DORO® Easy-Connect Navigation Adaptor 1204.002Recalls & reported problems

Black Forest Medical GmbH · DORO® Easy-Connect Navigation Adaptor, BRAINLAB · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04250435506738), received between Mar 14, 2025 and Mar 14, 2025 — classified as 1 malfunction.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K001808); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 14, 2025Class IIOpen, Classified

    Potential of compromised compatibility resulting in influence on the positioning of the reference array and an influence of the function of the navigation system cannot be excluded.

    Z-1860-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 14, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    DURING A COMPARISON OF THE DRAWINGS OF THE PRODUCTS 1204.002 DORO® EASY-CONNECT NAVIGATION ADAPTER, BRAINLAB AND 41725 STANDARD CRANIAL REFERENCE ARRAY, IT WAS DETERMINED THAT IN CERTAIN INSTANCES THE COMPATIBILITY OF BOTH DEVICES MIGHT BE COMPROMISED. THIS MIGHT HAVE AN INFLUENCE ON THE POSITIONING OF THE REFERENCE ARRAY. DUE TO THAT AN INFLUENCE OF THE FUNCTION OF THE NAVIGATION SYSTEM CANNOT BE EXCLUDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21607462 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001808. One submission often covers many models, so check the product and lot details in each record.

  • Z-0193-2018 · Nov 2, 2017 · Terminated
    pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into t

Other Holder, Head, Neurosurgical (Skull Clamp) devices

All Holder, Head, Neurosurgical (Skull Clamp) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page