i-ED COIL 393-030530Recalls & reported problems

Kaneka Corporation · The i-ED COIL is used for vascular embolization with a platinum coil at the target lesion in the patient’s blood vessel. · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04540778190429), received between Apr 24, 2025 and Apr 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2025Fracture; Fracture

    35MMX 30CM COIL WAS PLACED IN THE POSTERIOR DIVISION OF THE LEFT MMA AFTER A 2X 12CM COIL WAS SUCCESSFULLY DEPLOYED. COIL WAS COMPLETELY OUT OF THE SL10 CATHETER, BUT MD WANTED THE COIL TO FLIP INTO THE FRONTAL DIVISION. THE COIL WAS TAKEN OUT AND PUT ON THE BACK TABLE. THE MD THEN RESELECTED THE FRONTAL DIVISION WITH A WIRE AND THE SAME SL10. THE SAME 35MMX30CM, WAS THEN PLACED INTO THE SMALLER FRONTAL DIVISION. THE COIL HAD A VERY HARD TIME FORMING ITS SHAPE IN THE SMALL VESSEL. THE MD NOTICED THAT HE WAS USING A LOT OF FORCE TO PUSH THE COIL OUT TO THE POINT THAT HE BENT THE WIRE OF THE COIL. SO THAT HE WAS SUGGESTED TO REMOVE THE COIL AND USE ANOTHER 2X12CM COIL. AS HE WAS REMOVING THE COIL, THE CATHETER MOVED INTO A FAVORABLE POSITION, SO ONE LAST ATTEMPT WAS TRIED WITH35MM X 30CM (3RD ATTEMPT). AT THAT POINT, THE COIL BROKE WITH EXCESSIVE FORCE AND NOT A LOT OF SUPPORT (BENCHMARK 71 AND SL10). A 2MM SNARE WAS USED TO REMOVE THE COIL THAT WAS LEFT IN THE VESSEL WITH NO COMPLICATIONS NOTICED TO THE PATIENT. NO FURTHER ACTION WAS NEEDED BUT SINCE THE COIL BROKE (STRETCHED) AND A SNARE (RETRIEVAL DEVICE) WAS USED A COMPLAINT WAS DOCUMENTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898653 ↗

Other Device, Neurovascular Embolization devices

All Device, Neurovascular Embolization devices →