NA 4100125000Recalls & reported problems
Stryker Corporation · Pin Collet · Class 1
Quick summary
This Stryker pin collet, an accessory for orthopedic power tools, has 85 reports in the FDA database, with no deaths and no injuries recorded. Many reports describe the device shedding metal debris or breaking down, and some describe overheating or parts detaching, mostly found outside of patient use. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 85
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 85 adverse event reports that list this exact device (DI 04546540068163), received between Jul 24, 2018 and Oct 30, 2025 — classified as 85 malfunctions.
The most frequently reported specific device problem is “Material Disintegration” (49 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Disintegration49
- Overheating of Device14
- Detachment of Device or Device Component11
- Fluid/Blood Leak10
- Fracture3
Patient problems
- No Clinical Signs, Symptoms or Conditions47
- No Patient Involvement31
- No Consequences Or Impact To Patient8
- No Known Impact Or Consequence To Patient5
- No Information3
- Burn(s)2
- Skin Discoloration1
Recent adverse event reports
The 25 most recent of 85 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 TO JUNE 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23423550 ↗MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399758 ↗MalfunctionJul 29, 2025Material Disintegration; Material Disintegration
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30 2025. THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: MATERIAL DISINTEGRATION. - 5 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22647278 ↗MalfunctionApr 21, 2025Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 2 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. -1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876832 ↗MalfunctionJan 21, 2025Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 15 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 13 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 2 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21201048 ↗MalfunctionOct 22, 2024Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. -THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501974 ↗MalfunctionOct 22, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501490 ↗MalfunctionJul 24, 2024Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19820465 ↗MalfunctionJul 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19816089 ↗MalfunctionApr 15, 2024Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 1 EVENT WHICH HAD A MILD INJURY, ILLNESS OR IMPAIRMENT WHICH CAN BE TREATED WITH MINIMAL OR NO INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113355 ↗MalfunctionJan 22, 2024Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556926 ↗MalfunctionOct 23, 2023Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17987276 ↗MalfunctionJul 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401910 ↗MalfunctionJul 26, 2023Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT WHICH HAD A MILD INJURY, ILLNESS OR IMPAIRMENT WHICH CAN BE TREATED WITH MINIMAL OR NO INTERVENTION.
Patient problems: Burn(s)
MAUDE report 17401453 ↗MalfunctionJul 26, 2023Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17399010 ↗MalfunctionApr 25, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807816 ↗MalfunctionApr 24, 2023Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16802894 ↗MalfunctionJan 29, 2023Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261991 ↗MalfunctionJan 29, 2023Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261870 ↗MalfunctionOct 25, 2022Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666411 ↗MalfunctionOct 24, 2022Fluid/Blood Leak; Fluid/Blood Leak
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663861 ↗MalfunctionOct 24, 2022Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15663776 ↗MalfunctionJul 26, 2022Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 3 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15102047 ↗MalfunctionApr 25, 2022Material Disintegration; Material Disintegration
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14193536 ↗MalfunctionApr 25, 2022Overheating of Device; Overheating of Device
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14193515 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.