BURNS WERE ANALYZED BY A SIEMENS HEALTHINEERS MEDICAL OFFICER, WHO CONCLUDED THESE COULD BE THIRD-DEGREE BURNS. ADDITIONAL INFORMATION WAS PROVIDED TO SIEMENS HEALTHINEERS ON JULY 31, 2023, STATING THAT THE PATIENT HAD NOT COMPLAINED OF PAIN OR ALERTED THE MEDICAL TECHNICAL ASSISTANT PRESENT DURING THE EXAMINATION. IT WAS STATED THAT THE MEDICAL TECHNICAL ASSISTANT ROUTINELY WENT INTO THE SCANNING ROOM BEFORE ADMINISTERING THE CONTRAST AGENT AND NOTICED REDDENING OF THE PATIENT¿S SKIN. THE EXAMINATION WAS THEN STOPPED IMMEDIATELY. IT WAS FURTHER REPORTED THAT ALTHOUGH THE REDNESS ON THE INSIDE OF THE PATIENT¿S THIGHS HAS RECEDED, TWO DEEP ROUND THIRD-DEGREE BURNS REMAIN. EACH BURN IS APPROXIMATELY 2.5CM IN SIZE. ACCORDING TO THE DERMATOLOGICAL OUTPATIENT DEPARTMENT OF THE HOSPITAL,
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 17452218 ↗InjuryApr 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED FOLLOWING EXAMINATION ON THE MAGNETOM ALTEA MR SYSTEM. A PATIENT REPORTED LOUD NOISES DURING A SHOULDER EXAMINATION ON (B)(6) 2002. THE PATIENT ALREADY HAD TINNITUS WHICH WAS AGGRAVATED BY THE MR EXAMINATION. THE PATIENT WAS WEARING HEADPHONES DURING THE EXAMINATION AND IT WAS REPORTED THAT NO OTHER HEARING PROTECTION WAS PROVIDED. AT THIS TIME IT IS UNCLEAR IF THE PATIENTS TINNITUS IS TEMPORARY OR PERMANENT. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Tinnitus
MAUDE report 14237249 ↗
Other Siemens Healthcare GmbH products
- Artis zeego2 models · 31 reports · 6 recalls
- MAGNETOM Verio2 models · 2 reports · 4 recalls
- MAGNETOM Skyra2 models · 10 reports · 3 recalls
- MAGNETOM Cima.X2 models · 0 reports · 3 recalls
- MAGNETOM Vida2 models · 12 reports · 2 recalls
- MAGNETOM Lumina2 models · 3 reports · 2 recalls
- MAGNETOM Prisma2 models · 2 reports · 2 recalls
- MAGNETOM Sola2 models · 16 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page