NA 4100355000Recalls & reported problems
Stryker Corporation · Radiolucent Right Angle Drive · Class 1
Quick summary
This Stryker Radiolucent Right Angle Drive is a powered surgical attachment used in orthopedic procedures. The 30 reports mostly describe parts coming apart during surgery or service, and cutting accessories fracturing or breaking, with no patient harm described. There are no deaths and no injuries, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 04546540068279), received between Jul 24, 2018 and Aug 11, 2025 — classified as 30 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component24
- Fracture4
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions20
- No Patient Involvement9
- Insufficient Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING COMPONENT WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22777169 ↗MalfunctionAug 11, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING COMPONENT WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22775454 ↗MalfunctionApr 21, 2025Fracture; Fracture
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21876068 ↗MalfunctionJan 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21200898 ↗MalfunctionOct 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20501553 ↗MalfunctionJul 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19818723 ↗MalfunctionApr 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH.2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19113360 ↗MalfunctionJan 22, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18556922 ↗MalfunctionJan 29, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THIS REPORT SUMMARIZES 0 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16261932 ↗MalfunctionOct 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE RIVETS FELL OFF OF THE DEVICE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED AFTER A PROCEDURE AT THE END USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15652229 ↗MalfunctionOct 17, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE SCREW BROKE OFF DURING SURGERY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15618530 ↗MalfunctionOct 6, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS MISSING COMPONENTS DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15549946 ↗MalfunctionAug 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE WAS DISASSEMBLED. ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION ABOUT THE EVENT; NO FURTHER INFORMATION HAS BEEN RECEIVED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15190303 ↗MalfunctionMay 25, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS MISSING COMPONENTS DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14497912 ↗MalfunctionMay 6, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE WAS MISSING A SCREW DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14305604 ↗MalfunctionMar 28, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE PIECE WAS MISSING DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13934578 ↗MalfunctionNov 15, 2021Fracture; Fracture
THE USER FACILITY REPORTED THAT THE DEVICE DRILL BIT WAS BROKEN IN THE ATTACHMENT DURING A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12814336 ↗MalfunctionJan 6, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING COMPONENTS WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11127223 ↗MalfunctionDec 15, 2020Fracture; Fracture
THE USER FACILITY REPORTED THAT A BIT BROKE OFF IN THE DEVICE DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11006448 ↗MalfunctionNov 30, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE HAD A COMPONENT MISSING DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10922877 ↗MalfunctionMar 9, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS FOUND TO BE MISSING INTERNAL COMPONENTS DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 9806756 ↗MalfunctionNov 4, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE DEVICE HAD A COMPONENT MISSING AFTER PROCESSING. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 9271839 ↗MalfunctionJun 24, 2019Break; Break
THE USER FACILITY REPORTED THAT THE BIT BROKE OFF INSIDE THE ATTACHMENT DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Patient Involvement
MAUDE report 8727348 ↗MalfunctionMay 17, 2019Break; Break
THE DEVICE HAD A BIT BROKEN OFF INSIDE DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8619774 ↗MalfunctionMay 15, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
THE DEVICE WAS MISSING COMPONENTS DURING A SERVICE EVALUATION AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 8614164 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.