Skip to content

NA 2296033225Recalls & reported problems

Stryker Corporation · Precision (9.0 x 0.51 x 31.0mm) · Class 1

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540047014), received between Sep 21, 2021 and May 26, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2026

    THE SURGEON WAS USING THE STRYKER SAGITTAL SAW BLADE TO CUT THE MANDIBLE WHEN IT BROKE INTO THREE PIECES. NO RESULTANT PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25277258 ↗
  • MalfunctionJul 29, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647062 ↗
  • MalfunctionApr 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876045 ↗
  • MalfunctionJan 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200293 ↗
  • MalfunctionSep 21, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ACL (ANTERIOR CRUCIATE LIGAMENT) PROCEDURE, THE BLADE BROKE AT THE MOUNT. IT WAS ALSO REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12507115 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.