Breast BI 7 MR Coil Set 3T 10Recalls & reported problems

Noras MRI products GmbH · The intended use of the Breast Coil is, in conjunction with a Magnetic Resonance Scanner, the MR examination of the female breast. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the female breast. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis. The included Breast Biopsy Unit permits MR guided breast biopsy and placement of localization-wire by a trained physician. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04251269207389), received between Apr 23, 2026 and Aug 25, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Insufficient Device Problem Information3

Patient problems

  • Partial thickness (Second Degree) Burn5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem

    This is an initial 30-day report. This report is one of five related reports concerning local skin injuries reported in association with breast biopsy procedures performed using a NORAS Breast BI 7 biopsy coil, serial number (b)(6). The manufacturer's investigation has been completed, and the final investigation results are included in this initial report.The manufacturer was informed of this event as part of a cluster of five alleged local skin injuries occurring during the period from (b)(6) 2025 through (b)(6) 2026. No individual clinical questionnaire, photographs, procedure date, patient age, patient weight, detailed description of the injury, treatment information, or patient outcome information were made available for this specific patient. Therefore, the circumstances and clinical severity of this individual event could not be independently confirmed.The manufacturer completed its investigation of this event and the four related reports. The investigation included review of the available customer information, clinical information available for the related cases, imaging and system data, device history, service information, quality assurance testing, RF-related testing, and

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 50437731 ↗
  • InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 50437730 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    This is follow-up report 1 and the final investigation update for manufacturer report 3004929307-2026-00003. This report is one of five related reports concerning local skin injuries reported in association with breast biopsy procedures performed using a NORAS Breast BI 7 biopsy coil, serial number (b)(6).Additional information obtained during the investigation established that the event occurred on (b)(6) 2025. The patient was female; the patient's age and weight were not available. The injury was described by the customer as a second-degree burn. The skin injury was noticed approximately two days after the procedure. No additional medical treatment was documented, and complete recovery was reported after approximately one week.The patient underwent a breast biopsy in the prone, head-first position with symmetrical positioning and both arms placed in the Superman position. The left breast was biopsied. The procedure was performed using the Breast BI 7 biopsy coil with the standard grid and an EnCor/Enspire MRI vacuum biopsy system and associated accessories.The manufacturer completed its investigation of this event and the four related reports. The investigation included review of

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 24964883 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    This is follow-up report 1 and the final investigation update for manufacturer report 3004929307-2026-00002. This report is one of five related reports concerning local skin injuries associated with breast biopsy procedures performed at the same user facility using the same Breast BI 7 MR Coil, serial number (b)(6).The event occurred on (b)(6)2026. The patient was a 56-year-old female, 160 cm in height and weighing 65 kg. The reported skin lesion was located on the left breast above the nipple and measured approximately 6 cm by 2 cm. Blistering was observed within 24 hours, followed by ulceration. A lumpectomy was performed five days after the biopsy. The available records do not establish whether the lumpectomy was part of the treatment of the skin lesion.The procedure was performed in the prone, feet-first position with the arms in the Superman position. The left breast was biopsied. No contact with the inner wall of the MR bore and no skin-to-skin contact were reported. The patient did not report a sensation of heat during the procedure. The manufacturer's investigation has been completed. The investigation did not establish a device-related root cause. The final evaluation is p

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 24964862 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    THIS REPORT IS 1 OF 5 REPORTS ASSOCIATED WITH THE SAME DEVICE MODEL AND USER FACILITY. ON MARCH 24, 2026, THE MANUFACTURER WAS INFORMED OF A CLUSTER OF INCIDENTS. THE CUSTOMER ALLEGED THAT UP TO 5 PATIENTS SUSTAINED SUPERFICIAL BURNS DURING BREAST BIOPSY PROCEDURES BETWEEN (B)(6) 2025 AND (B)(6) 2026. THE EXAMINATIONS WERE PERFORMED USING THE NORAS BREAST BI 7 COIL (SN (B)(6)) IN CONJUNCTION WITH A SIEMENS MAGNETOM VIDA MRI SCANNER. PRELIMINARY QUALITY ASSURANCE (QA) CHECKS AND MR TESTING OF THE INVOLVED COIL (SN (B)(6)) CONDUCTED BY THE SYSTEM INTEGRATOR DID NOT REVEAL ANY FAULTS OR MALFUNCTIONS. HOWEVER, COMPREHENSIVE LABORATORY TESTING AND TECHNICAL EVALUATION BY THE LEGAL MANUFACTURER ARE STILL PENDING. THE ROOT CAUSE OF THE REPORTED BURNS REMAINS UNCONFIRMED. WE ARE ACTIVELY REVIEWING THE AVAILABLE CLINICAL DATA TO DETERMINE IF THE DEVICE CONTRIBUTED TO THE EVENT. AN UPDATED FOLLOW-UP REPORT WILL BE PROVIDED ONCE THE MANUFACTURER'S INVESTIGATION IS COMPLETED.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 24964839 ↗

Other Coil, Magnetic Resonance, Specialty devices