NA 1896-4024SRecalls & reported problems
Stryker GmbH · Locking Screw, Fully Threaded · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540202307), received between Sep 15, 2021 and Sep 3, 2025 — classified as 4 injuries.
The most frequently reported specific device problem is “Biocompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Biocompatibility2
- Device Dislodged or Dislocated1
- Fracture1
Patient problems
- Hypersensitivity/Allergic reaction2
- Pain2
- Deformity/ Disfigurement1
- Failure of Implant1
- Implant Pain1
- Loss of Range of Motion1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2025Biocompatibility; Biocompatibility
CUSTOMER SERVICE WAS CONTACTED BY A PATIENT INFORMING SHE HAS BEEN SUFFERING FROM SEVERE SKIN RASHES AND PAIN AFTER AN IMPLANT EMERGENCY SURGERY IN THE UPPER ARM AND REQUESTS SAMPLE OF THE COMPOSITION (METALS ETC) OF THE IMPLANTS AND SCREWS AS WELL AS A SAMPLE OF THE IMPLANTS SO THAT THE LAB/DERMATOLOGIST CAN DETERMINE WHETHER THE PARTS NEED TO BE REMOVED OR WHETHER THERE IS ANOTHER SOLUTION.
Patient problems: Hypersensitivity/Allergic reaction; Pain
MAUDE report 22950345 ↗InjurySep 3, 2025Biocompatibility; Biocompatibility
CUSTOMER SERVICE WAS CONTACTED BY A PATIENT INFORMING SHE HAS BEEN SUFFERING FROM SEVERE SKIN RASHES AND PAIN AFTER AN IMPLANT EMERGENCY SURGERY IN THE UPPER ARM AND REQUESTS SAMPLE OF THE COMPOSITION (METALS ETC) OF THE IMPLANTS AND SCREWS AS WELL AS A SAMPLE OF THE IMPLANTS SO THAT THE LAB/DERMATOLOGIST CAN DETERMINE WHETHER THE PARTS NEED TO BE REMOVED OR WHETHER THERE IS ANOTHER SOLUTION.
Patient problems: Hypersensitivity/Allergic reaction; Pain
MAUDE report 22950344 ↗InjuryJun 6, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE MANUFACTURER BECAME AWARE OF A LITERATURE PUBLISHED BY THE ¿DEPARTMENT OF ORTHOPAEDIC HAND AND MICROSURGERY CENTER, OSAKA EKISAIKAI HOSPITAL, JAPAN¿. THE TITLE OF THIS REPORT IS, ¿EVALUATION OF INTRAMEDULLARY NAIL FIXATION VIA THE NEVIASER PORTAL FOR PROXIMAL HUMERUS FRACTURE¿ PUBLISHED ON FEBRUARY 08, 2019, WHICH IS ASSOCIATED WITH THE STRYKER ¿T2 PROXIMAL HUMERAL NAILING SYSTEM¿. THE ARTICLE CAN BE FOUND AT HTTPS://DOI.ORG/10.1016/J.JCOT.2019.02.007. THIS REPORT INCLUDES AN ANALYSIS OF THE CLINICAL DATA THAT WAS COLLECTED ON 4 PATIENTS. THE CASES IN THIS STUDY RANGE FROM APRIL 2014 AND MARCH 2017. DURING THE REVIEW OF THE LITERATURE, IT WAS NOT POSSIBLE TO ESTABLISH A SPECIFIC DEVICE DETAIL, PATIENT INFORMATION, AND CURRENTLY NO ADDITIONAL DEVICE INFORMATION IS AVAILABLE. IT WAS REPORTED THAT 1 PATIENT EXPERIENCED PAIN, AND RESTRICTED RANGE OF MOTION (ROM) DUE TO FAILURE OF FIXATION WITH RECURRENT VARUS DEFORMITY AND SLIPPING OUT OF THE PROXIMAL INTERLOCKING SCREW. THE REPORT STATES, ¿IN THIS CASE, WE USED A BENT NAIL UNLIKE THE OTHER THREE CASES IN WHICH STRAIGHT NAILS WERE USED. WE CONSIDER THIS DIFFERENCE CAN BE A CAUSE OF THIS COMPLICATION. THIS PATIENT ALSO EXPERIENCED F
Patient problems: Deformity/ Disfigurement; Implant Pain; Loss of Range of Motion
MAUDE report 14610228 ↗InjurySep 15, 2021Fracture; Fracture
IT WAS REPORTED A REVISION OF THE SURGERY PERFORMED ABOUT A YEAR AGO WAS SCHEDULED DUE TO CONFIRMED POSTOPERATIVE SCREW FRACTURE. HOWEVER, BEFORE THE REVISION, A MISTAKE IN ORDERING THE INSTRUMENTS WAS DISCOVERED AND THE SURGERY WAS POSTPONED. THE PATIENT WAS UNDER GENERAL ANESTHESIA. REVISION SURGERY WAS RESCHEDULED FOR (B)(6) 2021. THIS RECORD CAPTURES THE POSTOPERATIVE SCREW FRACTURE.
Patient problems: Failure of Implant
MAUDE report 12478320 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page