exoplan 3.1 RijekaRecalls & reported problems
exocad GmbH · exoplan is a medical software, intended to support the pre-operative planning of dental implants using the visualization of the implant placement within images of the patient’s anatomy. The process is based on CT/CBCT data sets originating from other medical devices, and can be supported by optical scan(s) of the patient’s anatomy as well as a virtual prosthetic proposal. exoplan allows the design of surgical guides to support the placement of endosseous dental implants in guided surgery. The design of surgical guides is based on 3D surface data representing the preoperative situation and approved implant positions. Alternatively, instead of optical surface data a second CBCT/CT dataset can be used. The software exports the planning and design results as geometrical data and a digital 3D model of the surgical guide to support the manufacture of a separate physical product. exoplan does not extend or change indications of dental implants. Usage of a surgical guide designed with the software does not change the necessary due diligence required compared to conventional (non-guided) surgery. The software is intended to be used only by dental professionals with sufficient medical training in dental implantology and surgical dentistry in office environments suitable for reading diagnostic dental DICOM data sets. exoplan shall not be used for any purpose other than planning dental implant placement or design of surgical guides. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Aug 17, 2023Open, Classified
A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is selected, along with a Step-by-Step Full Drill Protocol export implant library, exoplan software does not filter out unsupported sleeve height positions for a particular sub-full drill protocol and instead shows ALL possible sleeve height positions for all sub-full drill protocols. The use of an improper sleeve height position in a surgical guide could result in patient injury
Z-2633-2023 · Root cause: Software design · Recall details · FDA record ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page