Skip to content

NA 2296101216Recalls & reported problems

Stryker Corporation · 1.6mm Cross-Cut Fissure Carbide Bur · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540046161), received between Jan 25, 2021 and Jan 25, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 25, 2021Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE BUR BROKE. IT WAS ALSO REPORTED THAT A FRAGMENT OF THE BUR REMAINED IN THE PATIENTS JAW AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE BUR WAS NOT REMOVED AND NO MEDICAL INTERVENTION WAS PERFORMED. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11225117 ↗

Other BUR, DENTAL devices

All BUR, DENTAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page