NA 4118127090Recalls & reported problems

Stryker Corporation · Dual Cut Sagittal Blade · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 04546540216885), received between May 3, 2019 and Oct 28, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product3
  • Break1
  • Fracture1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DELIVERED AS UNSTERILE PRODUCT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399876 ↗
  • MalfunctionOct 30, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY RECEIVED IN SUCH A MANNER TO INDICATE THAT ITS STERILITY HAD BEEN COMPROMISED. - 2 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20569737 ↗
  • MalfunctionDec 20, 2022Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED THAT THERE WAS A POTENTIAL STERILITY BREACH ON THE PACKAGING OF THE DEVICE, DURING RECEIPT OF THE DEVICE, A HOLE WAS IDENTIFIED IN THE PACKAGING. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16018496 ↗
  • MalfunctionOct 11, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT THERE WAS A POTENTIAL STERILITY BREACH ON THE PACKAGING OF THE DEVICE, DURING A PROCEDURE, THE POLYBAG DID NOT COMPLETE ASEPTIC TRANSFER AS IT WAS STUCK TO THE INNER PACKAGING SEAL. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15576761 ↗
  • MalfunctionAug 3, 2022Fracture; Fracture

    IT WAS REPORTED VIA MEDWATCH REPORT (B)(4), THAT DURING A RIGHT TOTAL KNEE ARTHROPLASTY, THE TOOTH BROKE FROM THE CUTTING END OF THE BLADE. IT WAS ALSO REPORTED THAT THE MISSING TOOTH WAS NOT LOCATED, AN X-RAY WAS TAKEN WHICH WAS NEGATIVE FOR FOREIGN BODY. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO DELAYS AS A RESULT OF THIS EVENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15157391 ↗
  • MalfunctionMay 3, 2019Break; Break

    IT WAS REPORTED THAT THE BLADE BROKE AT THE MOUNT DURING A KNEE REPLACEMENT SURGERY. IT WAS FURTHER REPORTED THAT THE PIECE WAS RETRIEVED AND THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT BLADE. IT WAS ALSO REPORTED THAT THERE WERE NO DELAYS AND NO ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8578531 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →