AquaBplus Water Purification System, Part Number 24-2000-0… recallZ-1364-2022 · Fresenius Medical Care Holdings, Inc. · initiated Aug 3, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

FDA root cause category: Labeling Change Control

What the company is doing

Customers were notified via telephone beginning 8/3/21. An FMCRTG, LLC Water Systems Specialist contacted the two consignees via phone and scheduled a time to perform the inspection on the AquaBplus 2000 devices. No formal written communication will be sent since calls and inspections have already performed.

Recall details

Recall number
Z-1364-2022
Status
Terminated
Initiated by the company
Aug 3, 2021
Recalling firm
Fresenius Medical Care Holdings, Inc.
Quantity
2 units
Product code
FIP
FDA submissions
K133829
Distribution
US Nationwide distribution in the states of NC and WV.

Product description

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Code information (models, lots, serial numbers)

UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.