GAMMA 3420-0280SRecalls & reported problems
Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540375186), received between Nov 16, 2018 and Feb 10, 2022 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Fracture1
- Insufficient Device Problem Information1
- Material Twisted/Bent1
Patient problems
- Failure of Implant2
- Device Embedded In Tissue or Plaque1
- Fall1
- Implant Pain1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryFeb 10, 2022Break; Fracture; Break; Fracture
AS REPORTED: "PATIENT FELL ON (B)(6) 2022 AND DEVELOPED PAIN. THE PATIENT WAS EXAMINED ON (B)(6) AND IT WAS FOUND THAT THE IMPLANT WAS BROKEN AND THE FRACTURE WAS DISLOCATED. THE FRACTURE WAS PSEUDO ARTICULAR, BUT THE PATIENT WAS TOO OLD, SO ONLY THE IMPLANT REMOVAL WAS DONE."
Patient problems: Fall; Implant Pain
MAUDE report 13508167 ↗InjuryDec 10, 2021Break; Break
AS REPORTED: "THE NAIL WAS BROKEN OFF AT THE LAG SCREW HOLE DUE TO PSEUDOARTHROSIS." DEVICE EXPLANTATION OF THE DEVICE WAS PERFORMED.
Patient problems: Failure of Implant
MAUDE report 12976145 ↗InjuryNov 16, 2018Insufficient Device Problem Information
ON (B)(6) 2018: REMOVAL OF NAIL REMAINED IN THE PATIENT'S BODY WAS DONE. HOWEVER, THE DISTAL PART OF THE NAIL WAS NOT REMOVED AND IT WAS LEFT AND CHS IMPLANT (SYNTHES) WAS DEPLOYED. REV # 2.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8081334 ↗InjuryNov 16, 2018Break; Material Twisted/Bent
AT (B)(6) 2018, OPERATION WAS DONE WITH LONG GAMMA NAIL. AT (B)(6) 2018, SUDDENLY SHE FELT A PAIN WHEN SHE WAS GOING DOWN THE HILL. THEREFORE, SHE VISITED THE HOSPITAL AND CONFIRMED BREAKING OF THE NAIL AROUND THE LAG SCREW HOLE. AT (B)(6) 2018, REVISION SURGERY IS PLANNED. AFTER THE PROXIMAL BROKEN NAIL AND LAG SCREW WERE REMOVED, EXTRACTION HOOK (18066160) WAS HANGED TO THE DISTAL PART OF NAIL AND HIT WITH HAMMER AND ABOUT 1MM OF THE TIP OF THE EXTRACTION HOOK WAS LACKED AND THE IT WAS NOT ABLE TO BE HANGED TO NAIL. THEREFORE, NEW HOOK WAS USED, BUT THE TIP OF THE HOOK WAS BENT AFTER IT WAS HIT WITH HAMMER AND NOT ABLE TO BE USED. THEREFORE, THE DISTAL PART OF THE NAIL AND BROKEN HOOK WAS LEFT IN THE PATIENT BODY AND OPERATION WAS FINISHED. REMOVAL OF NAIL REMAINED IN THE PATIENT'S BODY WAS DONE. HOWEVER, THE DISTAL PART OF THE NAIL WAS NOT REMOVED AND IT WAS LEFT AND CHS IMPLANT (SYNTHES) WAS DEPLOYED. REVISION # 2 REPORTED UNDER SEPARATE PI.
Patient problems: Failure of Implant; Pain
MAUDE report 8081152 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page