- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540375025), received between Jun 26, 2019 and Sep 1, 2026 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Fracture3
- Mechanical Jam1
Patient problems
- Failure of Implant2
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 1, 2026Fracture
ANSM reported that:Date of occurrence: (b)(6) 2026.Description of the incident: Implantation of a Gamma 3 nail on (b)(6) 2025, for a left femoral pertrochanteric fracture that had been progressing for one month and was complicated by a consolidated metastatic lesion from prostate cancer. (b)(6) 2026: Removal of the gamma nail, following the diagnosis of a left pertrochanteric pseudarthrosis resulting from radiation therapy to the metastases of the lesser trochanter, with fracture of a gamma nail implanted in (b)(6) 2025.Patient¿s current status: Revision surgery with removal of the nail and implantation of a total hip prosthesis with dual mobility.Actions taken at the healthcare facility for the patient¿s care: Removal of the nail on (b)(6) 2026."
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 50547353 ↗InjuryNov 7, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT ON (B)(6) 2022 THE PATIENT WAS COMMITTED IN THE HOSPITAL FOR A RUPTURE OF THE OSTHEOSYNTHESIS MATERIAL IN THE LEFT HIP THAT WAS INITIALLY IMPLANTED IN (B)(6) 2022, SPECIFICALLY A TITANIUM NAIL.
Patient problems: Failure of Implant
MAUDE report 15749508 ↗InjuryMay 11, 2022Break; Fracture; Break; Fracture
IT HAS BEEN REPORTED THAT THE NAIL WAS BROKEN AND THE PATIENT REQUIRED A REVISION SURGERY.
Patient problems: Failure of Implant
MAUDE report 14354172 ↗MalfunctionNov 19, 2021Mechanical Jam; Mechanical Jam
CUSTOMER REPORTED THE FOLLOWING ISSUE REGARDING A GAMMA NAIL REF. (B)(4), LOT K05CC65 AND THE ANCILLARY MATERIAL NECESSARY FOR ITS INSTALLATION (NAIL, TARGET DEVICE, BOLT): "DURING AN OPERATION ON (B)(6) 2021, THE MATERIAL WAS FITTED WITHOUT ANY PROBLEMS UNTIL THE NAIL HAD TO BE REMOVED FROM THE TARGET DEVICE TO COMPLETE THE OPERATION: IT WAS IMPOSSIBLE TO UNSCREW. THE CONSEQUENCES WERE A DOUBLING OF THE OPERATION TIME DUE TO THE NEED TO REMOVE EVERYTHING TO START THE WHOLE OPERATION AGAIN, AND THE NEED TO USE ANOTHER ANCILLARY AND ANOTHER NAIL LESS ADAPTED TO THE PATIENT".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12849558 ↗InjuryJun 26, 2019Break
PHARMACIST REPORTED THAT A GAMMA 3 NAIL, AFTER IMPLANTATION IN A PATIENT ON (B)(6) 2019, BROKE AT THE CERVICAL SCREW. FRACTURE OF THE GREATER TROCHANTER OCCURRED. MEDICAL INTERVENTION WAS REQUIRED TO REMOVE THE BROKEN NAIL. PROSTHETIC HIP WAS IMPLANTED DURING RECOVERY SURGERY.
Patient problems: Bone Fracture(s)
MAUDE report 8735323 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.