Stryker Corporation · Precision Thin (9.0 x 0.38 x 31.0mm) · Class 1
Quick summary
This Stryker Precision Thin saw blade is a surgical cutting blade. Reports mostly describe the blade breaking or fracturing during surgery, sometimes causing delays for replacement or imaging. There are 24 reports that name its identifier, with no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 04546540046550), received between Mar 26, 2019 and Apr 21, 2025 — classified as 23 malfunctions.
The most frequently reported specific device problem is “Fracture” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture17
- Break11
Patient problems
- No Consequences Or Impact To Patient16
- No Clinical Signs, Symptoms or Conditions10
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876034 ↗MalfunctionOct 21, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT REPORTEDLY REQUIRED IMAGING TO BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20500364 ↗MalfunctionSep 6, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TRANSMETATARSAL AMPUTATION PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THE CASE WAS DELAYED WAITING FOR NEW BLADE AND X-RAY TO ARRIVE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15362206 ↗MalfunctionJan 6, 2021Fracture; Fracture
IT WAS REPORTED THAT DURING A PELVIC OSTEOTOMY PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11127239 ↗MalfunctionJan 6, 2021Fracture; Fracture
IT WAS REPORTED THAT DURING AN OSTEOTOMY PROCEDURE, THE BLADE BROKE. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11127092 ↗MalfunctionNov 16, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING AN ACL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10844213 ↗MalfunctionAug 10, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A HALLUX PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10384221 ↗MalfunctionAug 10, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A HALLUX PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10384220 ↗MalfunctionAug 10, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING A HALLUX PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10384219 ↗MalfunctionAug 7, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY, THE BLADE BROKE. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10379008 ↗MalfunctionJul 20, 2020Fracture; Fracture
IT WAS REPORTED VIA MEDWATCH REPORT (B)(4) THAT DURING A TOTAL KNEE PROSTHESIS PROCEDURE, A PIECE OF THE MINI OSCILLATING SAW BLADE BROKE OFF DURING THE EXPLANT PORTION OF THE PROCEDURE. IT WAS ALSO REPORTED THAT AN INTRAOPERATIVE X-RAY WAS PERFORMED PRIOR TO FINAL CLOSURE. THE DELAY INCLUDED THE X-RAY AND THE ATTEMPT TO LOCATE THE MISSING PIECE. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AS A RESULTS OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10295054 ↗MalfunctionMar 19, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING AN OSTEOTOMY SURGERY, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9855820 ↗MalfunctionNov 7, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A COMMMANDO SURGICAL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9293293 ↗MalfunctionNov 7, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A COMMMANDO SURGICAL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9292111 ↗MalfunctionNov 7, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A COMMMANDO SURGICAL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9292108 ↗MalfunctionNov 7, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A COMMMANDO SURGICAL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9292102 ↗MalfunctionNov 7, 2019Break; Fracture; Break; Fracture
IT WAS REPORTED THAT DURING A COMMMANDO SURGICAL PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9292101 ↗MalfunctionJun 20, 2019Break; Break
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, IT WAS NOTED THAT THE BLADE BROKE WHERE IT FITS INTO THE HANDPIECE. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8717414 ↗MalfunctionMay 9, 2019Break; Break
IT WAS REPORTED THAT DURING AN ARTIFICIAL ANKLE JOINT SURGERY, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8595953 ↗MalfunctionApr 1, 2019Break; Break
IT WAS REPORTED THAT THE BLADE BROKE AT THE MOUNT DURING A FOOT/ANKLE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS A DELAY OF A COUPLE OF MINUTES OR LESS AS A RESULT TO GET A REPLACEMENT BLADE. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8470426 ↗MalfunctionApr 1, 2019Break; Break
IT WAS REPORTED THAT THE BLADE BROKE AT THE MOUNT DURING A FOOT/ANKLE PROCEDURE. IT WAS FURTHER REPORTED THAT THERE WAS A DELAY OF A COUPLE OF MINUTES OR LESS AS A RESULT TO GET A REPLACEMENT BLADE. IT WAS ALSO REPORTED THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8470418 ↗MalfunctionMar 26, 2019Break; Break
IT WAS REPORTED THAT TWO BLADES BROKE AT THE MOUNT DURING A HIGH TIBIAL OSTEOTOMY. IT WAS ALSO REPORTED THAT THERE WAS A DELAY OF A FEW MINUTES TO OBTAIN A REPLACEMENT BLADE. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. THIS REPORT IS FOR THE SECOND BLADE THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453345 ↗MalfunctionMar 26, 2019Break; Break
IT WAS REPORTED THAT TWO BLADES BROKE AT THE MOUNT DURING A HIGH TIBIAL OSTEOTOMY. IT WAS ALSO REPORTED THAT THERE WAS A DELAY OF A FEW MINUTES TO OBTAIN A REPLACEMENT BLADE. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THERE WERE NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. THIS REPORT IS FOR THE FIRST BLADE THAT BROKE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8453341 ↗
Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices
- NA 511.801Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL223 reports
- Formula 0375544000Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL98 reports
- NA 0277096325Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL95 reports
- NA 310.25Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL91 reports
- Arthrex® AR-8400TDArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL86 reports
- NA 0277096275Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL72 reports
- Arthrex® AR-7300DSArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL70 reports
- NA 0277096277Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL69 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.