GAMMA 3060-0090SRecalls & reported problems
Stryker GmbH · Lag Screw, Ti · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker Gamma3 lag screw is the screw that locks into the company's intramedullary nail used to repair hip and femur fractures. The 36 reports mostly describe hardware seating or sliding incorrectly inside the nail during insertion or removal, along with bone fractures and implants that fail to heal. None of the reports list a death, and 35 mention an injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 04546540374929), received between Aug 10, 2018 and Feb 28, 2022 — classified as 35 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available19
- Adverse Event Without Identified Device or Use Problem6
- Fracture5
- Insufficient Device Problem Information3
- Defective Device1
- Device Difficult to Setup or Prepare1
- Failure to Osseointegrate1
- Positioning Problem1
Patient problems
- Death19
- No Code Available8
- Failure of Implant4
- Nonunion/Delayed-union Bone Fracture3
- Bone Fracture(s)2
- Hip Fracture2
- No Information2
- Impaired Healing1
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 28, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
IT HAS BEEN REPORTED THE ASSEMBLY OF THE GAMMA NAIL DOES NOT COINCIDE WITH THE LAG SCREW: WHEN TRYING TO INSERT THE CEPHALIC SCREW, EXCESSIVE FORCE HAD TO BE USED AND IT WAS NOTICEABLE THAT IT DID NOT SLIDE WELL INSIDE THE NAIL. WE CHECKED THAT EVERYTHING WAS WELL CONNECTED, THAT IT WAS THE GUIDE THAT HAD TO BE, THE ANGULATION THAT IT HAD TO BE, EVERYTHING VERIFIABLE. WE ENDED UP REMOVING EVERYTHING, DISASSEMBLED AND REASSEMBLED OUTSIDE AND TESTING OUTSIDE, IT CONTINUED TO RUB AGAINST THE CEPHALIC SCREW WHEN PASSING INSIDE OF THE NAIL. IN THE END WE PUT A NAIL AND A NEW SCREW. THE PROCEDURE WAS DELAYED BY 60 MIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13620279 ↗InjuryMay 9, 2019Defective Device; Positioning Problem
THE FOLLOWING REPORT WAS RECEIVED FROM THE ITALY COMPETENT AUTHORITY ORG. (MOH): DURING THE REMOVAL OF A GAMMA NAIL LONG, USED TO REDUCE A MULTIFRAGMENTED RIGHT FEMUR FRACTURE, DISTALLY AND PROXIMALLY BLOCKED, THE SURGEON HAS NOTICED VISIBLE SIGNS OF RUBBING OF THE CEPHALIC SCREW ON THE NAIL IN AN ABNORMAL LOCATION: OUT OF THE CHANNEL THAT NORMALLY HOUSES THE CEPHALIC SCREW. THE SYSTEM IS NORMALLY USED PERCUTANEOUSLY, SO WITHOUT A DIRECT VIEW OF THE OPERATOR BUT ONLY VIA RX. THE SYSTEM IS COLLIMATED BY CANNULAS THAT ALLOW THIS PROCEDURE. SO CUSTOMER REPORTS THAT THE INCIDENT IS INEXPLICABLE EXCEPT WITH AN ALLEGED DEFECT IN THE INSTRUMENTS. ADVERSE EVENT: ASYMMETRY, LACK OF BONE CALCIFICATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8596809 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY AND RE-OSTEOSYNTHESIS WAS PERFORMED. THE 111 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7781156 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTESYNTHESIS WAS PERFORMED. THE 222 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death; Nonunion/Delayed-union Bone Fracture
MAUDE report 7781131 ↗InjuryAug 14, 2018Insufficient Device Problem Information
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTEOSYNTHESIS WAS PERFORMED. THE 198 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death; Nonunion/Delayed-union Bone Fracture
MAUDE report 7781118 ↗InjuryAug 14, 2018Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO PRIMARY OSTEOARTHRITIS AND SEQUELE AFTER HIP FRACTURE. THE 191 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7781081 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY (RE-OSTESYNTHESIS) DUE TO UNKNOWN REASON. THE 175TH OUT OF 290 CASES.
Patient problems: Failure of Implant
MAUDE report 7780931 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 141 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death; Failure of Implant
MAUDE report 7780818 ↗InjuryAug 14, 2018Failure to Osseointegrate
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTEOSYNTHESIS WAS PERFORMED. THE 46 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Bone Fracture(s); Death; Impaired Healing
MAUDE report 7780808 ↗InjuryAug 14, 2018Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(4) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(4) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO SEQUELE AFTER HIP FRACTURE. NINETY OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780790 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO CUTOUT AND MALUNION. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 247 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780687 ↗InjuryAug 14, 2018Insufficient Device Problem Information
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTEOSYNTHESIS WAS PERFORMED. THE 258 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780686 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. RE-OSTEOSYNTHESIS WAS PERFORMED. THE 250 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780617 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO NONUNION AND CUTOUT. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 288 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death; Nonunion/Delayed-union Bone Fracture
MAUDE report 7780522 ↗InjuryAug 14, 2018Fracture
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE (B)(4) COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE, NEW FRACTURE AROUND IMPLANT AND NONUNION. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 273 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780476 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTEOSYNTHESIS WAS PERFORMED. THE 274 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780474 ↗InjuryAug 14, 2018Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. RE-OSTESYNTHESIS WAS PERFORMED. THE 229 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780446 ↗InjuryAug 14, 2018Fracture
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE (B)(4) COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO SEQUELE AFTER HIP FRACTURE. THE 251 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780444 ↗InjuryAug 14, 2018Fracture
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO NEW FRACTURE AROUND IMPLANT. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 241 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Death
MAUDE report 7780410 ↗InjuryAug 14, 2018Fracture
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA 3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO NEW FRACTURE AROUND IMPLANT. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. THE 188 OUT OF 290 CASES.
Patient problems: Hip Fracture
MAUDE report 7780337 ↗InjuryAug 14, 2018Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO PRIMARY OSTEOARTHRITIS AND SEQUELA AFTER HIP FRACTURE. THE 153 OUT OF 290 CASES.
Patient problems: No Code Available
MAUDE report 7780240 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE (B)(4) COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO MALUNION. RE-OSTESYNTHESIS WAS PERFORMED; 168 OUT OF 290 CASES.
Patient problems: No Information
MAUDE report 7780239 ↗InjuryAug 14, 2018Insufficient Device Problem Information
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN (B)(6) 2005 AND (B)(6) 2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO OSTEOSYNTHESIS FAILURE. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY. ABOUT 104 OUT OF 290 CASES.
Patient problems: No Code Available
MAUDE report 7780238 ↗InjuryAug 14, 2018Appropriate Device Problem Term/Code Not Available
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE NORWEGIAN HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE NORWEGIAN HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE 290 COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY (REMOVAL OF IMPLANT) DUE TO AVASCULAR NECROSIS. THE 146 OUT OF 290 CASES.
Patient problems: Necrosis
MAUDE report 7780216 ↗InjuryAug 14, 2018Adverse Event Without Identified Device or Use Problem
THE MANUFACTURER BECAME AWARE OF ADVERSE EVENTS FROM THE (B)(6) HIP FRACTURE REGISTER. THE TITLE OF THIS REPORT IS ¿GAMMA 3 IN THE (B)(6) HIP FRACTURE REGISTER¿ WHICH IS ASSOCIATED WITH THE GAMMA3 LOCKING NAIL SYSTEM, WITHIN THAT REPORT, POST-OPERATIVE COMPLICATIONS/ ADVERSE EVENTS WERE REPORTED, WHICH OCCURRED BETWEEN 1-JAN-2005 AND 31-DEC-2016. A REVIEW OF THE COMPLAINT HANDLING DATABASE REVEALED THAT THE EVENTS HAVE NOT BEEN REPORTED PREVIOUSLY TO STRYKER, THEREFORE (B)(4) COMPLAINTS WERE INITIATED FOR THE REVISION SURGERIES RETROSPECTIVELY FOR DIFFERENT ADVERSE EVENTS MENTIONED IN THE REPORT. THIS PRODUCT INQUIRY ADDRESSES REVISION SURGERY DUE TO CUTOUT AND NEW FRACTURE AROUND IMPLANT. THE PATIENT RECEIVED A HIP BIPOLAR HEMIARTHROPLASTY; 78 OUT OF 290 CASES. IT WAS ALSO INDICATED THAT PATIENT DIED WITHIN 60 DAYS OF REVISION SURGERY HOWEVER THE CAUSE AND DATE OF DEATH WILL NOT BE DISCLOSED.
Patient problems: Bone Fracture(s); Death
MAUDE report 7780205 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page