Skip to content

OASYS 48552314Recalls & reported problems

Stryker Corporation · POLYAXIAL SCREW · Class 2 · Implant

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04546540371683), received between Jul 26, 2016 and Apr 30, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem5
  • Device Dislodged or Dislocated3
  • Break2
  • Difficult to Insert2
  • Failure to Advance2
  • Mechanical Jam2
  • Detachment Of Device Component1
  • Difficult To Position1

Patient problems

  • No Known Impact Or Consequence To Patient5
  • No Clinical Signs, Symptoms or Conditions3
  • Injury1
  • No Consequences Or Impact To Patient1
  • No Information1
  • Tissue Damage1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21939289 ↗
  • MalfunctionApr 30, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE TULIPS OF TWO OASYS POLYAXIAL SCREWS DISENGAGED INTRA-OPERATIVELY WHILE REMOVING THE SCREWS AND THAT THE SCREW SHANKS REMAIN IMPLANTED IN THE PATIENT. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21939283 ↗
  • MalfunctionApr 3, 2020Mechanical Jam; Mechanical Jam

    STRYKER REPRESENTATIVE REPORTED THAT AN OASYS 3.5 X 14MM POLYAXIAL SCREW BECAME JAMMED AFTER STERILIZATION PRE-OPERATIVELY. SURGERY WAS SUCCESSFULLY COMPLETED WITH A REPLACEMENT DEVICE. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 9924280 ↗
  • MalfunctionOct 10, 2016Mechanical Problem; Device Dislodged or Dislocated; Break; Mechanical Problem; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE SCREW WAS BEING REMOVED AND TULIP RIPPED OFF. I HAVE ONLY THE TULIP AND AM RETURNING IT. THE SURGICAL DELAY WAS TEN MINUTES WHILE THEY HAD TO GET THE SCREW SHANK OUT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6015666 ↗
  • MalfunctionSep 26, 2016Mechanical Problem; Detachment Of Device Component; Mechanical Problem

    IT WAS REPORTED THAT THE SURGEON PERFORMED AN OCCIPUT-T2 REVISION/EXTENSION TO T3.(CAPTURED IN (B)(4)). SURGEON'S PRE-OP PLAN WAS TO REMOVE AND RE-INSTRUMENT C1, T1, T2 AND EXTEND AND ADD SCREWS TO T3. ALONG WITH THIS, HE PLANNED TO HARVEST A RIB AND WIRE IT BETWEEN THE C1 ARCH AND SPINOUS PROCESS. LASTLY, CHECKING SCREWS FOR GENERAL PURCHASE AND POTENTIALLY ADJUSTING THE OCCIPITAL PLATE IF NEEDED. HE REMOVED AND RE-INSTRUMENTED ACCORDING TO PLAN WITH C1, T2 & T3. HE PLACED 5.5MM SCREWS INTO T2 & T3. HE ALSO REMOVED 3 BONES SCREWS FROM THE OCCIPITAL PLATE, ALL OF WHICH WERE 4.5MM RESCUE SCREWS TO BEGIN WITH. THIS ALLOWED HIM WITH NO RESCUE BUT TO SIMPLY ADD 2MM TO THE LENGTH AND RE-INSTRUMENT. HE THEN STARTED TO BEND HIS FIRST VITALLIUM ROD FOR THE LEFT SIDE OF THE CONSTRUCT. HE BEGAN PERSUADING THE ROD USING A GUN PERSUADER AND ROD FORK. HE PLACED 3 OR 4 BLOCKERS AND WAS WORKING HIS WAS UP TO C3 WHEN THE TULIP OF HIS C3 SCREW POPPED OFF OF THE SCREW SHANK. HE THEN HAD TO REMOVE THE BLOCKERS AND FISH OUT THE SHANK. HE ATTEMPTED TO PLACE A 4.5X24MM RESCUE SCREW WHERE THAT SCREW PREVIOUSLY WAS BUT WAS HAVING TROUBLE ADVANCING THE SCREW AFTER IT WAS ABOUT HALFWAY IN. IT THEN CROSS-THR

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5978439 ↗
  • InjuryAug 17, 2016Mechanical Problem; Difficult To Position; Malposition of Device; Mechanical Problem; Difficult To Position

    IT WAS REPORTED THAT AN ANGLE SCREW PLACED IN CERVICAL SPINE, ROD PLACED ON CONSTRUCT, BLOCKER PLACED, FINAL TIGHTENED. X-RAY TAKEN, SURGEON WAS UNHAPPY WITH THE SCREW, THEREFORE WANTED TO REPOSITION IT. THE SCREW WAS NO LONGER POLYAXIAL AND FIXED IN POSITION, SCREWDRIVER UNABLE TO ENGAGE WITH SCREW, OR POLY ADJUSTMENT DRIVER. THEREFORE SCREW HAD TO BE TURNED AND PULLED OUT, DAMAGE TO JOINT OF CERVICAL BONE. SCREW INSPECTED BY REP, IT APPEARS THE SHANK OF THE SCREW HAS BEEN PUSHED DOWN AND ALMOST POPPED THROUGH THE TULIP HEAD, THEREFORE FIXED IN POSITION.

    Patient problems: Injury; Tissue Damage

    MAUDE report 5884476 ↗
  • MalfunctionJul 28, 2016Difficult to Insert; Mechanical Problem; Failure to Advance; Mechanical Jam; Difficult to Insert; Mechanical Problem

    IT WAS REPORTED THAT; SURGEON COULD NOT ADVANCE OR BACK OUT SCREW AFTER INITIALLY BEING INSERTED. SURGEON USED POLYAXIAL DRIVER THEN USED SCREWDRIVER AND COULD NOT ADJUST SCREW. HAD TO HELICOPTER SCREW OUT TWICE. UPDATE; THERE WAS OVER AN HOUR SURGICAL DELAY TO HELICOPTER OUT SCREWS THEN RE INSERT THEM. (2X)

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5829179 ↗
  • InjuryJul 28, 2016Difficult to Insert; Mechanical Problem; Difficult to Remove; Failure to Advance; Difficult to Insert; Mechanical Problem

    IT WAS REPORTED THAT; SURGEON COULD NOT ADVANCE OR BACK OUT SCREW AFTER INITIALLY BEING INSERTED. SURGEON USED POLYAXIAL DRIVER THEN USED SCREWDRIVER AND COULD NOT ADJUST SCREW. HAD TO HELICOPTER SCREW OUT TWICE. UPDATE; THERE WAS OVER AN HOUR SURGICAL DELAY TO HELICOPTER OUT SCREWS THEN RE INSERT THEM. (2X)

    Patient problems: No Information; No Known Impact Or Consequence To Patient

    MAUDE report 5829177 ↗
  • MalfunctionJul 26, 2016Break

    IT WAS REPORTED THAT THE SCREWS STRIPPED DURING THE PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5821838 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.