GAMMA 3130-2170SRecalls & reported problems
Stryker GmbH · Trochanteric Nail Kit, Ti · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540375148), received between Nov 15, 2018 and Oct 25, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Patient-Device Incompatibility1
Patient problems
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
- No Known Impact Or Consequence To Patient1
- Pain1
- Perforation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "AFTER SURGERY, THE U-LAG SCREW CUT OUT THE FEMORAL HEAD WITH U-BLADE DEFORMATION AND THE REVISION SURGERY WAS PERFORMED."
Patient problems: Perforation
MAUDE report 18001391 ↗InjurySep 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING INSERTION OF THE NAIL, A FRACTURE OCCURRED AT ANOTHER SITE. IT WAS DETERMINED THE SURGERY COULD BE CONTINUED AND THE PROCEDURE WAS COMPLETED AS PLANNED. NO TREATMENT WAS PERFORMED FOR THE SUBTROCHANTERIC FRACTURE.
Patient problems: Bone Fracture(s)
MAUDE report 12478236 ↗InjuryMar 12, 2019Break; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE GAMMA NAIL (3130-2170S) WAS INSERTED ABOUT ONE YEAR BEFORE. THE DISTAL OF THE NAIL WAS BROKEN AND IT WAS REVISED TO GAMMA LONG NAIL AT (B)(6) 2019. AT REVISION SURGERY, IT WAS 130° NAIL TO DETAIN, BUT ONLY THE 125° NAIL WAS PREPARED. THE REASON OF THIS EVENT WAS MISCOMMUNICATION OF SURGEON AND SR. SURGERY STOPPED ABOUT 90 MIN WITH ANESTHESIA ON THE PATIENT TO WAIT FOR THE ARRIVAL OF THE IMPLANT. AFTER THAT, THE SURGERY WAS FINISHED WITH NO PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8414661 ↗InjuryNov 15, 2018Patient-Device Incompatibility
CUT OUT OF U LAG SCREW OCCURS. PATIENT FELT PAIN ON HIS HIP FOR SEVERAL WEEKS AND DIFFICULT TO WALK. THEN, WHEN FALLING ON (B)(6), VISITED IN THE HOSPITAL AND AN X - RAY FILM WAS TAKEN, IT TURNED OUT THAT THE LAG SCREW HAD CUT OUT AND REACHED THE ACETABULAR. THERE IS NO DISLOCATION OF THE PIECES OF THE BONY SKULL, AND ONLY THE LAG SCREW HAS CUT OUT. AS A MEASURE AFTER OCCURRENCE, ON (B)(6) REPLACED 120 ° GAMMA NAIL.
Patient problems: Ambulation or Postural Difficulties; Pain
MAUDE report 8074995 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page