Skip to content

Hudson 4100135000Recalls & reported problems

Stryker Corporation · Modified Trinkle Drill · Class 1

All models of Hudson (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 04546540068194), received between Jan 3, 2020 and Apr 22, 2026 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component7
  • Fracture3
  • Material Disintegration3
  • Overheating of Device3
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions14
  • No Patient Involvement3
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Detachment of Device or Device Component

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24953979 ↗
  • MalfunctionOct 28, 2025Material Fragmentation; Material Fragmentation

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: MATERIAL FRAGMENTATION - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399937 ↗
  • MalfunctionOct 28, 2025Material Disintegration; Material Disintegration

    NO CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399759 ↗
  • MalfunctionApr 21, 2025Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY OVERHEATED. - 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878627 ↗
  • MalfunctionApr 21, 2025Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 5 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876067 ↗
  • MalfunctionJan 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201186 ↗
  • MalfunctionOct 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20501552 ↗
  • MalfunctionJul 24, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822917 ↗
  • MalfunctionApr 15, 2024Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19113359 ↗
  • MalfunctionApr 25, 2023Overheating of Device; Overheating of Device

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY OVERHEATED.1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16808038 ↗
  • MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16261783 ↗
  • MalfunctionMar 24, 2022Fracture; Fracture

    THE USER FACILITY REPORTED THAT THE A DRILL BIT BROKE AND WAS LEFT IN THE PATIENT BONE DURING A PROCEDURE. THERE WAS NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 13890846 ↗
  • MalfunctionOct 27, 2021Overheating of Device; Overheating of Device

    THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12706031 ↗
  • MalfunctionJun 4, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE DEVICE WAS MISSING COMPONENTS DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11939586 ↗
  • MalfunctionAug 4, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING COMPONENTS WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 10363099 ↗
  • MalfunctionApr 22, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    THE DEVICE WAS FOUND TO BE MISSING COMPONENTS DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY.  THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9987199 ↗
  • MalfunctionJan 3, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD A MISSING COMPONENT WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 9547807 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates