Skip to content

GAMMA 3425-0360SRecalls & reported problems

Stryker GmbH · Long Nail Kit R1.5, Ti, RIGHT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04546540375353), received between Dec 18, 2018 and Dec 8, 2023 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Device Contaminated During Manufacture or Shipping1
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • Failure of Implant2
  • Nonunion/Delayed-union Bone Fracture2
  • Bone Fracture(s)1
  • Injury1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 8, 2023Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT DUE TO THE NAIL AND CONDYLE SCREW BREAKING, THEY WERE REMOVED AND REPLACED TO GMRS.

    Patient problems: Failure of Implant

    MAUDE report 18297559 ↗
  • InjuryDec 20, 2022Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT A GAMMA NAIL BROKE APPROX. 3 MONTHS POST SURGERY, RESULTING IN A REVISION SURGERY.

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 16014667 ↗
  • MalfunctionAug 5, 2020Tear, Rip or Hole in Device Packaging; Device Contaminated During Manufacture or Shipping; Tear, Rip or Hole in Device Packaging; Device Contaminated During Manufacture or Shipping

    AS REPORTED: "THE PHYSICIAN FOUND PLASTIC POWDER ON THE DEVICE, WHEN THE PACKAGING WAS OPENED." ADDITIONALLY REPORTED: - "THE PHYSICIAN CLEANED POWDER ON THE DEVICE AND USED IT." - "THE SPONGE [FOAM OF THE PACKAGING] WAS WORN"

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10370513 ↗
  • InjuryJun 12, 2020Fracture; Fracture

    AS REPORTED: A (B)(6) YEAR OLD MALE PATIENT HAD OPERATION WITH GAMMA NAIL FOR FEMORAL TROCHANTERIC FRACTURE ON (B)(6) 2019 AND WAS DISCHARGED WITHOUT POSTOPERATIVE PROBLEMS. THERE WAS NO PROBLEM WITH X-RAY AFTER 3 MONTHS. HOWEVER, WHEN HE REVISITED THE HOSPITAL ON (B)(6) 2020 AND TOOK AN X-RAY, IT WAS DISCOVERED THAT THE NAIL HAD BROKEN. IT SEEMS THAT THE NAIL WAS OVERLOADED DUE TO THE FALSE JOINT. REVISION (THA) IS SCHEDULED ON (B)(6) 2020.

    Patient problems: Injury; Nonunion/Delayed-union Bone Fracture

    MAUDE report 10150807 ↗
  • InjuryDec 18, 2018Break

    PATIENT HAD RIGHT GAMMA NAIL PUT IN (B)(6) AT ANOTHER HOSPITAL FOR A LEFT FEMUR FRACTURE AND SUBSEQUENTLY BROKE THE DISTAL END OF FEMUR. PATIENT HAD FRACTURE AND NEEDED REOPERATION.

    Patient problems: Bone Fracture(s)

    MAUDE report 8176744 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates