GAMMA 3525-0260SRecalls & reported problems
Stryker GmbH · Long Nail Kit R1.5, Ti, LEFT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 04546540375698), received between Sep 6, 2018 and Dec 22, 2025 — classified as 7 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Fracture3
- Break2
Patient problems
- Injury2
- Nonunion/Delayed-union Bone Fracture2
- Limb Fracture1
- No Code Available1
- Pain1
- Perforation1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryDec 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "DURING INSERTION OF 260MM GAMMA3 LONG NAIL, THE BONE HAD A HAIRLINE FRACTURE. THE 260MM NAIL WAS REMOVED AND THE 320MM NAIL WAS INSERTED INSTEAD TO COMPLETE THE PROCEDURE."
Patient problems: Limb Fracture
MAUDE report 23869626 ↗InjuryOct 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "OVER-TELESCOPING OF THE LAG SCREW DUE TO NON-UNION. THE SCREW DOES NOT HAVE A CUTOUT. THE LAG SCREW PROTRUDED TO LATERAL DUE TO COLLAPSE OF THE FEMORAL NECK. THE PATIENT REQUIRED A REVISION SURGERY."
Patient problems: Perforation
MAUDE report 23318037 ↗InjuryMay 7, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT A NONUNION OCCURRED AND THE NAIL BROKE OFF AT THE LAG SCREW HOLE. IT WAS A SPECIAL CASE WITH REDUCTION DIFFICULTY.
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 14323697 ↗InjuryJan 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "ON (B)(6) 2020 THE INITIAL SURGERY UNDER THE TROCHANTER OF THE FEMUR (LEFT) WAS PERFORMED. ON (B)(6) 2021 RE-OPERATION WAS PERFORMED. THE REMAINING GAP IN THE FRACTURE WOULD HAVE PUT A LOAD ON THE NAIL, CAUSING NONUNION AND NAIL BREAKAGE."
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 11223039 ↗InjuryMay 20, 2020Fracture; Fracture
IT WAS REPORTED, SUBTROCHANTERIC FRACTURE OF THE FEMUR ON (B)(6) 2019. THE LONG GAMMA NAIL WAS USED. ON (B)(6) 2020, THE NAIL WAS BROKEN STARTING FROM THE NAIL LAG SCREW INSERTION HOLE. THA SCHEDULED FOR REVISION.
Patient problems: Injury
MAUDE report 10076710 ↗InjuryMar 27, 2020Fracture; Fracture
IT WAS REPORTED, ON (B)(6) 2019 THE PATIENT UNDERWENT OPERATION WITH GAMMA 3 NAIL. A FEW MONTHS AFTER THE SURGERY, THE PATIENT¿S PAIN CONTINUED. ON (B)(6) 2020 AT THE TIME OF THE HOSPITAL VISIT, BREAKAGE OF THE GAMMA NAIL WAS CONFIRMED. THE REVISION BHA PROCEDURE WAS PERFORMED WITH OTHER COMPANY PRODUCT.
Patient problems: Injury; Pain
MAUDE report 9890722 ↗InjurySep 6, 2018Break
THE PATIENT UNDERWENT A PRIMARY SURGERY WITH GAMMA NAIL ON (B)(6) 2017. ON (B)(6) 2018, IT WAS FOUND THAT THE NAIL WAS BROKEN AT THE TIME OF VISIT FOR FOLLOW-UP. THEREFORE, REVISION SURGERY WAS PERFORMED ON (B)(6) 2018.
Patient problems: No Code Available
MAUDE report 7852940 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page