Skip to content

1077 DPOAE/TEOAE PROBE 8-69-41100Recalls & reported problems

PATH medical GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Nov 6, 2025Class IIOpen, Classified

    Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

    Z-0940-2026 · Root cause: Component change control · Recall details · FDA record ↗

Other Stimulator, Auditory, Evoked Response devices

All Stimulator, Auditory, Evoked Response devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page