T2 1830-0725SRecalls & reported problems

Stryker GmbH · Humeral Nail · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04546540201317), received between Oct 30, 2024 and Dec 20, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Fracture2

Patient problems

  • Bone Fracture(s)1
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 20, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED, T2 HUMERAL NAIL BROKE AFTER PRIMARY SURGERY ON (B)(6), INTRAOPERATIVE X RAYS WERE TAKEN ON (B)(6) WITH MINIMAL SIGN OF FRACTURE BUT WHEN PATIENT WAS MONITORED ON (B)(6) FRACTURE WAS EVIDENT. THE IMPLANT WAS REMOVED FROM THE PATIENT AND WAS ELIMINATED.

    Patient problems: Bone Fracture(s); Failure of Implant

    MAUDE report 20995679 ↗
  • MalfunctionOct 30, 2024Break; Fracture; Break; Fracture

    AS REPORTED: "NAIL BREAKS WHEN PLACED IN THE NAIL INSERTION SPACE". ADDITIONAL INFORMATION RECEIVED 10/17/24; IT WAS A PRIMARY SURGERY. SURGEON DRILLED THE CANAL, PUT A HUMERUS NAIL, AND WHEN ADJUSTING THE COMPRESSION SCREW, HE LEANED 2.7 SCREWDRIVER ON THE PROXIMAL PART OF THE NAIL AND THE NAIL MATERIAL BECAME DEFORMED. NAIL HAD TO BE REMOVED AND ANOTHER ONE WITH SAME MEASURES WAS REPLACED. THERE WAS NO DEBRIS, ONLY MATERIAL FATIGUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20571068 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →