Trokamed WA2PS20LRecalls & reported problems

TROKAMED GmbH · Sheath, 20 Fr., 160 mm · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
2
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04251303810919), received between May 8, 2024 and Jun 14, 2024 — classified as 2 deaths.

The most frequently reported specific device problem is “Inability to Irrigate” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 27, 2024Open, Classified

    If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.

    Z-0714-2025 · Root cause: Labeling False and Misleading · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Inability to Irrigate1
  • Restricted Flow rate1

Patient problems

  • Cardiac Arrest2
  • Hemorrhage/Blood Loss/Bleeding2
  • Respiratory Failure2
  • Aneurysm1
  • Ascites1
  • Disseminated Intravascular Coagulation (DIC)1
  • Hypoxia1
  • Increased Intra-Peritoneal Volume (IIPV)1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • DeathJun 14, 2024Restricted Flow rate; Inability to Irrigate; Restricted Flow rate; Inability to Irrigate

    ON (B)(6) 2024: THE PATIENT WAS TREATED IN HOSPITAL WITH THE OLYMPUS MINI-PCNL TRAY MINI NEPHROSCOPE DUE TO THE FOLLOWING DIAGNOSIS: 4-CM OBSTRUCTIVE LEFT PELVIC STAGHORN CALCULUS WITH CHRONIC HYDRONEPHROSIS AND RENAL CORTICAL THINNING, MULTIPLE NONOBSTRUCTIVE LEFT RENAL CALYCEAL CALCULI, MARGINAL KIDNEY FUNCTION AT 20%. DURING THE PROCEDURE, THE PHYSICIAN RECOGNISED THAT THERE WAS NOT ENOUGH OUTFLOW FROM THE WORKING SHEATH TOGETHER WITH POOR VISUALIZATION, RESULTING IN INCREASED INTRARENAL PRESSURE. THE PHYSICIAN MADE SEVERAL ATTEMPTS TO SWITCH IRRIGATION PORT AND TESTED FOR POSSIBLE CONTINUOUS FLOW, BUT THIS WAS UNSUCCESSFUL. AS A FINAL SOLUTION DURING SURGERY, THE PHYSICIAN HAD TO REMOVE THE NEPHROSCOPE INTERMITTENTLY TO LET THE IRRIGATION DRAIN OUT AND PRESSURE IRIGATION WAS USED. THE SURGERY CONTINUED AND THE STONE WAS BROKEN COMPLETELY WITH A LARGE BORE LASER. THE PHYSICIAN LATER REALIZED THAT HE WAS NO LONGER GETTING FLOW OUT, EVEN WHEN HE REMOVED THE NEPHROSCOPE. HE DECIDED TO PLACE STENTS, STOP THE PROCEDURE, AND PERFORM THE REST OF THE PROCEDURE AT ANOTHER TIME. THE PATIENT DEVELOPED ASCITES AND ABOUT 1 LITER WAS DRAINED A RUPTURE OF THE LOWER KIDNEY AND OOZING WAS VISIBL

    Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding; Respiratory Failure; Shock

    MAUDE report 19536011 ↗
  • DeathMay 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE, THERE WAS NO FLOW OUT OF THE OUTER SHEATH RESULTING IN INCREASED INTRARENAL PRESSURE CAUSING COMPLICATIONS. WHEN AN OLYMPUS REPRESENTATIVE WAS ONSITE AT THE CUSTOMER, THEY WERE INFORMED THAT AFTER THE PCNL PROCEDURE THE PATIENT HAD ADDITIONAL SURGERIES AND ENDED UP PASSING AWAY. THE PATIENT WAS INITIALLY DIAGNOSED WITH A 4-CENTIMETER OBSTRUCTIVE LEFT PELVIC STAGHORN CALCULUS WITH CHRONIC HYDRONEPHROSIS AND RENAL CORTICAL THINNING WITH MULTIPLE NON-OBSTRUCTIVE LEFT RENAL CALYCEAL CALCULI. THE PATIENT HAD MARGINAL KIDNEY FUNCTION AT 20%. THE PATIENT DID NOT WANT TO LOSE HER KIDNEY. THE SURGEON DISCUSSED OPTIONS WITH THE PATIENT AND INFORMED HER THAT SHE WOULD LIKELY LOSE THE KIDNEY IN THE FUTURE, BUT THEY DECIDED "TO GIVE THIS PROCEDURE A TRY". ON (B)(6) 2024 THE PATIENT UNDERWENT A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE USING AN OLYMPUS MINI-PCNL TRAY WITH MINI NEPHROSCOPE. DURING THE PROCEDURE, IT WAS REPORTED THAT THERE WAS NOT ENOUGH OUTFLOW FROM THE WORKING SHEATH AND THERE WAS POOR VISUALIZATION. SEVERAL ATTEMPTS WERE MADE TO SWITCH THE IRRIGATION PORT AND TEST FOR POSSIBLE CONTINUOUS FLOW, BUT THEY

    Patient problems: Aneurysm; Ascites; Cardiac Arrest; Disseminated Intravascular Coagulation (DIC); Hemorrhage/Blood Loss/Bleeding; Hypoxia; Increased Intra-Peritoneal Volume (IIPV); Ischemia; Necrosis; Respiratory Failure; Rupture; Septic Shock; Thrombosis/Thrombus; Unspecified Kidney or Urinary Problem; Vascular Dissection

    MAUDE report 19273992 ↗

Other Endoscopic Access Overtube, Gastroenterology-Urology devices

All Endoscopic Access Overtube, Gastroenterology-Urology devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.