Skip to content

NA 1896-4028SRecalls & reported problems

Stryker GmbH · Locking Screw, Fully Threaded · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540202338), received between Dec 22, 2020 and May 11, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration2
  • Positioning Failure1

Patient problems

  • Failure of Implant2
  • Nonunion/Delayed-union Bone Fracture2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Positioning Failure

    DURING T2PH SURGERY, THE SCREW PLACED IN THE MOST DISTAL OVAL HOLE PROTRUDED ANTERIORLY BEYOND THE BONE. AS THE EVENT OCCURRED AFTER SCREW INSERTION INTO THE DISTAL ROUND HOLE, THE SCREW PLACED IN THE DISTAL OBLONG HOLE WAS REMOVED FROM THE BODY. SINCE ONE SCREW REMAINS INSERTED DISTALLY, THE PHYSICIAN ASSESSED THAT STABLE FIXATION HAD BEEN ACHIEVED. UPON INSPECTION OF THE TARGETING DEVICE AT THE JAPAN DISTRIBUTION CENTER, IT WAS CONFIRMED THAT THE DRILL INTERFERES WITH THE NAIL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25138086 ↗
  • InjuryDec 22, 2020Migration; Migration

    AS REPORTED: "THE INDICATION WAS LEFT UPPER ARM BONE FRACTURE. AFTER THE INITIAL OPERATION(PERFORMED ON (B)(6) 2020), ON (B)(6) 2020, DURING CLINICAL OBSERVATION THE BACK OUT OF THE SCREW WAS FOUND OUT IN X-RAYS. IT IS JUDGED THAT THE BONE UNION HAS NOT BEEN ACHIEVED AND THE UNION WAS POOR. SO, THE SCREW WAS REMOVED AND REINSERTED WITH ANOTHER SCREW ON THE SAME DATE. THE PROCEDURE COMPLETED WITHOUT PROBLEM.".

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 11060355 ↗
  • InjuryDec 22, 2020Migration; Migration

    AS REPORTED: "THE INDICATION WAS LEFT UPPER ARM BONE FRACTURE. AFTER THE INITIAL OPERATION(PERFORMED ON (B)(6) 2020). ON (B)(6) 2020, DURING CLINICAL OBSERVATION THE BACK OUT OF THE SCREW WAS FOUND OUT IN X-RAYS. IT IS JUDGED THAT THE BONE UNION HAS NOT BEEN ACHIEVED AND THE UNION WAS POOR. SO, THE SCREW WAS REMOVED AND REINSERTED WITH ANOTHER SCREW ON THE SAME DATE. THE PROCEDURE COMPLETED WITHOUT PROBLEM."

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 11060344 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page