MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100) recallZ-0940-2026 · Path · initiated Nov 6, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Path or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome.

FDA root cause category: Component change control

What the company is doing

On November 6, 2025 URGENT: MEDICAL DEVICE Field Corrective Action letters were sent to their two distributors. Actions to be taken: - Users will be asked to confirm the DPOAE protocol in use. - Verify the DPOAE Protocol used is not affected by the issue (DP1, DP2, DP3, DP4, DP17, DP18, DP19, DP20, DP21). - If any other protocol is used, please replace it by using AccuLink Software. Follow instructions in IFU to do so. - Remove all affected protocols (DP5-DP16) using AccuLink Software. Follow instructions in IFU to do so. - Complete and return the attached form to confirm these steps were followed. - The expected lifetime of the probes is two years from date of manufacture. Please put the products out of use, once this is reached.

Recall details

Recall number
Z-0940-2026
Status
Open, Classified
Initiated by the company
Nov 6, 2025
Posted by the FDA
Dec 18, 2025
Recalling firm
Path
Product code
GWJ
FDA submissions
K100661
Distribution
Domestic: WI. International: Ireland.

Product description

MADSEN ACCUSCREEN TEOAE/DPOAE/ABR Probe (P/N 8-69-41100)

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Part Number: 8-69-41100; UDI-DI: 04260223141355; Serial Numbers: 1008852, 1008888, 1008916, 1008925, 1008935, 1008945, 1008951, 1008957, 1008960, 1008962, 1008966, 1008967, 1008987, 1008990, 1009054, 1009065, 1009066, 1009131, 1009137, 1009138, 1009139, 1009145, 1009148, 1009155, 1009159, 1009161, 1009169, 1009184, 1009188, 1009191, 1009305, 1009315, 1009316, 1009319, 1009321, 1009322, 1009325, 1009328, 1009343, 1009352, 1009356, 1009357, 1009359, 1009366, 1009372, 1009375, 1009376, 1009386, 1009388, 1009390, 1009396, 1009399, 1009400, 1009404, 1009411, 1009415, 1009419, 1009426, 1009427, 1009432, 1009439, 1009442, 1009454, 1009456, 1009460, 1009463, 1009464, 1009468, 1009475, 1009481, 1009488, 1009494, 1009495, 1009497, 1009848, 1009855.