HFD100 119695-000Recalls & reported problems
IMRIS Imaging, Inc. · Head Fixation Device HFD100 -ORT ORT100/200/300 METRIC · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00857534006608), received between May 14, 2022 and May 14, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Unintended Movement” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K103493); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Dec 8, 2025Open, Classified
Torque screw may crack or separate. If used, an affected screw could lead to inadequate head stabilization, compromised surgical precision, procedural delays, requirement for alternate head fixation methods, and risk of patient injury.
Z-1034-2026 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2022Unintended Movement; Unintended Movement
A CUSTOMER REPORTED THAT THE FEMALE AND MALE SKULL CLAMP COMPONENTS OF THEIR HFD100 DEVICE APPEAR TO EXHIBIT SLIGHT MOVEMENT IN RELATION TO EACH OTHER AFTER BEING FIXED WITH THE LOCKING MECHANISM. THE CUSTOMER ALSO REPORTED THIS MOVEMENT COULD IMPACT THEIR LITT CASES. NO ADVERSE PATIENT CONSEQUENCES OR PROCEDURAL OUTCOMES HAVE BEEN REPORTED IN CONNECTION WITH THIS INFORMATION. FURTHER INFORMATION AND ACCESS TO THE SPECIFIC HFD100 DEVICE AND COMPONENTS IS BEING SOUGHT FROM THE CUSTOMER TO INVESTIGATE THE CONCERN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14402887 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103493. One submission often covers many models, so check the product and lot details in each record.
- Z-0934-2022 · Feb 17, 2022 · TerminatedTrumpf Mount Clamp Assembly (table adapter component), part #118704-000, a component of the IMRIS HFD100 head fixation device, models #119629-000 and #119630-000, which are used wi
- Z-3211-2017 · Aug 14, 2017 · TerminatedIMRIS DEERFIELD IMAGING, REF 113803, Starburst adaptor thumb screw
Other Holder, Head, Neurosurgical (Skull Clamp) devices
- MAYFIELD® A1059Integra Lifesciences Corporation · HOLDER, HEAD, NEUROSURGICAL (SKULL CLAMP)506 reports
- MAYFIELD® A3059Integra Lifesciences Corporation · HOLDER, HEAD, NEUROSURGICAL (SKULL CLAMP)196 reports
- DORO® Skull Clamp 1001.001Black Forest Medical GmbH · Holder, Head, Neurosurgical (Skull Clamp)173 reports
- Head Fixation Device HFD100IMRIS Imaging, Inc. · Holder, Head, Neurosurgical (Skull Clamp)1 recall13 reports
- HFD100 119629-000IMRIS Imaging, Inc. · Holder, Head, Neurosurgical (Skull Clamp)1 recall6 reports
- DORO® Skull Clamp 3003-009Black Forest Medical GmbH · Holder, Head, Neurosurgical (Skull Clamp)52 reports
- DORO® Easy-Connect Navigation Adaptor 1204.002Black Forest Medical GmbH · Holder, Head, Neurosurgical (Skull Clamp)1 recall1 report
- HFD100 119630-000IMRIS Imaging, Inc. · Holder, Head, Neurosurgical (Skull Clamp)1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.