Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04546540375308), received between Sep 4, 2015 and Oct 10, 2025 — classified as 7 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Break3
- Improper or Incorrect Procedure or Method1
- Inadequacy of Device Shape and/or Size1
- Migration1
Patient problems
- Failure of Implant3
- No Known Impact Or Consequence To Patient2
- Pain2
- Fall1
- Impaired Healing1
- Implant Pain1
- Limb Fracture1
- No Code Available1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryOct 10, 2025Fracture; Fracture
AS REPORTED: "A PATIENT WHO UNDERWENT SURGERY ON (B)(6) 2025, VISITED THE HOSPITAL ON (B)(6) 2025 COMPLAINING OF PAIN. AN X-RAY REVEALED A BROKEN NAIL. THE BROKEN AREA WAS NEAR THE LAG SCREW HOLE.".
Patient problems: Pain
MAUDE report 23265547 ↗InjuryAug 17, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE NAIL BROKE WHICH RESULTED IN BHA REVISION SURGERY. THE NAIL BREAKAGE WAS NOTICED DUE TO PAIN APPEARING WITHOUT TRAUMATIC FACTORS AND IT WAS NOTICED THAT THERE WAS A DELAYED UNION.
Patient problems: Failure of Implant; Impaired Healing; Implant Pain
MAUDE report 15240186 ↗InjuryFeb 22, 2022Break; Break
AS REPORTED: "THE FALL RESULTED IN A NEW FRACTURE AT THE DISTAL END OF THE NAIL AND SUBTROCHANTERIC RE-FRACTURE AND NAIL DEFORMITY. REVISION SURGERY IS SCHEDULED ON JANUARY 28 OR 29."
Patient problems: Failure of Implant; Fall; Limb Fracture
MAUDE report 13577965 ↗InjuryOct 20, 2020Fracture; Fracture
AS REPORTED: "ON (B)(6) 2020, THE PATIENT UNDERWENT THE OPERATION FOR FRACTURE OF THE PROXIMAL FEMUR. ON (B)(6) 2020, THE PATIENT UNDERWENT REVISION SURGERY DUE TO BREAKAGE OF THE IMPLANT. IT IS UNCERTAIN WHEN THE IMPLANT BROKE." THE ADDITIONAL INFORMATION WAS RECEIVED: "TROCHANTERIC FRACTURE REVERSE TYPE (REVERSE OBLIQUE FRACTURE UNDER LUG SCREW)NAIL WAS USED IN THE OPERATION OF THE SECOND REVISION, AND THE PROGRESS IS GOOD. A GAP REMAINED IN THE FRACTURED PART IN A DIFFICULT REDUCTION CASE IT IS BELIEVED THAT THE IMPLANT IS LIKELY TO HAVE BROKEN [...]." THE ADDITIONAL INFORMATION ON PATIENT'S HISTORY WAS RECEIVED: "THE PATIENT HAD HISTORY OF LEFT THA, LUMBAR COMPRESSION FRACTURE (CONSERVATIVE THERAPY)."
Patient problems: Failure of Implant
MAUDE report 10708428 ↗MalfunctionApr 3, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
AS REPORTED: "THE IMPLANTING PROCEDURE WAS [PERFORMED] FOR THE LEFT FEMUR, BUT THE RIGHT IMPLANT WAS IMPLANTED. SR NOTICED MISS IMPLANTING THE RIGHT NAIL FOR LEFT FEMUR AFTER PROCEDURE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9924838 ↗InjuryMar 5, 2019Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2018 A LONG GAMMA NAIL WAS IMPLANTED. DURING FOLLOW UP 2 WEEKS AFTER SURGERY, AND THERE WAS NO PROBLEM. HOWEVER DUE TO FALL OVER ON (B)(6) 2019, PATIENT WAS COMPLAINING OF PAIN. DURING REVISION SURGERY ON (B)(6) 2018 CUT OUT THE SCREW WAS DISCOVERED.
Patient problems: Pain
MAUDE report 8392758 ↗InjuryNov 29, 2017Fracture; Improper or Incorrect Procedure or Method
THE PHYSICIAN ACCIDENTALLY DEPLOYED THE RIGHT LONG GAMMA NAIL ON THE LEFT. ALTHOUGH IT FIXED AS IT WAS, RECURRENT FRACTURE WAS CONFIRMED AT AFTER OPERATION XP AND THE NAIL WAS REPLACED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7069103 ↗InjurySep 4, 2015Break
IT WAS REPORTED THAT ON (B)(6) 2015, GAMMA3 LONG NAIL SURGERY WAS PERFORMED. SEVEN MONTHS AFTER THE SURGERY, BREAKAGE OF THE NAIL WAS FOUND AND REVISION SURGERY TO BIPOLAR WAS PERFORMED.
Patient problems: No Code Available
MAUDE report 5055710 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page