K-Zero® Needleless ConnectorRecalls & reported problems
Fresenius Kabi AG · The K-Zero Neutral Needleless Connector is intended for single patient use in intravenous and blood administration and aspiration without the need for needles, thus eliminating the potential for needle-stick injuries during use. The K-Zero Neutral Needleless Connector may be used with low pressure power injectors having a maximum pressure of 325 psi and a maximum flow rate of 10 mL/sec. The subject device does ot have to be changed subsequent to use with a low pressure power injector. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 04250273735949), received between Dec 2, 2025 and Sep 4, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Apparent Adverse Event9
- Break3
- Leak/Splash3
- Blocked Connection2
- Device Contamination with Body Fluid1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Break; Leak/Splash; No Apparent Adverse Event
Cases of detachment/fracture between the transparent and coloured components of the check valve have been reported, resulting in blood leakage due to backflow or leakage of the intravenous therapy being administered. Furthermore, several samples were found to be occluded and unusable upon opening of the primary packaging.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50603297 ↗MalfunctionSep 4, 2026Break; Leak/Splash; No Apparent Adverse Event
Cases of detachment/fracture between the transparent and coloured components of the check valve have been reported, resulting in blood leakage due to backflow or leakage of the intravenous therapy being administered. Furthermore, several samples were found to be occluded and unusable upon opening of the primary packaging.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50603296 ↗MalfunctionSep 4, 2026Break; Leak/Splash; No Apparent Adverse Event
Cases of detachment/fracture between the transparent and coloured components of the check valve have been reported, resulting in blood leakage due to backflow or leakage of the intravenous therapy being administered. Furthermore, several samples were found to be occluded and unusable upon opening of the primary packaging.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50603291 ↗MalfunctionMar 20, 2026Blocked Connection
THE FOLLOWING HAS BEEN REPORTED: IT WAS OBSERVED THAT THE DEVICE SHOWS RESISTANCE DURING FLUSHING AND/OR DURING THE COLLECTION OF SAMPLES FOR LABORATORY TESTS (BLOOD). DUE TO THIS NONCONFORMITY, IT HAS BEEN NECESSARY TO REMOVE THE DEVICE IN ORDER TO PERFORM BOTH THE FLUSHING AND THE SAMPLE COLLECTION. THIS SITUATION SIGNIFICANTLY INCREASES THE RISK OF BLOODSTREAM INFECTION DUE TO THE ADDITIONAL MANIPULATION OF THE ACCESS, THEREBY COMPROMISING PATIENT SAFETY. FRESENIUS KABI INVESTIGATION RESULTS AS FOLLOWS: REPORTED COMPLAINT: FLUID INGRESS IN NEEDLE FREE PORT. QUANTITY OF THE SAMPLE(S): NO COMPLAINT SAMPLE OR PHOTO AVAILABLE FOR INVESTIGATION. CONCERNED PRODUCT CODE/BATCH: K-ZERO NEEDLELESS CONNECTOR - M79400849Y/32442575. INVESTIGATION REPORT: WE HAVE NOT RECEIVED A COMPLAINT SAMPLE OR PHOTO FOR INVESTIGATION. BASED ON THE REASON FOR THE COMPLAINT, WE PERFORMED A VISUAL INSPECTION OF ALL RETAIN SAMPLES AND DO NOT SEE ANY DEFECTS. THE SETS ARE COMPLETE AND FREE FROM CONTAMINATION. IN THE NEXT STEP, WE PERFORMED A FREE FLOW AND TIGHTNESS TEST ON ONE SAMPLE AND THE RESULTS WERE CORRECT. ADDITIONALLY, WE CONDUCTED A PRACTICAL TEST USING A SYRINGE AND WATER. WE DIDN'T SEE ANY WATER RES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651361 ↗MalfunctionMar 20, 2026Blocked Connection
THE FOLLOWING HAS BEEN REPORTED: IT WAS OBSERVED THAT THE DEVICE SHOWS RESISTANCE DURING FLUSHING AND/OR DURING THE COLLECTION OF SAMPLES FOR LABORATORY TESTS (BLOOD). DUE TO THIS NONCONFORMITY, IT HAS BEEN NECESSARY TO REMOVE THE DEVICE IN ORDER TO PERFORM BOTH THE FLUSHING AND THE SAMPLE COLLECTION. THIS SITUATION SIGNIFICANTLY INCREASES THE RISK OF BLOODSTREAM INFECTION DUE TO THE ADDITIONAL MANIPULATION OF THE ACCESS, THEREBY COMPROMISING PATIENT SAFETY. FRESENIUS KABI INVESTIGATION RESULTS AS FOLLOWS: 1. REPORTED COMPLAINT FLUID INGRESS IN NEEDLE FREE PORT 2. QUANTITY OF THE SAMPLE(S) NO COMPLAINT SAMPLE OR PHOTO AVAILABLE FOR INVESTIGATION 3. CONCERNED PRODUCT CODE/BATCH K-ZERO NEEDLELESS CONNECTOR - M79400849Y / (B)(6) 4. INVESTIGATION REPORT WE HAVE NOT RECEIVED A COMPLAINT SAMPLE OR PHOTO FOR INVESTIGATION. BASED ON THE REASON FOR THE COMPLAINT, WE PERFORMED A VISUAL INSPECTION OF ALL RETAIN SAMPLES AND DO NOT SEE ANY DEFECTS. THE SETS ARE COMPLETE AND FREE FROM CONTAMINATION. IN THE NEXT STEP, WE PERFORMED A FREE FLOW AND TIGHTNESS TEST ON ONE SAMPLE AND THE RESULTS WERE CORRECT. ADDITIONALLY, WE CONDUCTED A PRACTICAL TEST USING A SYRINGE AND WATER. WE DIDN'T SEE ANY WATE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651360 ↗MalfunctionMar 20, 2026Device Contamination with Body Fluid
THE FOLLOWING HAS BEEN REPORTED: DURING MANIPULATION OF THE KEY DEVICE DURING TEST COLLECTION OR MEDICATION ADMINISTRATION, THE NURSING TEAM OBSERVED DIFFICULTY IN WASHING THE DEVICE, WITH BLOOD ACCUMULATION EVEN AFTER NUMEROUS WASHES, CHANGES IN THE CALIBRE OF THE SYRINGES, AND THE WHIRLING METHOD. FRESENIUS KABI INVESTIGATION RESULTS AS FOLLOWS: 1. REPORTED COMPLAINT FLUID INGRESS IN NEEDLE FREE PORT 2. QUANTITY OF THE SAMPLE(S) NO COMPLAINT SAMPLE OR PHOTO AVAILABLE FOR INVESTIGATION 3. CONCERNED PRODUCT CODE/BATCH K-ZERO NEEDLELESS CONNECTOR - M79400849Y / (B)(6) 4. INVESTIGATION REPORT WE HAVE NOT RECEIVED A COMPLAINT SAMPLE OR PHOTO FOR INVESTIGATION. BASED ON THE REASON FOR THE COMPLAINT, WE PERFORMED A VISUAL INSPECTION OF ALL RETAIN SAMPLES AND DO NOT SEE ANY DEFECTS. THE SETS ARE COMPLETE AND FREE FROM CONTAMINATION. IN THE NEXT STEP, WE PERFORMED A FREE FLOW AND TIGHTNESS TEST ON ONE SAMPLE AND THE RESULTS WERE CORRECT. ADDITIONALLY, WE CONDUCTED A PRACTICAL TEST USING A SYRINGE AND WATER. WE DIDN'T SEE ANY WATER RESIDUE INSIDE THE K-ZERO. BATCH DOCUMENTATION REVIEW WAS PERFORMED AND FOLLOWING DATA WAS CHECKED: PRODUCTION ORDERS, IN PROCESS CONTROL AND FINAL INSPECTION R
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24651359 ↗MalfunctionDec 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: PER EMAIL FROM THE FRESENIUS KABI SALES DEVELOPMENT REPRESENTATIVE SOUTHEAST ON 11/7/2025: "I HAVE HAD A FEW COMPLAINTS COME FROM OUR CUSTOMER (B)(6) HOSPITAL, REGARDING THE K-ZERO NEEDLESS CONNECTORS "POPPING OFF" AND NOT SECURING ONTO NON-DEDICATED SETS. THIS HAPPENED DURING PATIENT CARE AND CONCERNED THE HOSPITAL, NURSES, PATIENT, ETC. PER PREVIOUS ORDER HISTORY FROM THIS CUSTOMER, THE K-ZERO CONNECTOR THEY ORDER IS THE (B)(4) CONNECTOR." NO ADVERSE EVENTS REPORTED NO SAMPLE AVAILABLE NO PICTURE AVAILABLE. 1. REPORTED COMPLAINT PROBLEMS WITH MEDICATION PORT 2. QUANTITY OF SAMPLE(S) WE DID NOT RECEIVE ANY SAMPLE OR PICTURE FOR INVESTIGATION 3. CONCERNED PRODUCT CODE / BATCH K-ZERO NEEDLELESS CONNECTOR, (M79400849), UNKNOWN 4. INVESTIGATION REPORT WE HAVE NOT RECEIVED A COMPLAINT SAMPLE FOR INVESTIGATION. AS WE DID NOT RECEIVE THE BATCH NUMBER, AN INSPECTION OF RETAINED SAMPLES AND REVIEW OF BATCH DOCUMENTATION WAS NOT POSSIBLE. WITHOUT FAULTY SAMPLE AND BATCH NUMBER, A MORE DETAILED INVESTIGATION WAS NOT POSSIBLE. IN THE LAST 12 MONTHS (FROM 20.11.2024 TO 19.11.2025) WE HAVE RECEIVED 2 COMPLAINTS FOR THIS ARTICLE WITH THE SAME REASON AS "PROBLEMS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23722354 ↗MalfunctionDec 4, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: PER EMAIL FROM THE FRESENIUS KABI SALES DEVELOPMENT REPRESENTATIVE SOUTHEAST ON 11/7/2025: "I HAVE HAD A FEW COMPLAINTS COME FROM OUR CUSTOMER (B)(6) HOSPITAL, REGARDING THE K-ZERO NEEDLESS CONNECTORS "POPPING OFF" AND NOT SECURING ONTO NON-DEDICATED SETS. THIS HAPPENED DURING PATIENT CARE AND CONCERNED THE HOSPITAL, NURSES, PATIENT, ETC. PER PREVIOUS ORDER HISTORY FROM THIS CUSTOMER, THE K-ZERO CONNECTOR THEY ORDER IS THE M79400849 CONNECTOR." NO ADVERSE EVENTS REPORTED. NO SAMPLE AVAILABLE. NO PICTURE AVAILABLE. 1. REPORTED COMPLAINT: PROBLEMS WITH MEDICATION PORT. 2. QUANTITY OF SAMPLE(S): WE DID NOT RECEIVE ANY SAMPLE OR PICTURE FOR INVESTIGATION. 3. CONCERNED PRODUCT CODE / BATCH: K-ZERO NEEDLELESS CONNECTOR, (M79400849), UNKNOWN. 4. INVESTIGATION REPORT: WE HAVE NOT RECEIVED A COMPLAINT SAMPLE FOR INVESTIGATION. AS WE DID NOT RECEIVE THE BATCH NUMBER, AN INSPECTION OF RETAINED SAMPLES AND REVIEW OF BATCH DOCUMENTATION WAS NOT POSSIBLE. WITHOUT FAULTY SAMPLE AND BATCH NUMBER, A MORE DETAILED INVESTIGATION WAS NOT POSSIBLE. IN THE LAST 12 MONTHS (FROM 20.11.2024 TO 19.11.2025) WE HAVE RECEIVED 2 COMPLAINTS FOR THIS ARTICLE WITH THE SAME REASON A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23722293 ↗MalfunctionDec 2, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: ON 11/3/2025 FENWAL INTERNATIONAL QE COMPLAINTS ENGINEER RECEIVED A [FRESENIUS KABI] IVENIX COMPLAINT/DISPOSABLE FOR EVALUATION AND DURING EVALUATION THEY FOUND A DEFECT WITH A BLONIE PRODUCT. NO ADVERSE EVENT REPORTED. NO SAMPLE AVAILABLE. PICTURES AVAILABLE. FINDING: DURING SAMPLE EVALUATION THE CAP WAS FOUND TO BE SEPARATED FROM Y-PORT. CRACKED CAP. NOTE: THIS WAS A K-ZERO CONNECTOR ON A FRESENIUS KABI IVENIX ADMINISTRATION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23700011 ↗MalfunctionDec 2, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: ON (B)(6) 2025 FENWAL INTERNATIONAL QE COMPLAINTS ENGINEER RECEIVED A [FRESENIUS KABI] IVENIX COMPLAINT/DISPOSABLE FOR EVALUATION AND DURING EVALUATION THEY FOUND A DEFECT WITH A BLONIE PRODUCT. NO ADVERSE EVENT REPORTED NO SAMPLE AVAILABLE PICTURES AVAILABLE FINDING: DURING SAMPLE EVALUATION THE CAP WAS FOUND TO BE SEPARATED FROM Y-PORT. CRACKED CAP. NOTE: THIS WAS A K-ZERO CONNECTOR ON A FRESENIUS KABI IVENIX ADMINISTRATION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23700009 ↗MalfunctionDec 2, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: ON 11/3/2025 THEY RECEIVED AN IVENIX COMPLAINT/DISPOSABLE FOR EVALUATION AND DURING EVALUATION THEY FOUND A DEFECT WITH A BLONIE PRODUCT. SEE ATTACHED FOR DESCRIPTION OF ISSUE. NO ADVERSE EVENT REPORTED NO SAMPLE AVAILABLE PICTURES AVAILABLE FINDINGS: DURING SAMPLE EVALUATION, TWO DEFECTS WERE IDENTIFIED: -MISSING SILICONE BARRIER -CRACKED CAP NOTE: THIS WAS A K-ZERO CONNECTOR ON A FRESENIUS KABI IVENIX ADMINISTRATION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23699854 ↗MalfunctionDec 2, 2025No Apparent Adverse Event; No Apparent Adverse Event
THE FOLLOWING HAS BEEN REPORTED: ON 11/3/2025 THEY RECEIVED AN IVENIX COMPLAINT/DISPOSABLE FOR EVALUATION AND DURING EVALUATION THEY FOUND A DEFECT WITH A BLONIE PRODUCT. SEE ATTACHED FOR DESCRIPTION OF ISSUE. NO ADVERSE EVENT REPORTED. NO SAMPLE AVAILABLE. PICTURES AVAILABLE. FINDINGS: DURING SAMPLE EVALUATION, TWO DEFECTS WERE IDENTIFIED: MISSING SILICONE BARRIER, AND A CRACKED CAP. NOTE: THIS WAS A K-ZERO CONNECTOR ON A FRESENIUS KABI IVENIX ADMINISTRATION SET.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23699853 ↗
Other Set, Administration, Intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page