Trokamed WA2PS20SRecalls & reported problems
TROKAMED GmbH · Sheath, 20 Fr., 130 mm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 2
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04251303810926), received between May 8, 2024 and Jun 14, 2024 — classified as 2 deaths.
The most frequently reported specific device problem is “Inability to Irrigate” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 27, 2024Open, Classified
If the inflow speed is too quick, it could result in reduced outflow through the shaft due to blood, tissue fragments and kidney stone fragments, and quickly build up of intrarenal pressure.
Z-0713-2025 · Root cause: Labeling False and Misleading · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Inability to Irrigate1
- Restricted Flow rate1
Patient problems
- Cardiac Arrest2
- Hemorrhage/Blood Loss/Bleeding2
- Respiratory Failure2
- Septic Shock2
- Aneurysm1
- Ascites1
- Disseminated Intravascular Coagulation (DIC)1
- Hypoxia1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
DeathJun 14, 2024Restricted Flow rate; Inability to Irrigate; Restricted Flow rate; Inability to Irrigate
ON (B)(6) 2024: THE PATIENT WAS TREATED IN HOSPITAL WITH THE OLYMPUS MINI-PCNL TRAY MINI NEPHROSCOPE DUE TO THE FOLLOWING DIAGNOSIS: 4-CM OBSTRUCTIVE LEFT PELVIC STAGHORN CALCULUS WITH CHRONIC HYDRONEPHROSIS AND RENAL CORTICAL THINNING, MULTIPLE NONOBSTRUCTIVE LEFT RENAL CALYCEAL CALCULI, MARGINAL KIDNEY FUNCTION AT 20%. DURING THE PROCEDURE, THE PHYSICIAN RECOGNISED THAT THERE WAS NOT ENOUGH OUTFLOW FROM THE WORKING SHEATH TOGETHER WITH POOR VISUALIZATION, RESULTING IN INCREASED INTRARENAL PRESSURE. THE PHYSICIAN MADE SEVERAL ATTEMPTS TO SWITCH IRRIGATION PORT AND TESTED FOR POSSIBLE CONTINUOUS FLOW, BUT THIS WAS UNSUCCESSFUL. AS A FINAL SOLUTION DURING SURGERY, THE PHYSICIAN HAD TO REMOVE THE NEPHROSCOPE INTERMITTENTLY TO LET THE IRRIGATION DRAIN OUT AND PRESSURE IRIGATION WAS USED. THE SURGERY CONTINUED AND THE STONE WAS BROKEN COMPLETELY WITH A LARGE BORE LASER. THE PHYSICIAN LATER REALIZED THAT HE WAS NO LONGER GETTING FLOW OUT, EVEN WHEN HE REMOVED THE NEPHROSCOPE. HE DECIDED TO PLACE STENTS, STOP THE PROCEDURE, AND PERFORM THE REST OF THE PROCEDURE AT ANOTHER TIME. THE PATIENT DEVELOPED ASCITES AND ABOUT 1 LITER WAS DRAINED A RUPTURE OF THE LOWER KIDNEY AND OOZING WAS VISIBL
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding; Respiratory Failure; Septic Shock
MAUDE report 19536211 ↗DeathMay 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED, DURING A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. THERE WAS NO FLOW OUT OF THE OUTER SHEATH. RESULTING IN, INCREASED INTRARENAL PRESSURE. CAUSING COMPLICATIONS. WHEN AN OLYMPUS REPRESENTATIVE WAS ONSITE AT THE CUSTOMER, THEY WERE INFORMED, THAT AFTER THE PCNL PROCEDURE, THE PATIENT HAD ADDITIONAL SURGERIES. AND ENDED UP PASSING AWAY. THE PATIENT, WAS INITIALLY DIAGNOSED WITH A 4-CENTIMETER OBSTRUCTIVE LEFT PELVIC STAGHORN CALCULUS. WITH CHRONIC HYDRONEPHROSIS AND RENAL CORTICAL THINNING WITH MULTIPLE NON-OBSTRUCTIVE LEFT RENAL CALYCEAL CALCULI. THE PATIENT, HAD MARGINAL KIDNEY FUNCTION AT 20%. THE PATIENT, DID NOT WANT TO LOSE HER KIDNEY. THE SURGEON, DISCUSSED OPTIONS WITH THE PATIENT AND INFORMED HER, THAT SHE WOULD LIKELY, LOSE THE KIDNEY IN THE FUTURE, BUT THEY DECIDED "TO GIVE THIS PROCEDURE A TRY". ON (B)(6) 2024, THE PATIENT UNDERWENT A PERCUTANEOUS NEPHROLITHOTOMY (PCNL) PROCEDURE. USING AN OLYMPUS MINI-PCNL TRAY WITH MINI NEPHROSCOPE. DURING THE PROCEDURE, IT WAS REPORTED, THAT THERE WAS NOT ENOUGH OUTFLOW FROM THE WORKING SHEATH. AND THERE WAS POOR VISUALIZATION. SEVERAL ATTEMPTS WERE MADE TO SWITCH THE IRRIGATION PORT AND TEST FOR POSSIBLE CONTINUO
Patient problems: Aneurysm; Ascites; Cardiac Arrest; Disseminated Intravascular Coagulation (DIC); Hemorrhage/Blood Loss/Bleeding; Hypoxia; Increased Intra-Peritoneal Volume (IIPV); Ischemia; Necrosis; Respiratory Failure; Rupture; Septic Shock; Thrombosis/Thrombus; Unspecified Kidney or Urinary Problem; Vascular Dissection
MAUDE report 19273996 ↗
Other Endoscopic Access Overtube, Gastroenterology-Urology devices
- EVIS EXERA II SMALL INTESTINAL VIDEOSCOPE OLYMPUS SIF TYPE Q180Olympus Medical Systems Corp. · endoscopic access overtube, gastroenterology-urology1 recall42 reports
- Trokamed WA2PS20LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall2 deaths2 reports
- Trokamed WA2PS18LTROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Trokamed WA2PS18STROKAMED GmbH · Endoscopic Access Overtube, Gastroenterology-Urology1 recall0 reports
- Navigator HD M0062502220Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology1 death31 reports
- Navigator HD M0062502230Boston Scientific Corporation · endoscopic access overtube, gastroenterology-urology29 reports
- Proxis™ Ureteral Access Sheath 231045C. R. Bard, Inc. · endoscopic access overtube, gastroenterology-urology22 reports
- Flexor Parallel G34874Cook Incorporated · endoscopic access overtube, gastroenterology-urology21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.