Skip to content

i-ED COIL 373-040815Recalls & reported problems

Kaneka Corporation · The i-ED COIL is used for vascular embolization with a platinum coil at the target lesion in the patient’s blood vessel. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04540778177635), received between Oct 7, 2020 and Jan 18, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Ruptured Aneurysm1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 18, 2024Fracture; Fracture

    BALLOON ASSISTED COIL EMBOLIZATION OF AN LEFT ICA (INTERNAL CAROTID ARTERY) ANEURYSM, DIAMETER OF 6MM WIDTH X 4MM HEIGHT X 5MM DEPTH. AFTER DELIVERY OF A BALLOON CATHETER (TRANSFORM SUPER COMPLIANT 4MM X 10CM), PLACED IN THE VESSEL PAST THE ANEURYSM FOR SUPPORT DURING THE PROCEDURE, AN SL10 STRAIGHT TIP MICROCATHETER FOR COIL DELIVERY WAS PLACED INTO THE ANEURYSM. AN I-ED COIL COMPLEX SOFT (392-0620), LOT# KR033037 WAS DELIVERED THROUGH THE MICROCATHETER AND PLACED INTO THE ANEURYSM WITH ASSIST FROM THE BALLOON CATHETER TO DELIVER THE COIL INTO PLACE. THE COIL WAS DETACHED AND IMPLANTED WITHOUT PROBLEM. THE CONCERNED I-ED COIL 14 EXTRASOFT (373-040815) WAS SELECTED AS THE SECOND COIL AND DELIVERED THROUGH THE MICROCATHETER. THE FIRST HALF OF THE COIL WAS DELIVERED WITHOUT PROBLEM BUT FROM THERE THE PHYSICIAN EXPERIENCED CHALLENGES IN DELIVERING THE REMAINDER OF THE COIL AS SOME RESISTANCE WAS MET AND THE MICROCATHETER WAS KICKED OUT OF THE ANEURYSM AS THE COIL HERNIATED OUT. THE BALLOON CATHETER WAS INFLATED A NUMBER OF TIMES IN EFFORT TO KEEP THE COIL IN THE ANEURYSM BUT COULD NOT. AFTER SEVERAL ATTEMPT OF PULLING BACK AND REDELIVERY OF THE COIL, STRETCHING OF THE COIL WAS CONFIRM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18533695 ↗
  • MalfunctionJan 17, 2024Fracture; Fracture

    BALLOON ASSISTED COIL EMBOLIZATION OF AN LEFT ICA (INTERNAL CAROTID ARTERY) ANEURYSM, DIAMETER OF 6MM WIDTH X 4MM HEIGHT X 5MM DEPTH. AFTER DELIVERY OF A BALLOON CATHETER (TRANSFORM SUPER COMPLIANT 4MM X 10CM), PLACED IN THE VESSEL PAST THE ANEURYSM FOR SUPPORT DURING THE PROCEDURE, AN SL10 STRAIGHT TIP MICROCATHETER FOR COIL DELIVERY WAS PLACED INTO THE ANEURYSM. AN I-ED COIL COMPLEX SOFT (392-0620), LOT# KR033037 WAS DELIVERED THROUGH THE MICROCATHETER AND PLACED INTO THE ANEURYSM WITH ASSIST FROM THE BALLOON CATHETER TO DELIVER THE COIL INTO PLACE. THE COIL WAS DETACHED AND IMPLANTED WITHOUT PROBLEM. THE CONCERNED I-ED COIL 14 EXTRASOFT (373-040815) WAS SELECTED AS THE SECOND COIL AND DELIVERED THROUGH THE MICROCATHETER. THE FIRST HALF OF THE COIL WAS DELIVERED WITHOUT PROBLEM BUT FROM THERE THE PHYSICIAN EXPERIENCED CHALLENGES IN DELIVERING THE REMAINDER OF THE COIL AS SOME RESISTANCE WAS MET AND THE MICROCATHETER WAS KICKED OUT OF THE ANEURYSM AS THE COIL HERNIATED OUT. THE BALLOON CATHETER WAS INFLATED A NUMBER OF TIMES IN EFFORT TO KEEP THE COIL IN THE ANEURYSM BUT COULD NOT. AFTER SEVERAL ATTEMPT OF PULLING BACK AND REDELIVERY OF THE COIL, STRETCHING OF THE COIL WAS CONFIRM

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18531822 ↗
  • InjuryOct 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    COIL EMBOLIZATION OF WAITING CASES OF INTERNAL CAROTID ARTERY C2 AND UNRUPTURED ANEURYSM. AFTER THE ASSIST STENT WAS PLACED, THE PRODUCT WAS USED AS THE FIRST FRAMING FROM THE MICROCATHETER THROUGH THE STENT STRUT. THE 3 CM TO 4 CM PART AT THE REAR END OF THE COIL DID NOT FOLD WELL AND THE LOOP CAME OUT OF THE STENT. WHILE MANIPULATING THE CATHETER TO CORRECT IT, THE COIL PIERCED THE AREA NEAR THE BLEB AT THE TIP OF THE ANEURYSM. THE PRODUCT WAS LEFT IN PLACE AS IT WAS, BUT THE COIL EXTENDED INTO THE SUBARACHNOID SPACE. WHILE EXPANDING THE BALLOON CATHETER NEAR THE ANEURYSM TO STOP BLEEDING, ANOTHER 6 COILS WERE PLACED FOR THE PURPOSE OF STOPPING BLEEDING, BUT THE BLEEDING DID NOT STOP EASILY. PROTAMINE WAS ADMINISTERED, AND FINALLY HEMOSTASIS WAS MAINTAINED FOR ABOUT 1 HOUR. ALTHOUGH THE PATIENT'S BLOOD PRESSURE WAS LOW, THE AMOUNT OF BLEEDING WAS LARGE, SO EXTERNAL VENTRICULAR DRAIN WAS PERFORMED AS IT WAS. AFTER DRAINAGE, A SLIGHT REDUCTION IN THE PUPIL WAS CONFIRMED, BUT SEVERE DISTURBANCE OF CONSCIOUSNESS REMAINED AND HE WAS HOSPITALIZED FOR FOLLOW-UP.

    Patient problems: Ruptured Aneurysm

    MAUDE report 10640576 ↗

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →

Sources and updates