Skip to content

GAMMA 3425-0280SRecalls & reported problems

Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 04546540375315), received between Mar 29, 2018 and Sep 20, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Fracture2
  • Device Damaged by Another Device1
  • Material Perforation1
  • Positioning Problem1

Patient problems

  • Failure of Implant2
  • Impaired Healing1
  • Implant Pain1
  • No Code Available1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 20, 2024Break; Fracture; Device Damaged by Another Device; Break; Fracture; Device Damaged by Another Device

    IT WAS REPORTED, WHEN THE PATIENT VISITED THE HOSPITAL BECAUSE OF THE PROXIMAL HUMERUS FRACTURE, HE ALSO HAD PAIN IN HIS FEMUR, WHERE THE SURGERY WAS PERFORMED AT ANOTHER HOSPITAL A YEAR AND A HALF AGO. THUS, X-RAY WAS TAKEN AND IT REVEALED PSEUDOARTICULAR OF THE FEMUR AND THE GAMMA3 NAIL FRACTURE AT THE LAG SCREW HOLE PART. THE GAMMA3 NAIL WILL BE REMOVED AT THE SAME TIME AS THE SURGERY FOR THE PROXIMAL HUMERUS FRACTURE ON (B)(6) 2024.

    Patient problems: Failure of Implant; Implant Pain; Nonunion/Delayed-union Bone Fracture

    MAUDE report 20269627 ↗
  • InjuryAug 16, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A NAIL BROKE, RESULTING IN BHA REVISION SURGERY. THE BREAK WAS NOTICED DURING A REGULAR CHECKUP AND IT WAS ALSO NOTED THAT THERE WAS DELAYED UNION OF THE FRACTURE.

    Patient problems: Failure of Implant; Impaired Healing

    MAUDE report 15237939 ↗
  • InjuryMar 29, 2018Material Perforation; Positioning Problem

    THE PATIENT UNDERWENT GAMMA3 SURGERY ON (B)(6) 2017. WHEN THE PATIENT VISITED A HOSPITAL WITH ANOTHER DISEASE, SHORTENING OF RIGHT LEG WAS POINTED OUT, AND THEN TRANSFERRED TO THE HOSPITAL THAT THE INITIAL SURGERY WAS PERFORMED. X-RAY SHOWED A FEMORAL HEAD PERFORATION BUT THE PATIENT DID NOT FEEL ANY PAIN DUE TO SCHIZOPHRENIA, THIS WAS ONE OF THE REASON WHY THE PERFORATION WAS ADVANCED IN THE SURGEON'S OPINION. REVISION SURGERY WAS PERFORMED ON (B)(6) 2018, THE LAG SCREW WHICH REACHED INTESTINAL TRACT WAS REMOVED WITH LAPAROSCOPIC SURGERY, AND THE GAMMA3 NAIL WAS REMOVED SUBSEQUENTLY. THERE WERE NO DAMAGES NOTED TO THE LAG SCREW BY A SET SCREW. IN THE SURGEON'S OPINION, THE MAIN REASON OF THE PERFORATION WAS THE SET SCREW WAS NOT REACHED TO THE LAG SCREW. FEMORAL HEAD PERFORATION, REVISION.

    Patient problems: No Code Available

    MAUDE report 7383206 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates