NA 0400820000Recalls & reported problems
Stryker Corporation · T4 Zipper Toga, (L/XL) · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 04546540152688), received between May 30, 2018 and Dec 11, 2018 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn14
- Material Separation9
- Torn Material9
- Material Integrity Problem5
Patient problems
- No Patient Involvement13
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2018Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING SETUP PRIOR TO A SURGICAL PROCEDURE AT THE USER FACILITY, A TEAR WAS FOUND ON THE TOGA. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; THERE WERE NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8153752 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562450 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562449 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562443 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562320 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562232 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562230 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7562226 ↗MalfunctionJun 1, 2018Material Separation; Torn Material; Material Split, Cut or Torn; Material Separation; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7560376 ↗MalfunctionMay 31, 2018Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7556916 ↗MalfunctionMay 31, 2018Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7556915 ↗MalfunctionMay 31, 2018Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7556914 ↗MalfunctionMay 31, 2018Material Integrity Problem; Material Split, Cut or Torn; Material Integrity Problem; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7556912 ↗MalfunctionMay 30, 2018Material Separation; Material Integrity Problem; Torn Material; Material Split, Cut or Torn; Material Separation; Material Integrity Problem; Torn Material; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING INSPECTION CONDUCTED AT THE USER FACILITY PRIOR TO THE START OF A SURGICAL PROCEDURE, THE OUTER COATING ON THE FRONT AND SLEEVES OF THE GOWN APPEARS TO HAVE COME AWAY FROM THE BACKING MATERIAL, AS WELL AS THINNING AND TEARING OF THE OUTER LAYER. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7554418 ↗
Other GOWN, SURGICAL devices
- Medline DYNJP2001SMedline Industries, Inc. · GOWN, SURGICAL2 recalls9 reports
- Medline DYNJP2202SMedline Industries, Inc. · GOWN, SURGICAL2 recalls5 reports
- MEDLINE DYNJP2002SMedline Industries, Inc. · GOWN, SURGICAL2 recalls4 reports
- Medline DYNJP2602Medline Industries, Inc. · GOWN, SURGICAL2 recalls3 reports
- Medline Industries, Inc. DYNJP2402Medline Industries, Inc. · GOWN, SURGICAL2 recalls2 reports
- MEDLINE DYNJP2002Medline Industries, Inc. · GOWN, SURGICAL2 recalls2 reports
- Medline DYNJP2103Medline Industries, Inc. · GOWN, SURGICAL2 recalls1 report
- Medline DYNJP2203Medline Industries, Inc. · GOWN, SURGICAL2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page