Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 04546540375322), received between Nov 10, 2021 and Jan 5, 2026 — classified as 4 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Break2
- Off-Label Use1
Patient problems
- Failure of Implant3
- Bone Fracture(s)1
- Implant Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 5, 2026Off-Label Use
AS REPORTED: "BOTH SIDE NAILS (RIGHT AND LEFT) WERE PREPARED. A NAIL OPENED ON THE WRONG SIDE BY THE NURSE, BUT THE PHYSICIAN WAS NOT AWARE OF THIS PRIOR TO USE AND ONLY REALISED DURING THE PROCEDURE. AS A RESULT OF INSERTING THE NAIL AS IS, A FRACTURE OCCURRED ON THE FRONT OUTER SIDE DUE TO COLLISION AT THE NAIL TIP. AFTER INSERTING THE CORRECT NAIL, TWO CCS PINS WERE FIXED AND THEN WRAPPED THE WIRE."
Patient problems: Bone Fracture(s)
MAUDE report 23968280 ↗InjurySep 27, 2022Break; Fracture; Break; Fracture
A PATIENT WAS ADMITTED TO (B)(6) MEDICAL AND (B)(6) HOSPITAL. A GAMMA 3 LONG NAIL IN THE PATIENT'S BODY WAS FOUND TO BE BROKEN. ON (B)(6) 2022 THE PATIENT WAS REVISED TO REMOVE THE BROKEN NAIL. THE IMPLANT WAS INSERTED AT (B)(6) HOSPITAL IN (B)(6) 2021- THE DATE OF THE INITIAL SURGERY IS UNSPECIFIED, BUT IS ESTIMATED TO BE ON (B)(6) 2021.
Patient problems: Failure of Implant
MAUDE report 15493443 ↗InjuryMar 15, 2022Break; Fracture; Break; Fracture
AS REPORTED: "ON (B)(6) 2021, GAMMA 3 LONG NAIL WAS IMPLANTED FOR FEMORAL TROCHANTERIC FRACTURE DUE TO BONE TUMOR. IN FEBRUARY 2022, WITHOUT A FALL, THE PATIENT FELT PAIN DURING DAILY ACTIVITIES AND THE X-RAYS REVEALED NAIL FRACTURE. DURING REVISION, THE BROKEN NAIL WAS REMOVED AND GMRS WAS USED."
Patient problems: Failure of Implant; Implant Pain
MAUDE report 13770790 ↗InjuryNov 10, 2021Fracture; Fracture
AS REPORTED; "DAMAGE TO GAMMA NAILS IN THE BODY. INSERTED ABOUT HALF A YEAR AGO. THE DOCTOR'S VIEW WAS THAT IT WAS A VERY COMPLICATED FRACTURE AND THAT IT WAS DAMAGED BECAUSE THE PATIENT WAS YOUNG AND VERY ACTIVE." THE PATIENT REQUIRED AN IMPLANT EXPLANTATION TO REMOVE THE DEVICE.
Patient problems: Failure of Implant
MAUDE report 12785449 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page