NA 2296003108Recalls & reported problems
Stryker Corporation · Precision Thin (13.0 x 0.38 x 39.0mm) · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540046482), received between Dec 16, 2019 and Sep 12, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 8, 2025Open, Classified
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
Z-1138-2025 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Fracture3
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 12, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15403295 ↗MalfunctionDec 16, 2019Fracture; Fracture
IT WAS REPORTED THAT DURING A COMMANDO PROCEDURE (COMBINED MANDIBULECTOMY AND NECK DISSECTION OPERATION), TWO BLADES BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR THE SECOND BLADE THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9474792 ↗MalfunctionDec 16, 2019Fracture; Fracture
IT WAS REPORTED THAT DURING A COMMANDO PROCEDURE (COMBINED MANDIBULECTOMY AND NECK DISSECTION OPERATION), TWO BLADES BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR THE FIRST BLADE THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9474767 ↗
Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices
- NA 511.801Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL223 reports
- Formula 0375544000Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL98 reports
- NA 0277096325Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL95 reports
- NA 310.25Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL91 reports
- Arthrex® AR-8400TDArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL86 reports
- NA 0277096275Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL72 reports
- Arthrex® AR-7300DSArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL70 reports
- NA 0277096277Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.