Skip to content

T2 1830-0003SRecalls & reported problems

Stryker GmbH · End Cap, Standard · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 04546540201140), received between Oct 19, 2023 and Oct 19, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • Limb Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 19, 2023Positioning Failure; Positioning Failure

    AS REPORTED: "IN A RETROGRADE CASE, THE TIMING OF COMPRESSION SCREW INSERTION WAS INCORRECT. THE COMPRESSION SCREW WAS INSERTED TO THE FRONT OF THE DYNAMIC HOLE WITH THE NAIL HOLDING SCREW REMOVED FROM THE NAIL. BECAUSE THE DEVICE AND THE NAIL WERE NOT SECURED, THE NAIL MIGRATED DEEPLY, RESULTING IN A FRACTURE WITH A THIRD BONE FRAGMENT WHEN ATTEMPTING TO REATTACH THE DEVICE AND THE NAIL IN THE MEDULLARY CAVITY. THE NAIL DEPTH WAS THEN READJUSTED USING A UNIVERSAL ROD AND PROXIMAL SCREW FIXATION WAS PERFORMED. WHEN COMPRESSION WAS ATTEMPTED WITH THE COMPRESSION SCREW AFTER INSERTION OF THE DISTAL SCREW, THE COMPRESSION SCREW COULD NOT MOVE AND COMPRESSION COULD NOT BE PERFORMED, PROBABLY DUE TO CROSS-THREADING. IN ADDITION, BECAUSE THE COMPRESSION SCREW HAD BLOCKED THE SCREW HOLE, THE PLANNED SCREW COULD NOT BE INSERTED AND ONLY ONE DISTAL SCREW FIXATION WAS NEEDED. AS FOR THE END CAP, IT WAS ALSO IMPLANTED WITH ABOUT 2 MM PROTRUDING FROM THE NAIL END DUE TO POOR PLACEMENT OF THE COMPRESSION SCREW." ADDITIONALLY A 60 MINUTE SURGICAL DELAY WAS REPORTED.

    Patient problems: Limb Fracture

    MAUDE report 17971556 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page