Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540202338), received between Dec 22, 2020 and May 11, 2026 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration2
- Positioning Failure1
Patient problems
- Failure of Implant2
- Nonunion/Delayed-union Bone Fracture2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2026Positioning Failure
DURING T2PH SURGERY, THE SCREW PLACED IN THE MOST DISTAL OVAL HOLE PROTRUDED ANTERIORLY BEYOND THE BONE. AS THE EVENT OCCURRED AFTER SCREW INSERTION INTO THE DISTAL ROUND HOLE, THE SCREW PLACED IN THE DISTAL OBLONG HOLE WAS REMOVED FROM THE BODY. SINCE ONE SCREW REMAINS INSERTED DISTALLY, THE PHYSICIAN ASSESSED THAT STABLE FIXATION HAD BEEN ACHIEVED. UPON INSPECTION OF THE TARGETING DEVICE AT THE JAPAN DISTRIBUTION CENTER, IT WAS CONFIRMED THAT THE DRILL INTERFERES WITH THE NAIL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25138086 ↗InjuryDec 22, 2020Migration; Migration
AS REPORTED: "THE INDICATION WAS LEFT UPPER ARM BONE FRACTURE. AFTER THE INITIAL OPERATION(PERFORMED ON (B)(6) 2020), ON (B)(6) 2020, DURING CLINICAL OBSERVATION THE BACK OUT OF THE SCREW WAS FOUND OUT IN X-RAYS. IT IS JUDGED THAT THE BONE UNION HAS NOT BEEN ACHIEVED AND THE UNION WAS POOR. SO, THE SCREW WAS REMOVED AND REINSERTED WITH ANOTHER SCREW ON THE SAME DATE. THE PROCEDURE COMPLETED WITHOUT PROBLEM.".
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture
MAUDE report 11060355 ↗InjuryDec 22, 2020Migration; Migration
AS REPORTED: "THE INDICATION WAS LEFT UPPER ARM BONE FRACTURE. AFTER THE INITIAL OPERATION(PERFORMED ON (B)(6) 2020). ON (B)(6) 2020, DURING CLINICAL OBSERVATION THE BACK OUT OF THE SCREW WAS FOUND OUT IN X-RAYS. IT IS JUDGED THAT THE BONE UNION HAS NOT BEEN ACHIEVED AND THE UNION WAS POOR. SO, THE SCREW WAS REMOVED AND REINSERTED WITH ANOTHER SCREW ON THE SAME DATE. THE PROCEDURE COMPLETED WITHOUT PROBLEM."
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture
MAUDE report 11060344 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page